Identification and quantification of hemoglobins in whole blood: the analytical and organizational aspects of Capillarys 2 Flex Piercing compared with agarose electrophoresis and HPLC methods.

Sara Altinier1, Mariacristina Varagnolo, Martina Zaninotto, Mario Plebani.   

Abstract

BACKGROUND: The present study was conducted to evaluate the analytical performance and the organizational aspects of Capillarys 2 Flex Piercing system (CFP) respect to agarose electrophoresis and HPLC methods in hemoglobinopathies screening.
METHODS: The measurement of imprecision in HbA 2 and HbF quantification was verified on HbA 2 CFP control and on three samples; 74 whole blood samples were used to evaluate migration time imprecision of hemoglobin variants S, C and E (HbS, HbC, and HbE); to compare methods, 451 samples were tested on CFP and HPLC; reference values were verified as value distribution in 160 blood donors and at ROC curve analysis on 449 samples from routine analysis.
RESULTS: Imprecision: the analytical CV % s ranged from 1.25 to 3.9 at HbA 2 quantification, the CV % was 3.78 at HbF quantification; the running time imprecision for HbS and HbC and HbE ranged from 0.20 to 0.69 % . Method comparison: at regression analysis findings were HbA 2: CFP=1.21×HPLC–0.64, HbF: CFP=1.31×HPLC−0.75, HbS: CFP=1.10×HPLC−3.24. Reference values: the HbA 2 95th percentile range was 2.5–2.8; HbF was undetectable in 154 out 160 samples tested; at ROC curve analysis the best combination of sensitivity and diagnostic efficiency was obtained using 2.2 and 3.0, as reference values, for HbA 2 and 1.1 as the upper reference limit for HbF. Organizational aspects: with respect to the procedures currently implemented in our laboratory CFP requires 2 h less time and obviates the need for some manual steps.
CONCLUSIONS: The quantification, reproducibility and diagnostic efficiency provided by CFP in identification and quantification of hemoglobins appear accurate. In addition, the use of primary tubes allows improved safety, and the avoidance of some manual steps, that prolong working time and are a source of possible errors.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23072852     DOI: 10.1515/cclm-2012-0061

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  2 in total

1.  Unusual pattern in haemoglobin electrophoresis in Croatian population: a case report.

Authors:  Dragana Segulja; Danica Matisic; Lorena Honovic; Josip Batinic; Dunja Rogic
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

2.  Molecular prevalence of HBB-associated hemoglobinopathy among reproductive-age adults and the prenatal diagnosis in Jiangxi Province, southern central China.

Authors:  Haiyan Luo; Ting Huang; Qing Lu; Liuyang Zhang; Yonghua Xu; Yan Yang; Zhen Guo; Huizhen Yuan; Yinqin Shen; Shuhui Huang; Bicheng Yang; Yongyi Zou; Yanqiu Liu
Journal:  Front Genet       Date:  2022-09-28       Impact factor: 4.772

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.