Evaluation of 4 corneal astigmatic marking methods.

Nina Popp1, Nino Hirnschall, Sophie Maedel, Oliver Findl.   

Abstract

PURPOSE: To compare 4 devices used to mark the cornea before astigmatism-reducing surgery.
SETTING: Hanusch Krankenhaus, Vienna, Austria.
DESIGN: Randomized examiner-masked clinical trial.
METHODS: Patients were randomly allocated to 1 of 4 groups for preoperative corneal marking in the sitting position. The 4 methods used were marking at the slitlamp with an insulin needle, a pendular marker, a bubble marker, and a tonometer marker. The marks were then documented with a standardized photographic technique, and the rotational deviation and vertical misalignment were assessed.
RESULTS: The study enrolled 60 patients. The pendular-marking device showed the least rotational deviation to the reference meridian (mean 1.8 degrees). There was no statistically significant difference between slitlamp marking and pendular marking (P = .05); however, there was a significant difference between the pendular marker and the bubble marker and between the pendular marker and the tonometer marker (P = .01 and P < .01, respectively). The least vertical misalignment was observed with the slitlamp-marking device (mean 0.28 mm). There was no statistically significant difference in vertical misalignment between the 4 groups.
CONCLUSIONS: All marking devices showed a slight deviation to the horizontal reference meridian. Because small deviations of the meridian can result in a relevant reduction in the astigmatism-reducing effect with toric intraocular lenses, accurate marking of the cornea before surgery is critical due to the variable cyclotorsion caused by a change from the upright to the supine position. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2012 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 23098629     DOI: 10.1016/j.jcrs.2012.07.039

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  24 in total

1.  An easy and practical method for toric intraocular lens implantation: marking corneal astigmatic axis at slit-lamp.

Authors:  Huseyin Bayramlar; Yaşar Dag; Remzi Karadag; Ozgur Cakici
Journal:  Int Ophthalmol       Date:  2016-05-12       Impact factor: 2.031

2.  Comparison of two different methods of preoperative marking for toric intraocular lens implantation: bubble marker versus pendulum marker.

Authors:  Javed Hussain Farooqui; Archana Koul; Ranjan Dutta; Noshir Minoo Shroff
Journal:  Int J Ophthalmol       Date:  2016-05-18       Impact factor: 1.779

3.  Comparison of clinical outcomes among 3 marking methods for toric intraocular lens implantation.

Authors:  Hideyuki Onishi; Hidemasa Torii; Kazuhiro Watanabe; Kazuo Tsubota; Kazuno Negishi
Journal:  Jpn J Ophthalmol       Date:  2016-03-09       Impact factor: 2.447

4.  Toric IOL positioning with a no-touch head-up display axis alignment.

Authors:  J Luebke; D Boehringer; P Maier; T Reinhard; P Eberwein
Journal:  Int Ophthalmol       Date:  2019-11-23       Impact factor: 2.031

5.  Introduction of a Toric Intraocular Lens to a Non-Refractive Cataract Practice: Challenges and Outcomes.

Authors:  Clare Kirwan; John M Nolan; Jim Stack; Ian Dooley; Johnny Moore; Tara Cb Moore; Stephen Beatty
Journal:  Int J Ophthalmol Clin Res       Date:  2016-05-25

6.  Factors Influencing Efficacy of Peripheral Corneal Relaxing Incisions during Cataract Surgery.

Authors:  Nino Hirnschall; Jörg Wiesinger; Petra Draschl; Oliver Findl
Journal:  J Ophthalmol       Date:  2015-06-25       Impact factor: 1.909

7.  Evaluation of a new electronic preoperative reference marker for toric intraocular lens implantation by two different methods of analysis: Adobe Photoshop versus iTrace.

Authors:  Javed Hussain Farooqui; Mansi Sharma; Archana Koul; Ranjan Dutta; Noshir Minoo Shroff
Journal:  Oman J Ophthalmol       Date:  2017 May-Aug

8.  A comparison of three different corneal marking methods used to determine cyclotorsion in the horizontal meridian.

Authors:  Hung-Yuan Lin; Yi-Ting Fang; Ya-Jung Chuang; Justin N Karlin; Hsin-Yang Chen; Szu-Yuan Lin; Pi-Jung Lin; Ming Chen
Journal:  Clin Ophthalmol       Date:  2017-02-08

9.  Comparison of Preoperative Measurements with Intraoperative Aberrometry in Predicting Need for Correction in Eyes with Low Astigmatism Undergoing Cataract Surgery.

Authors:  John A Hovanesian
Journal:  Clin Ophthalmol       Date:  2021-05-26

10.  Comparison Between Digital and Manual Marking for Toric Intraocular Lenses: A Randomized Trial.

Authors:  Abdel Hamid Elhofi; Hany Ahmed Helaly
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.