Extremely high-pressure dilation with a new noncompliant balloon.

José F Díaz1, Antonio Gómez-Menchero, Rosa Cardenal, Carlos Sánchez-González, Amit Sanghvi.   

Abstract

Calcified nondilatable lesions remain a challenge for the interventional cardiologist, and they are becoming more frequent in the catheterization laboratory as the overall complexity of interventions increases. Stent underexpansion is the main risk factor for restenosis and thrombosis, which is the most feared complication for the interventionalist.In this report, we present our initial experience (8 patients) with a new noncompliant high-pressure balloon, the OPN NC balloon, a double-layered device. When we used this device at 40 atm, our rate of success was reasonable (75%) and our patients experienced no adverse sequelae. The OPN NC balloon appears to offer a new means of dilating lesions or underexpanded stents when other noncompliant balloons have failed. Safety seems reasonable, even at pressures as high as 40 atm.

Entities:  

Keywords:  Angioplasty, balloon, coronary/instrumentation/methods; balloon dilation; calcinosis/therapy; compliance; coronary stenosis/therapy; prosthesis failure; stents

Mesh:

Year:  2012        PMID: 23109756      PMCID: PMC3461667     

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  11 in total

1.  Cutting balloon angioplasty for treatment of calcified coronary lesions.

Authors:  E Karvouni; G Stankovic; R Albiero; T Takagi; N Corvaja; M Vaghetti; C Di Mario; A Colombo
Journal:  Catheter Cardiovasc Interv       Date:  2001-12       Impact factor: 2.692

2.  Nonrandomized comparison of coronary stenting under intravascular ultrasound guidance of direct stenting without predilation versus conventional predilation with a semi-compliant balloon versus predilation with a new scoring balloon.

Authors:  Jose de Ribamar Costa; Gary S Mintz; Stéphane G Carlier; Roxana Mehran; Paul Teirstein; Koichi Sano; Xuebo Liu; Joanna Lui; Yingbo Na; Celia Castellanos; Sinan Biro; Lockeshi Dani; Jason Rinker; Issam Moussa; George Dangas; Alexandra J Lansky; Edward M Kreps; Michael Collins; Gregg W Stone; Jeffrey W Moses; Martin B Leon
Journal:  Am J Cardiol       Date:  2007-06-13       Impact factor: 2.778

3.  Two hugging balloons at high pressures successfully dilate a lesion refractory to routine coronary angioplasty.

Authors:  H Feld; L Valerio; J Shani
Journal:  Cathet Cardiovasc Diagn       Date:  1991-10

4.  Two "buddies" may be better than one: use of two buddy wires to expand an underexpanded left main coronary stent.

Authors:  Jason B Lindsey; Subhash Banerjee; Emmanouil S Brilakis
Journal:  J Invasive Cardiol       Date:  2007-12       Impact factor: 2.022

5.  Successful predilation of a resistant, heavily calcified lesion with cutting balloon for coronary stenting: a case report.

Authors:  Y Asakura; Y Furukawa; S Ishikawa; K Asakura; K Sueyoshi; M Sakamoto; S Takagi; S Takatsuki; K Oyamada; T Okabe; H Mitamura; S Ogawa; T Hinohara
Journal:  Cathet Cardiovasc Diagn       Date:  1998-08

6.  Undefeatable coronary lesion.

Authors:  Muhammad Tariq Farman; Jawaid Akbar Sial; Naveed Ullah Khan; Tariq Masood; Tahir Saghir
Journal:  J Pak Med Assoc       Date:  2011-02       Impact factor: 0.781

7.  Management of resistant coronary lesions by the cutting balloon catheter: initial experience.

Authors:  O F Bertrand; R Bonan; L Bilodeau; J F Tanguay; J C Tardif; J Rodés; M Joyal; J Crépeau; G Côté
Journal:  Cathet Cardiovasc Diagn       Date:  1997-06

8.  Rotational ablation of balloon angioplasty failures.

Authors:  J Rosenblum; S H Stertzer; R E Shaw; B Hidalgo; H N Hansell; M C Murphy; R K Myler
Journal:  J Invasive Cardiol       Date:  1992 Jul-Aug       Impact factor: 2.022

9.  A noncompliant, high pressure balloon to manage undilatable coronary lesions.

Authors:  Yogesh Raja; Helen C Routledge; Sagar N Doshi
Journal:  Catheter Cardiovasc Interv       Date:  2010-06-01       Impact factor: 2.692

10.  Intravascular ultrasound assessment of the novel AngioSculpt scoring balloon catheter for the treatment of complex coronary lesions.

Authors:  Alberto Fonseca; José de Ribamar Costa; Alexandre Abizaid; Fausto Feres; Andrea S Abizaid; Ricardo Costa; Rodolfo Staico; Luis Alberto Mattos; Amanda G R M Sousa; Eberhard Grube; J Edouardo Sousa
Journal:  J Invasive Cardiol       Date:  2008-01       Impact factor: 2.022

View more
  6 in total

1.  Triple-Guidewire Technique for Treating Stent Underexpansion in Severely Calcified Coronary Artery Lesions.

Authors:  Maheedhar Gedela; Shenjing Li; Chirag Desai; Tomasz Stys; Adam Stys
Journal:  Tex Heart Inst J       Date:  2020-04-01

2.  Bail-out therapy in ST-segment elevation myocardial infarction due to calcified lesion causing stent underexpansion: Intravascular lithotripsy is in the lead.

Authors:  Thomas Baudinet; Benjamin Seguy; Laura Cetran; Muhammad Khaled Luttoo; Pierre Coste; Edouard Gerbaud
Journal:  J Cardiol Cases       Date:  2021-01-09

Review 3.  How Do We Treat Complex Calcified Coronary Artery Disease?

Authors:  Paul N Fiorilli; Saif Anwaruddin
Journal:  Curr Treat Options Cardiovasc Med       Date:  2016-12

Review 4.  Current understanding of coronary artery calcification.

Authors:  Wei Liu; Yue Zhang; Cheuk-Man Yu; Qing-Wei Ji; Meng Cai; Ying-Xin Zhao; Yu-Jie Zhou
Journal:  J Geriatr Cardiol       Date:  2015-11       Impact factor: 3.327

5.  The Clinical and Angiographic Outcomes of Postdilation after Percutaneous Coronary Intervention in Patients with Acute Coronary Syndrome: A Systematic Review and Meta-Analysis.

Authors:  Yan Li; Xiying Liang; Wenjiao Zhang; Xuan Qiao; Zhilu Wang
Journal:  J Interv Cardiol       Date:  2021-04-09       Impact factor: 2.279

6.  Long-term outcomes of rotational atherectomy of underexpanded stents. A single center experience.

Authors:  Marco Hernández-Enríquez; Francisco Campelo-Parada; Thibault Lhermusier; Frédéric Bouisset; Jérôme Roncalli; Meyer Elbaz; Didier Carrié; Nicolas Boudou
Journal:  J Interv Cardiol       Date:  2018-01-25       Impact factor: 2.279

  6 in total