| Literature DB >> 23082296 |
Neil T Clancy1, Danail Stoyanov, David R C James, Aimee Di Marco, Vincent Sauvage, James Clark, Guang-Zhong Yang, Daniel S Elson.
Abstract
Sequential multispectral imaging is an acquisition technique that involves collecting images of a target at different wavelengths, to compile a spectrum for each pixel. In surgical applications it suffers from low illumination levels and motion artefacts. A three-channel rigid endoscope system has been developed that allows simultaneous recording of stereoscopic and multispectral images. Salient features on the tissue surface may be tracked during the acquisition in the stereo cameras and, using multiple camera triangulation techniques, this information used to align the multispectral images automatically even though the tissue or camera is moving. This paper describes a detailed validation of the set-up in a controlled experiment before presenting the first in vivo use of the device in a porcine minimally invasive surgical procedure. Multispectral images of the large bowel were acquired and used to extract the relative concentration of haemoglobin in the tissue despite motion due to breathing during the acquisition. Using the stereoscopic information it was also possible to overlay the multispectral information on the reconstructed 3D surface. This experiment demonstrates the ability of this system for measuring blood perfusion changes in the tissue during surgery and its potential use as a platform for other sequential imaging modalities.Entities:
Keywords: (170.2150) Endoscopic imaging; (170.3010) Image reconstruction techniques; (170.6510) Spectroscopy, tissue diagnostics
Year: 2012 PMID: 23082296 PMCID: PMC3469985 DOI: 10.1364/BOE.3.002567
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732
Fig. 1(a) Optical set-up of the system [22]. (b) Photograph of the experimental arrangement.
Fig. 2Illustration of trinocular endoscope imaging geometry. Geometric calibration of the system means that image points in the left and right white light images can be used to triangulate the 3D position of points on the tissue surface. These can then be reprojected into multispectral image coordinates. We track the motion of points in the white light images as these have consistent light appearance and we use the reprojection capability of the calibrated system to maintain a track of the respective region in the multispectral image.
Fig. 3Image processing algorithm schematic. The raw image stacks are acquired simultaneously using the synchronised endoscope cameras. Features are identified and tracked in 3D throughout the image stack. The transformations needed to align these features are derived and implemented. Each feature is then projected through 3D space onto the multispectral camera using the calibration relations, so that they are automatically aligned in the multispectral 2D images. These can then be processed to extract the reflectance spectrum at each pixel location over the stack and compute relative concentrations of chromophores of interest. The processed image may then be reprojected onto 3D space to aid visualisation.
Fig. 4(a) Reconstructed colour image of checker card using misaligned multispectral images. (b) Colour checker reconstructed using multispectral images of stationary target. (c) Image of colour checker card acquired using colour CCD camera. (d) Colour image of moving chart reconstructed using the aligned multispectral camera images along with reflectance spectra (normalised intensity vs. wavelength) for each panel. The reflectance spectrum for each colour panel calculated using the aligned images (blue dots) is compared with that calculated from images of a stationary target (red dots).
Bland-Altman analysis of the agreement between spectra measured using multispectral images of a static colour checker card, and images of a moving card aligned using the 3D reconstruction and tracking algorithm
| 0.0218 | 0.0609 | 0.0018 | 0.0489 | 0.0066 | 0.1437 | 0.0387 | 0.0692 | 0.0125 | |
| 0.1552 | 0.1334 | 0.0962 | 0.1184 | 0.0856 | 0.1225 | 0.1102 | 0.0847 | 0.0438 | |
| 0.3042 | 0.2614 | 0.1885 | 0.2321 | 0.1677 | 0.2400 | 0.2160 | 0.1661 | 0.0859 | |
The average absolute difference between the normalised reflected intensity across all wavelengths is presented, along with the standard deviation of the differences (σ) and the limits of agreement (L.O.A.), which are defined as ±1.96 × σ and represent the range in which 95% of the differences are found.
Fig. 5Motion tracking in vivo. The location of the patch is tracked in the colour stereo cameras (right camera; top row). The feature is back-projected onto the multispectral camera (middle row). Multispectral images are aligned using the back-projected feature (bottom row). See Media 1, Media 2, and Media 3.
Fig. 6Colour images of selected patches of intestinal tissue reconstructed from the multispectral stack of images, and corresponding maps of total haemoglobin. Arrows indicate smaller blood vessels that are washed out in the raw images but become visible after alignment. The smear effect on Patch 4 and 5 is due to pixel padding after the region of interest around those features moved outside the boundary of the raw multispectral image.
Fig. 7Stereo pair (left and right) used for feature tracking and 3D reconstruction shown alongside the finished surface. The processed Hbt data from the patches analysed in Fig. 6 are overlaid onto the surface.