Comparison of fixed tilt and tuned defibrillation waveforms: the PROMISE study.

Michael R Gold1, Jesus Val-Mejias, Frank Cuoco, Mukkaram Siddiqui.   

Abstract

BACKGROUND: All modern defibrillation systems use biphasic shock waveforms. Typically a fixed tilt waveform is used for implantable defibrillators (ICDs), but a tuned waveform with duration based on shock impedance may be superior based on theoretical calculations.
OBJECTIVE: The objective of this study was to compare defibrillation efficacy of fixed tilt and tuned waveforms.
METHODS: PROMISE was designed as a prospective, within-patient, randomized study of defibrillation thresholds (DFTs) comparing a tuned (assuming a 3.5 milliseconds membrane time constant) versus a 50/50% tilt waveform. All patients had a left pectoral implant (active can) and testing was performed with a single coil shocking configuration ("SVC coil OFF"). DFTs were measured in random order with a binary search method in 52 patients, using the high-voltage lead impedance to select the pulse widths for both waveforms.
RESULTS: At the DFT, the tuned waveform had similar delivered energy (10.5 ± 6.3 vs 9.5 ± 5.5 J, P = 0.47), stored energy (13.6 ± 7.9 vs 11.3 ± 6.3 J, P = 0.06), peak current (7.5 ± 3.0 vs 6.8 ± 2.2 A, P = 0.09), and delivered voltage (451.0 ± 134.5 vs 411.5 ± 120.7 V, P = 0.05) compared with the 50/50% tilt waveform.
CONCLUSION: The DFTs for 3.5-millisecond time constant based tuned and 50/50% tilt waveforms are similar using a single coil, left pectoral active can.
© 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 23210764     DOI: 10.1111/jce.12041

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873