| Literature DB >> 23077121 |
Elizabeth L Palavecino1, Jacqueline M Burnell.
Abstract
This study correlated the daptomycin MIC results obtained by Microscan and by Etest in Staphylococcus aureus and enterococci and found that the Microscan panel GP 29 had a high rate of false nonsusceptible results. Of the isolates interpreted as nonsusceptible by Microscan, 87% of Staphylococcus aureus, 90% of Enterococcus faecalis, and 88% of Enterococcus faecium isolates were interpreted as susceptible by Etest. In turn, Etest also has false nonsusceptible results compared to reference methods.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23077121 PMCID: PMC3536226 DOI: 10.1128/JCM.01721-12
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948