| Literature DB >> 23200355 |
Stéphane Fierro1, Christos Comninellis, Yasuaki Einaga.
Abstract
Individual and simultaneous electrochemical detection of iodide and iodine has been performed via cyclic voltammetry on boron doped diamond (BDD) electrodes in a 1M NaClO(4) (pH 8) solution, representative of typical environmental water conditions. It is feasible to compute accurate calibration curve for both compounds using cyclic voltammetry measurements by determining the peak current intensities as a function of the concentration. A lower detection limit of about 20 μM was obtained for iodide and 10 μM for iodine. Based on the comparison between the peak current intensities reported during the oxidation of KI, it is probable that iodide (I(-)) is first oxidized in a single step to yield iodine (I(2)). The latter is further oxidized to obtain IO(3)(-). This technique, however, did not allow for a reasonably accurate detection of iodate (IO(3)(-)) on a BDD electrode.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23200355 DOI: 10.1016/j.talanta.2012.10.002
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057