Suitability of Exoseal vascular closure device for antegrade femoral artery puncture site closure.

Christopher Schmelter1, Andrea Liebl, Nektarios Poullos, Volker Ruppert, Dierk Vorwerk.   

Abstract

PURPOSE: To assess the efficacy and safety of the Exoseal vascular closure device for antegrade puncture of the femoral artery.
METHODS: In a prospective study from February 2011 to January 2012, a total of 93 consecutive patients received a total of 100 interventional procedures via an antegrade puncture of the femoral artery. An Exoseal vascular closure device (6F) was used for closure in all cases. Puncture technique, duration of manual compression, and use of compression bandages were documented. All patients were monitored by vascular ultrasound and color-coded duplex sonography of their respective femoral artery puncture site within 12 to 36 h after angiography to check for vascular complications.
RESULTS: In 100 antegrade interventional procedures, the Exoseal vascular closure device was applied successfully for closure of the femoral artery puncture site in 96 cases (96 of 100, 96.0 %). The vascular closure device could not be deployed in one case as a result of kinking of the vascular sheath introducer and in three cases because the bioabsorbable plug was not properly delivered to the extravascular space adjacent to the arterial puncture site, but instead fully removed with the delivery system (4.0 %). Twelve to 36 h after the procedure, vascular ultrasound revealed no complications at the femoral artery puncture site in 93 cases (93.0 %). Minor vascular complications were found in seven cases (7.0 %), with four cases (4.0 %) of pseudoaneurysm and three cases (3.0 %) of significant late bleeding, none of which required surgery.
CONCLUSION: The Exoseal vascular closure device was safely used for antegrade puncture of the femoral artery, with a high rate of procedural success (96.0 %), a low rate of minor vascular complications (7.0 %), and no major adverse events.

Entities:  

Mesh:

Year:  2012        PMID: 23070109     DOI: 10.1007/s00270-012-0501-2

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  8 in total

Review 1.  Access and hemostasis: femoral and popliteal approaches and closure devices-why, what, when, and how?

Authors:  Iacopo Barbetta; Jos C van den Berg
Journal:  Semin Intervent Radiol       Date:  2014-12       Impact factor: 1.513

2.  Haemostasis with the FISH Vascular Closure Device after 6 French Transfemoral Accesses in Interventional Radiology: Clinical Results.

Authors:  Peter C Kamusella; Christopher W Lüdtke; Fabian Scheer; Reimer Andresen; Christian Wissgott
Journal:  J Clin Diagn Res       Date:  2017-02-01

3.  Thermic sealing in femoral catheterization: First experience with the Secure Device.

Authors:  Michael Sacherer; Ewald Kolesnik; Friederike von Lewinski; Nicolas Verheyen; Karin Brandner; Markus Wallner; Deborah M Eaton; Olev Luha; Robert Zweiker; Dirk von Lewinski
Journal:  Cardiol J       Date:  2018-04-03       Impact factor: 2.737

4.  Comparison of Exo-Seal(®) and Angio-Seal (®) for arterial puncture site closure: A randomized, multicenter, single-blind trial.

Authors:  Johannes Ketterle; Harald Rittger; Inga Helmig; Lutz Klinghammer; Stefan Zimmermann; Wolfgang Hohenforst-Schmidt; Johannes Brachmann; Holger Nef; Stephan Achenbach; Christian Schlundt
Journal:  Herz       Date:  2015-06-13       Impact factor: 1.443

5.  [Arterial closure devices. What device for which clinical situation?].

Authors:  M Treitl; K M Eberhardt; D Maxien; B Behrends; M F Reiser
Journal:  Radiologe       Date:  2013-03       Impact factor: 0.635

6.  Exoseal for puncture site closure after antegrade procedures in peripheral arterial disease patients.

Authors:  Gerald Hackl; Thomas Gary; Klara Belaj; Franz Hafner; Peter Rief; Hannes Deutschmann; Marianne Brodmann
Journal:  Diagn Interv Radiol       Date:  2014 Sep-Oct       Impact factor: 2.630

7.  Percutaneous Vascular Closure System Based on an Extravascular, Bioabsorbable Polyglycolic Plug (ExoSeal): Results from 1000 Patients.

Authors:  P Kamusella; C Wissgott; T Jahnke; J Brossmann; F Scheer; C W Lüdtke; R Andresen
Journal:  Clin Med Insights Cardiol       Date:  2015-06-10

8.  Korean single-center experience with femoral access closure using the ExoSeal device.

Authors:  Yoonhee Han; Jae Hyun Kwon; Surin Park
Journal:  World J Radiol       Date:  2018-09-28
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.