Robotic knee laxity testing: reliability and normative data.

Renata Vauhnik1, Maja Pohar Perme, Massimo G Barcellona, Darja Rugelj, Matthew C Morrissey, France Sevsek.   

Abstract

PURPOSE: To evaluate the reliability of the GeNouRoB knee arthrometer and present normative values of knee anterior laxity using this device on young females.
METHODS: Anterior laxity in both knees was tested in two groups of young, uninjured females using the hamstrings electromyography biofeedback feature of the device. There were 13 participants in the group tested for reliability and 23 for the normative study. Laxity (mm of movement of the proximal tibia in the anterior direction relative to the femur) was calculated at test forces of 134 N and 250 N with values presented for the unstandardised and standardised (relative to stabilisation force) conditions.
RESULTS: The relative reliability (95% limits of agreement) of the device for laxity at a test force of 134 N was 2 to 3mm. Left knee anterior laxity was almost 1mm greater than the right.
CONCLUSIONS: The relative reliability of the GeNouRoB arthrometer is comparable to the KT device. In agreement with previous work on the nonrobotic KT arthrometer, the knee anterior laxity values found with the GeNouRoB are greater in the left as compared to the right knee.
Copyright © 2012 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23178112     DOI: 10.1016/j.knee.2012.10.010

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  5 in total

Review 1.  Anterior cruciate ligament assessment using arthrometry and stress imaging.

Authors:  Eric M Rohman; Jeffrey A Macalena
Journal:  Curr Rev Musculoskelet Med       Date:  2016-06

2.  Multiplanar Knee Laxity and Perceived Function During Activities of Daily Living and Sport.

Authors:  Jeffrey B Taylor; Hsin-Min Wang; Randy J Schmitz; Christopher K Rhea; Scott E Ross; Sandra J Shultz
Journal:  J Athl Train       Date:  2015-11-05       Impact factor: 2.860

3.  Evaluation of reproducibility of robotic knee testing device (GNRB) on 60 healthy knees.

Authors:  Dany Mouarbes; Etienne Cavaignac; Philippe Chiron; Emilie Bérard; Jérôme Murgier
Journal:  J Orthop       Date:  2018-01-30

4.  Combined anterior and rotational laxity measurements allow characterizing personal knee laxity profiles in healthy individuals.

Authors:  Caroline Mouton; Romain Seil; Tim Meyer; Hélène Agostinis; Daniel Theisen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-08-26       Impact factor: 4.342

5.  The Reliability of the GNRB® Knee Arthrometer in Measuring ACL Stiffness and Laxity: Implications for Clinical Use and Clinical Trial Design.

Authors:  Kayla Smith; Nicole Miller; Steve Laslovich
Journal:  Int J Sports Phys Ther       Date:  2022-10-02
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.