Comparison of the SphygmoCor and Omron devices in the estimation of pressure amplification against the invasive catheter measurement.

Feng-Hua Ding1, Yan Li, Rui-Yan Zhang, Qi Zhang, Ji-Guang Wang.   

Abstract

BACKGROUND: We previously tested the accuracy of the SphygmoCor and Omron HEM-9000AI devices in the estimation of central blood pressure. In the present study, we investigated these two devices in the estimation of central-to-brachial pressure amplification against the invasive catheter measurement.
METHODS: In 33 individuals undergoing cardiac catheterization, we measured central blood pressure simultaneously by the invasive catheter and each of the two noninvasive devices and brachial blood pressure by the invasive catheter and an automated oscillometric blood pressure monitor of the Omron device. Pressure amplification was calculated as central-to-brachial systolic pressure difference and pulse pressure difference and ratio. The agreement between each of these two noninvasive devices and the invasive catheter was evaluated using the Student's t-test, intraclass correlation analysis, and the Bland-Altman method.
RESULTS: The mean central-to-brachial systolic pressure difference and pulse pressure difference and ratio estimated by Omron were significantly lower than those measured by the catheter (P < 0.001), whereas no difference was observed for SphygmoCor (P ≥ 0.07). Nonetheless, the intraclass correlation coefficients for systolic pressure difference and pulse pressure difference and ratio between the noninvasive and invasive catheter measurements were similar for the two devices, being 0.11 (P = 0.56), 0.38 (P = 0.03), and 0.40 (P = 0.02), respectively, for SphygmoCor, and 0.15 (P = 0.41), 0.23 (P = 0.20), and 0.53 (P = 0.002), respectively, for Omron.
CONCLUSION: If the invasive catheter measurement would be considered as standard, the two noninvasive devices have similar accuracy in the estimation of pressure amplification, but apparently require device-specific criteria for diagnosis. Pulse pressure ratio seems to be a more consistent measure of central-to-brachial pressure amplification.

Entities:  

Mesh:

Year:  2013        PMID: 23188416     DOI: 10.1097/HJH.0b013e32835a8eca

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  8 in total

1.  Older age is associated with greater central aortic blood pressure following the exercise stress test in subjects with similar brachial systolic blood pressure.

Authors:  Masatake Kobayashi; Kazutaka Oshima; Yoichi Iwasaki; Yuto Kumai; Alberto Avolio; Akira Yamashina; Kenji Takazawa
Journal:  Heart Vessels       Date:  2015-08-21       Impact factor: 2.037

2.  Women have significantly greater difference between central and peripheral arterial pressure compared with men: the Bogalusa Heart Study.

Authors:  Rebecca Chester; Gary Sander; Camilo Fernandez; Wei Chen; Gerald Berenson; Thomas Giles
Journal:  J Am Soc Hypertens       Date:  2013-07-12

3.  Central and peripheral blood pressures in relation to plasma advanced glycation end products in a Chinese population.

Authors:  Q-F Huang; C-S Sheng; Y-Y Kang; L Zhang; S Wang; F-K Li; Y-B Cheng; Q-H Guo; Y Li; J-G Wang
Journal:  J Hum Hypertens       Date:  2015-06-18       Impact factor: 3.012

4.  A systematic review of invasive, high-fidelity pressure studies documenting the amplification of blood pressure from the aorta to the brachial and radial arteries.

Authors:  Denis Chemla; Sandrine Millasseau
Journal:  J Clin Monit Comput       Date:  2020-10-09       Impact factor: 1.977

5.  Blood pressure measurements-Shifting the focus from periphery to center.

Authors:  Tibor Fülöp; Sohail Abdul Salim; Johann Herberth; Roberto Pisoni
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-03-10       Impact factor: 3.738

6.  Estimation of coronary wave intensity analysis using noninvasive techniques and its application to exercise physiology.

Authors:  Christopher J Broyd; Sukhjinder Nijjer; Sayan Sen; Ricardo Petraco; Siana Jones; Rasha Al-Lamee; Nicolas Foin; Mahmud Al-Bustami; Amarjit Sethi; Raffi Kaprielian; Punit Ramrakha; Masood Khan; Iqbal S Malik; Darrel P Francis; Kim Parker; Alun D Hughes; Ghada W Mikhail; Jamil Mayet; Justin E Davies
Journal:  Am J Physiol Heart Circ Physiol       Date:  2015-12-18       Impact factor: 4.733

7.  The value of a BP determination method using a novel non-invasive BP device against the invasive catheter measurement.

Authors:  Jinsong Xu; Yanqing Wu; Hai Su; Weitong Hu; Juxiang Li; Wenying Wang; Xin Liu; Xiaoshu Cheng
Journal:  PLoS One       Date:  2014-06-23       Impact factor: 3.240

8.  Association of thyroid hormones with blood pressure and arterial stiffness in the general population: The Dali study.

Authors:  Md Tasneem Jamal; Qing-Lu Li; Qi-Yan Li; Wan-Ying Liang; Li-Hong Wang; Jian-Hang Wei; Quan Liang; Nai-Qing Hu; Li-Hua Li
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-12-27       Impact factor: 3.738

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.