An experimental study of the interface pressure profile during level walking of a new suspension system for lower limb amputees.

Arezoo Eshraghi1, Noor Azuan Abu Osman, Hossein Gholizadeh, Sadeeq Ali, Stefán Karl Sævarsson, Wan Abu Bakar Wan Abas.   

Abstract

BACKGROUND: Different suspension systems that are used within prosthetic devices may alter the distribution of pressure inside the prosthetic socket in lower limb amputees. This study aimed to compare the interface pressure of a new magnetic suspension system with the pin/lock and Seal-In suspension systems.
METHODS: Twelve unilateral transtibial amputees participated in the study. The subjects walked on a level walkway at a self-selected speed. The resultant peak pressure with the three different suspension systems was recorded using F-socket transducers.
FINDINGS: There were significant statistical differences between the three studied suspension systems. Pair-wise analyses revealed that the mean peak pressure (kPa) was lower with the magnetic system than it was with the pin/lock system over the anterior and posterior aspects during one gait cycle (89.89 vs. 79.26 and 47.22 vs. 26.01, respectively). Overall, the average peak pressure values were higher with the Seal-In system than they were with the new magnetic lock and pin/lock system.
INTERPRETATION: The new magnetic system might reduce the pressure within the prosthetic socket in comparison to the pin/lock and Seal-In system during one gait cycle. This is particularly important during the swing phase of gait and may reduce the pain and discomfort at the distal residual limb in comparison to the pin/lock system.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23157843     DOI: 10.1016/j.clinbiomech.2012.10.002

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  7 in total

1.  Textile-based Pressure Sensors for Monitoring Prosthetic-Socket Interfaces.

Authors:  Jordan Tabor; Talha Agcayazi; Aaron Fleming; Brendan Thompson; Ashish Kapoor; Ming Liu; Michael Lee; He Helen Huang; Alper Bozkurt; Tushar Ghosh
Journal:  IEEE Sens J       Date:  2021-01-21       Impact factor: 3.301

2.  Self-reported socket comfort, mobility, and balance-confidence of individuals with transtibial amputation using pinlock vs suction suspension.

Authors:  Mayank Seth; Emma Haldane Beisheim; Maximilian Tobias Spencer; John Robert Horne; Frank Bernard Sarlo; Jaclyn Megan Sions
Journal:  Prosthet Orthot Int       Date:  2021-06-01       Impact factor: 1.672

3.  Evaluation of new suspension system for limb prosthetics.

Authors:  Hossein Gholizadeh; Noor Azuan Abu Osman; Arezoo Eshraghi; Sadeeq Ali; Nooranida Arifin; Wan Abu Bakar Wan Abas
Journal:  Biomed Eng Online       Date:  2014-01-10       Impact factor: 2.819

4.  Gait biomechanics of individuals with transtibial amputation: effect of suspension system.

Authors:  Arezoo Eshraghi; Noor Azuan Abu Osman; Mohammad Karimi; Hossein Gholizadeh; Ehsan Soodmand; Wan Abu Bakar Wan Abas
Journal:  PLoS One       Date:  2014-05-27       Impact factor: 3.240

5.  Clinical implication of interface pressure for a new prosthetic suspension system.

Authors:  Hossein Gholizadeh; Noor Azuan Abu Osman; Arezoo Eshraghi; Nasrul Anuar Abd Razak
Journal:  Biomed Eng Online       Date:  2014-06-30       Impact factor: 2.819

6.  Interface Pressure System to Compare the Functional Performance of Prosthetic Sockets during the Gait in People with Trans-Tibial Amputation.

Authors:  Salvador Ibarra Aguila; Gisel J Sánchez; Eric E Sauvain; B Alemon; Rita Q Fuentes-Aguilar; Joel C Huegel
Journal:  Sensors (Basel)       Date:  2020-12-09       Impact factor: 3.576

Review 7.  Techniques for Interface Stress Measurements within Prosthetic Sockets of Transtibial Amputees: A Review of the Past 50 Years of Research.

Authors:  Ebrahim A Al-Fakih; Noor Azuan Abu Osman; Faisal Rafiq Mahmad Adikan
Journal:  Sensors (Basel)       Date:  2016-07-20       Impact factor: 3.576

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.