Reuse of explanted, resterilized implantable cardioverter-defibrillators: a cohort study.

Behzad B Pavri1, Yash Lokhandwala, Gaurav V Kulkarni, Mandar Shah, Bharat K Kantharia, Daniel A N Mascarenhas.   

Abstract

BACKGROUND: Implantable cardioverter-defibrillators (ICDs) often have clinically useful battery life remaining when explanted because of upgrades, infection, or patient death.
OBJECTIVE: To show that explanted ICDs can be resterilized and reused.
DESIGN: Retrospective cohort study.
SETTING: Multicenter ICD acquisition and single-center ICD reimplantation. PATIENTS: Indigent persons in India who had class I indications for cardiac resynchronization therapy with an ICD and were unable to afford such a device. MEASUREMENTS: Device longevity after reimplantation, device-related complications, number of appropriate therapies, patient clinical characteristics, and deaths.
RESULTS: Eighty-one consecutive consenting patients (mean age, 52.6 years; 66 male patients) received 106 explanted devices. Twenty-two patients received a second device and 3 patients received a third device after the prior one reached replacement voltage. Mean time to ICD replacement was 1287.4 days. Follow-up data were available for 75 of 81 (92.6%) patients. Mean follow-up duration for all devices was 824.9 days. No infectious complications occurred; 1 lead dislodgement and 1 lead fracture required repeated surgery. Appropriate therapy (shocks or antitachycardia pacing) was delivered by 64 of 106 (60.4%) devices in 44 of 81 (54.3%) patients. Nine of 81 (11.1%) patients died; mean time from implantation to death was 771.3 days. LIMITATIONS: This is a retrospective report of a single-center experience with a modest number of patients and devices. Follow-up data were missing for 6 patients. No records were kept of the number of devices obtained through postmortem versus antemortem explantation or whether explantation was due to infection or upgrade. Complete data were not available on exact battery voltage at the time of reimplantation, left ventricular ejection fraction, or number of inappropriate shocks. A control group was not possible.
CONCLUSION: Explanted ICDs with 3 or more years of estimated remaining battery life can be reused after they are cleaned and resterilized. These devices functioned normally and delivered life-saving therapies, without an increased risk for complications. These preliminary data deserve further validation and, if confirmed, could have important societal and economic implications. PRIMARY FUNDING SOURCE: None.

Entities:  

Mesh:

Year:  2012        PMID: 23070487     DOI: 10.7326/0003-4819-157-8-201210160-00004

Source DB:  PubMed          Journal:  Ann Intern Med        ISSN: 0003-4819            Impact factor:   25.391


  10 in total

1.  Necessity of databases in context of Indian guidelines.

Authors:  Sridevi Chennapragada
Journal:  Indian Heart J       Date:  2015-10-21

Review 2.  CSI position statement on management of heart failure in India.

Authors:  Santanu Guha; S Harikrishnan; Saumitra Ray; Rishi Sethi; S Ramakrishnan; Suvro Banerjee; V K Bahl; K C Goswami; Amal Kumar Banerjee; S Shanmugasundaram; P G Kerkar; Sandeep Seth; Rakesh Yadav; Aditya Kapoor; Ajaykumar U Mahajan; P P Mohanan; Sundeep Mishra; P K Deb; C Narasimhan; A K Pancholia; Ajay Sinha; Akshyaya Pradhan; R Alagesan; Ambuj Roy; Amit Vora; Anita Saxena; Arup Dasbiswas; B C Srinivas; B P Chattopadhyay; B P Singh; J Balachandar; K R Balakrishnan; Brian Pinto; C N Manjunath; Charan P Lanjewar; Dharmendra Jain; Dipak Sarma; G Justin Paul; Geevar A Zachariah; H K Chopra; I B Vijayalakshmi; J A Tharakan; J J Dalal; J P S Sawhney; Jayanta Saha; Johann Christopher; K K Talwar; K Sarat Chandra; K Venugopal; Kajal Ganguly; M S Hiremath; Milind Hot; Mrinal Kanti Das; Neil Bardolui; Niteen V Deshpande; O P Yadava; Prashant Bhardwaj; Pravesh Vishwakarma; Rajeeve Kumar Rajput; Rakesh Gupta; S Somasundaram; S N Routray; S S Iyengar; G Sanjay; Satyendra Tewari; Sengottuvelu G; Soumitra Kumar; Soura Mookerjee; Tiny Nair; Trinath Mishra; U C Samal; U Kaul; V K Chopra; V S Narain; Vimal Raj; Yash Lokhandwala
Journal:  Indian Heart J       Date:  2018-06-08

3.  Device therapy: Reuse of ICDs in developing countries--'proof of concept' data published.

Authors:  Alexandra King
Journal:  Nat Rev Cardiol       Date:  2012-11-13       Impact factor: 32.419

Review 4.  Relevance of guideline-based ICD indications to clinical practice.

Authors:  Nora Al-Jefairi; Haran Burri
Journal:  Indian Heart J       Date:  2013-12-22

5.  Reuse of pacemakers, defibrillators and cardiac resynchronisation devices.

Authors:  Raja J Selvaraj; R Sakthivel; Santhosh Satheesh; Ajith Ananthakrishna Pillai; Pascal Sagnol; Xavier Jouven; Bernard Dodinot; Jayaraman Balachander
Journal:  Heart Asia       Date:  2017-01-23

6.  Safety and feasibility of biventricular devices reuse in general and elderly population--a single-center retrospective cohort study.

Authors:  Raluca Şoşdean; Cristian Mornoş; Bogdan Enache; Răzvan I Macarie; Raluca Ianoş; Ana-Maria Ştefea; Sorin Pescariu
Journal:  Clin Interv Aging       Date:  2015-08-12       Impact factor: 4.458

7.  Performance of re-used pacemakers and implantable cardioverter defibrillators compared with new devices at Groote Schuur Hospital in Cape Town, South Africa.

Authors:  Zimasa V Jama; Ashley Chin; Motasim Badri; Bongani M Mayosi
Journal:  Cardiovasc J Afr       Date:  2015 Jul-Aug       Impact factor: 1.167

Review 8.  Channelopathies - emerging trends in the management of inherited arrhythmias.

Authors:  Priya Chockalingam; Yuka Mizusawa; Arthur Am Wilde
Journal:  Indian Pacing Electrophysiol J       Date:  2015-04-01

Review 9.  Guidance on reuse of cardio-vascular catheters and devices in India: A consensus document.

Authors:  Aditya Kapoor; Amit Vora; Gita Nataraj; Sundeep Mishra; Prafulla Kerkar; C N Manjunath
Journal:  Indian Heart J       Date:  2017-04-13

10.  Patient preferences regarding device reuse and potential of devices for reuse - a study in a veteran population.

Authors:  Indiresha R Iyer; Judith Mackall
Journal:  Indian Pacing Electrophysiol J       Date:  2013-06-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.