| Literature DB >> 29694622 |
Adriana Marco Antonio1, Carlos Andre Pereira Vieira1.
Abstract
Objective To test performance of SurgiSafe®, a radiofrequency electronic device to detect surgical textiles during operations as compared to manual counting. Methods Surgical sponges with radiofrequency TAGs were placed in the abdominal cavity of a pig submitted to laparotomy, in randomly distributed sites. The TAGs were counted manually and also using SurgiSafe®. Positive and negative predictive values, sensitivity, specificity and time required for counting were analyzed for both methods. Results Through the analysis of 35 surgical cycles, SurgiSafe® immediately identified all sponges, with specificity, sensitivity, positive and negative predictive values of 100%. Although not statistically significant, the manual count had sensitivity of 99.72% and specificity of 99.90%. Conclusion SurgiSafe® proved to be an effective device to identify surgical sponges in vivo, in real time; and its use as an adjuvant to manual counting is very helpful to increase patient's safety.Entities:
Mesh:
Year: 2018 PMID: 29694622 PMCID: PMC6066152 DOI: 10.1590/s1679-45082018ao3997
Source DB: PubMed Journal: Einstein (Sao Paulo) ISSN: 1679-4508
Figure 1SurgiSafe® apparatus. (A) SurgiSafe® image, (B) tagged surgical sponges, (C) operating room, and (D) manual count of surgical textile items
Comparing results of SurgiSafe® count (immediate counting) and manual count
| Cycle number | Total number of gauzes | Total SurgiSafe® on table | Total manual count | Manual count time |
|---|---|---|---|---|
| 1 | 23 | 23 | 23 | 44 |
| 2 | 26 | 26 | 26 | 52 |
| 3 | 30 | 30 | 30 | 72 |
| 4 | 22 | 22 | 22 | 51 |
| 5 | 31 | 31 | 31 | 90 |
| 6 | 33 | 33 | 33 | 109 |
| 7 | 34 | 34 | 33 | 68 |
| 8 | 33 | 33 | 33 | 101 |
| 9 | 38 | 38 | 38 | 91 |
| 10 | 36 | 36 | 36 | 100 |
| 11 | 30 | 30 | 29 | 60 |
| 12 | 34 | 34 | 33 | 91 |
| 13 | 31 | 31 | 31 | 66 |
| 14 | 49 | 49 | 49 | 134 |
| 15 | 26 | 26 | 26 | 93 |
| 16 | 32 | 32 | 32 | 84 |
| 17 | 42 | 42 | 42 | 84 |
| 18 | 23 | 23 | 23 | 43 |
| 19 | 18 | 18 | 18 | 62 |
| 20 | 15 | 15 | 15 | 29 |
| 21 | 35 | 35 | 35 | 85 |
| 22 | 34 | 34 | 34 | 87 |
| 23 | 35 | 35 | 35 | 75 |
| 24 | 35 | 35 | 35 | 91 |
| 25 | 33 | 33 | 33 | 71 |
| 26 | 24 | 24 | 25 | 60 |
| 27 | 31 | 31 | 31 | 95 |
| 28 | 23 | 23 | 23 | 101 |
| 29 | 34 | 34 | 34 | 245 |
| 30 | 32 | 32 | 32 | 104 |
| 31 | 35 | 35 | 35 | 134 |
| 32 | 32 | 32 | 32 | 85 |
| 33 | 42 | 42 | 42 | 123 |
| 34 | 8 | 8 | 8 | 14 |
| 35 | 31 | 31 | 31 | 79 |
Figure 2Time of manual count per gauze unit versus count cycles in chronological order