Assessment of radiofrequency device sensitivity for the detection of retained surgical sponges in patients with morbid obesity.

Victoria M Steelman1, Mohammad H Alasagheirin.   

Abstract

HYPOTHESIS: Retained surgical sponges are serious medical errors that result in negative patient outcomes. A radiofrequency (RF) mat for the operating room bed has recently been introduced to detect the presence of a retained surgical sponge. The study objectives were to evaluate the sensitivity and specificity of the RF mat for the detection of surgical sponges through the torso of individuals with varying body habitus and to compare the sensitivity of the RF mat with that of the RF wand. We hypothesized that the sensitivity and specificity of the RF technology would be comparable to published findings of the manual sponge count by operating room personnel.
DESIGN: A prospective, crossover, double-blinded study design was used. Participants served as their own controls. SETTING Large Midwestern academic medical center. PARTICIPANTS: In total, the first phase of the study enrolled 203 participants, including 129 (63.5%) with morbid obesity. One hundred seventeen of 203 participants were also enrolled in the second phase of the study. MAIN OUTCOME MEASURES: The study participants reclined in a supine position on top of an RF mat. Four surgical sponges were sequentially placed on top of the torso in locations approximating the abdominal quadrants. The torso was scanned for sponges. In a subset of participants, 4 surgical sponges were sequentially placed underneath the torso, and an RF wand was passed over the abdomen.
RESULTS: Overall, 812 readings were obtained with the RF mat, and 468 readings were obtained with the RF wand. Twelve false-negative readings were obtained with the RF mat, exclusively in participants with super morbid obesity (body mass index [calculated as weight in kilograms divided by height in meters squared] >50.0). Overall, the sensitivity of the RF mat was 98.1%, and the specificity of the RF mat was 100.0%. In the subset of 117 participants in whom the RF wand was also used, the sensitivity and specificity of the wand were each 100.0%.
CONCLUSIONS: The sensitivity and specificity of RF device technology are much higher than those of surgical sponge counts or published findings on the use of intraoperative radiographs to identify retained surgical sponges. The RF wand is more sensitive than the RF mat in individuals with morbid obesity.

Entities:  

Mesh:

Year:  2012        PMID: 23070411     DOI: 10.1001/archsurg.2012.1556

Source DB:  PubMed          Journal:  Arch Surg        ISSN: 0004-0010


  8 in total

1.  Occult risk of broken instruments for endoscopy-assisted surgery.

Authors:  Hiroshi Yasuhara; Kazuhiko Fukatsu; Takami Komatsu; Satoshi Murakoshi; Yuhei Saito; Yushi Uetera
Journal:  World J Surg       Date:  2014-11       Impact factor: 3.352

2.  Retained surgical sponges, needles and instruments.

Authors:  V M Steelman
Journal:  Ann R Coll Surg Engl       Date:  2014-03       Impact factor: 1.951

3.  Gossypibomas in India - A systematic literature review.

Authors:  T Patial; V Thakur; N Vijhay Ganesun; M Sharma
Journal:  J Postgrad Med       Date:  2017 Jan-Mar       Impact factor: 1.476

Review 4.  Exploring Risk, Antecedents and Human Costs of Living with a Retained Surgical Item: A Narrative Synthesis of Australian Case Law 1981-2018.

Authors:  Sonya R Osborne; Tina Cockburn; Juliet Davis
Journal:  J Multidiscip Healthc       Date:  2021-08-31

Review 5.  Risk factors and preventive strategies for unintentionally retained surgical sharps: a systematic review.

Authors:  Samuel Weprin; Fabio Crocerossa; Dielle Meyer; Kaitlyn Maddra; David Valancy; Reginald Osardu; Hae Sung Kang; Robert H Moore; Umberto Carbonara; Fernando J Kim; Riccardo Autorino
Journal:  Patient Saf Surg       Date:  2021-07-12

6.  Intraluminal migration of retained surgical sponge as a cause of intestinal obstruction.

Authors:  Basmah A Rafie; Omar J AbuHamdan; Nawal S Trengganu
Journal:  J Surg Case Rep       Date:  2013-05-09

7.  Gossypiboma revisited: A never ending issue.

Authors:  M Ezzedien Rabie; Mohammad Hassan Hosni; Alaa Al Safty; Manea Al Jarallah; Fadel Hussain Ghaleb
Journal:  Int J Surg Case Rep       Date:  2015-12-23

8.  Evaluation of radiofrequency electronic system in intraoperative monitoring of surgical textiles.

Authors:  Adriana Marco Antonio; Carlos Andre Pereira Vieira
Journal:  Einstein (Sao Paulo)       Date:  2018-04-26
  8 in total