Literature DB >> 23218695

Change in efficiency of aspheric intraocular lenses based on pupil diameter.

Youngsub Eom1, Eunjoo Yoo, Su-Yeon Kang, Hyo-Myung Kim, Jong-Suk Song.   

Abstract

PURPOSE: To measure the effect of spherical aberration correction by aspheric intraocular lenses (IOLs) based on pupil diameter, and to determine the minimum pupil diameter for each aspheric IOL.
DESIGN: Retrospective cross-sectional study.
METHODS: Eight-six patients (169 eyes) who were implanted with a HOYA AF-1 NY-60 (HOYA Corporation) or Tecnis ZCB00 1-piece IOL (Abbott Medical Optics Inc) were enrolled. Ocular, corneal, and internal spherical aberrations were measured at the 1-month postoperative visit using the Wavefront Analyzer KR-1W (Topcon). Minimum pupil diameter, which is required for each aspheric IOL to be effective, was calculated using a regression equation.
RESULTS: The mean value of internal spherical aberration of the Tecnis ZCB00 group (-0.09 ± 0.094 μm) was lower than that of the HOYA NY-60 group (-0.05 ± 0.072 μm) (P = .005). The original negative spherical aberrations of the HOYA NY-60 (-0.18 μm) were measured at a pupil diameter of 5.6 mm, and for the Tecnis ZCB00 (-0.27 μm) at a pupil diameter of 6.1 mm. The aspheric IOL efficiency dropped to 0% when the pupil diameter was 3.47 mm for the Tecnis ZCB00 group and 3.71 mm for the HOYA NY-60 group.
CONCLUSIONS: When the pupil diameters of patients are smaller than 3.4 mm for the Tecnis ZCB00 and 3.7 mm for the HOYA NY-60, the spherical aberration correction using these aspheric IOLs seems to be ineffective. Approximately 10% of the eyes showed smaller pupil size than the minimum effective diameter under mesopic conditions.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23218695     DOI: 10.1016/j.ajo.2012.09.024

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  3 in total

1.  Comparison of two types of visual quality analyzer for the measurement of high order aberrations.

Authors:  Jing Hao; Lin Li; Fang Tian; Hong Zhang
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

2.  Comparison of visual outcomes after two types of mix-and-match implanted trifocal extended-depth-of-focus and trifocal intraocular lenses.

Authors:  Jung Wan Kim; Youngsub Eom; Wonkyung Park; Jong Suk Song; Ji Won Jeong; Seh Kwang Park; Hyo Myung Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-05-28       Impact factor: 3.535

3.  The efficiency of aspheric intraocular lens according to biometric measurements.

Authors:  Woong-Joo Whang; Junjie Piao; Young-Sik Yoo; Choun-Ki Joo; Geunyoung Yoon
Journal:  PLoS One       Date:  2017-10-16       Impact factor: 3.240

  3 in total