Reducing subsidence risk by using rapid manufactured patient-specific intervertebral disc implants.

Neal de Beer1, Cornie Scheffer.   

Abstract

BACKGROUND CONTEXT: Intervertebral disc implant size, shape, and position during total disc replacement have been shown to affect the risk of implant subsidence or vertebral fracture. Rapid manufacturing has been successfully applied to produce patient-specific implants for craniomaxillofacial, dental, hip, and knee requirements, but very little has been published on its application for spinal implants.
PURPOSE: This research was undertaken to investigate the improved load distribution and stiffness that can be achieved when using implants with matching bone interface geometry as opposed to implants with flat end plate geometries. STUDY
DESIGN: The study design comprises a biomechanical investigation and comparison of compressive loads applied to cadaveric vertebrae when using two different end plate designs.
METHODS: Four spines from male cadavers (ages 45-65 years, average 52 years), which had a total of n=88 vertebrae (C3-L5), were considered during this study. Bone mineral density scans on each spine revealed only one to be eligible for this study. Twenty remaining vertebrae (C3-L3) were potted and subjected to nondestructive compression tests followed by destructive compression tests. Custom-made nonfunctional implants were designed for this experiment. Ten implants were designed with matching end plate-to-bone interface geometry, whereas the other 10 were designed with flat end plates. Testing did not incorporate the use of a keel in either design type. I-Scan pressure sensors (Tekscan, Inc., MA, USA) were used during the nondestructive tests to assess the load distribution and percentage surface contact.
RESULTS: Average percent contact area measured during nondestructive tests was 45.27% and 10.49% for conformal and flat implants, respectively-a difference that is statistically significant (p<.001). A higher percent contact area was especially observed for cervical vertebrae because of their pronounced end plate concavity. During destructive compression tests, conformal implants achieved higher failure loads than flat implants. Conformal implants also performed significantly better when stiffness values were compared (p<.0001).
CONCLUSIONS: One of the main expected benefits from customizing the end plate geometry of disc implants is the reduced risk and potential for subsidence into the vertebral bone end plate. Subsidence depends in part on the stiffness of the implant-bone construct, and with a 137% increase in stiffness, the results of this study show that there are indeed significant potential benefits that can be achieved through the use of customization during the design and manufacture of intervertebral disc implants.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23103407     DOI: 10.1016/j.spinee.2012.10.003

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  5 in total

1.  Morphometry evaluations of cervical osseous endplates based on three dimensional reconstructions.

Authors:  Hang Feng; Haoxi Li; Zhaoyu Ba; Zhaoxiong Chen; Xinhua Li; Desheng Wu
Journal:  Int Orthop       Date:  2018-08-09       Impact factor: 3.075

2.  Parathyroid hormone enhances gap healing and osseointegration in orthopedic porous coated titanium implants: a correlative micro-computed tomographic, histomorphometric and biomechanical analysis.

Authors:  Xinlin Gao; Yang Meng; Dingjun Hao; Hao Liu
Journal:  BMC Musculoskelet Disord       Date:  2022-01-03       Impact factor: 2.362

3.  A morphometric study of the middle and lower cervical vertebral endplates and their components.

Authors:  Hang Feng; Xiang-Yi Fang; Da-Geng Huang; Cheng-Cheng Yu; Hou-Kun Li; Song-Chuan Zhao; Chao-Yuan Ge; Ru-Hai Bai; Ding-Jun Hao
Journal:  Medicine (Baltimore)       Date:  2017-03       Impact factor: 1.889

4.  Comparison of Two Surgical Techniques for the Treatment of Canine Disc Associated-Cervical Spondylomyelopathy.

Authors:  Cristian Falzone; Vito Tranquillo; Nicola Gasparinetti
Journal:  Front Vet Sci       Date:  2022-06-20

5.  Biomechanical evaluation of cervical disc replacement with a novel prosthesis based on the physiological curvature of endplate.

Authors:  Jigang Lou; Yuanchao Li; Beiyu Wang; Yang Meng; Quan Gong; Hao Liu
Journal:  J Orthop Surg Res       Date:  2018-02-27       Impact factor: 2.359

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.