Evaluation of torque maintenance of abutment and cylinder screws with Morse taper implants.

Mayara Barbosa Ferreira1, Juliana Aparecida Delben, Valentim Adelino Ricardo Barão, Leonardo Perez Faverani, Paulo Henrique Dos Santos, Wirley Gonçalves Assunção.   

Abstract

The screw loosening of implant-supported prostheses is a common mechanical failure and is related to several factors as insertion torque and preload. The aim of this study was to evaluate the torque maintenance of retention screws of tapered abutments and cylinders of Morse taper implants submitted to retightening and detorque measurements. Two groups were obtained (n = 12): group I-tapered abutment connected to the implant with titanium retention screw and group II-cylinder with metallic base connected to tapered abutment with titanium retention screw. The detorque values were measured by an analogic torque gauge after 3 minutes of torque insertion. The detorque was measured 10 times for each retention screw of groups I and II, totalizing 120 detorque measurements in each group. Data were submitted to ANOVA and Fisher exact test (P < 0.05). Both groups presented reduced detorque value (P < 0.05) in comparison to the insertion torque in all measurement periods. There was a statistically significant difference (P < 0.05) between the detorque values of the first measurement and the other measurement periods for the abutment screw. However, there was no statistically significant difference (P > 0.05) for the detorque values of all measurement periods for the cylinder screw. In conclusion, the abutment and cylinder screws exhibited torque loss after insertion, which indicates the need for retightening during function of the implant-supported prostheses.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23172504     DOI: 10.1097/SCS.0b013e318270fe0e

Source DB:  PubMed          Journal:  J Craniofac Surg        ISSN: 1049-2275            Impact factor:   1.046


  4 in total

1.  Wear of Morse taper and external hexagon implant joints after abutment removal.

Authors:  Abraão M Prado; Jorge Pereira; Filipe S Silva; Bruno Henriques; Rubens M Nascimento; Cesar A M Benfatti; José López-López; Júlio C M Souza
Journal:  J Mater Sci Mater Med       Date:  2017-03-20       Impact factor: 3.896

Review 2.  Systematic analysis of factors that cause loss of preload in dental implants.

Authors:  S Nithyapriya; A S Ramesh; A Kirubakaran; Jeevitha Mani; J Raghunathan
Journal:  J Indian Prosthodont Soc       Date:  2018 Jul-Sep

3.  Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading.

Authors:  Mai Ahmed Yousry El-Sheikh; Tamer Mohamed Nasr Mostafa; Mohamed Maamoun El-Sheikh
Journal:  Int J Implant Dent       Date:  2018-12-03

Review 4.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.