| Literature DB >> 23105618 |
P K Nigam1, S Sharma, R Sareen, V Paul, A Lal.
Abstract
This study was carried out to check the precision of HbA(1c) values on Bio-Rad Variant II in cases of a rare hemoglobin variant Q India. The study was carried out over a three month period on samples collected for HbA(1c) estimation. Seven out of eleven patients showed variable results of HbA(1c) with a very high and unacceptable intraday mean coefficient of variation (CV) of 9.93%. We conclude, that the results of HbA(1c) on Variant II can not be reported without adversely affecting HbA(1c) as a marker of long-term glycemic control in patients who have hemoglobin Q India. The HbA(1c) value of these patients needs to be assessed by a different instrument/method or the glycemic control be monitored by an alternate test like serum Fructosamine.Entities:
Keywords: HbA1c; diabetes mellitus; precision; variant hemoglobin
Year: 2006 PMID: 23105618 PMCID: PMC3453997 DOI: 10.1007/BF02912916
Source DB: PubMed Journal: Indian J Clin Biochem ISSN: 0970-1915