Comparison of two different generations of "NIRS" devices and transducers in healthy volunteers and ICU patients.

Cecilia Luengo1, Matthieu Resche-Rigon, Charles Damoisel, Sébastien Kerever, Jacques Creteur, Didier Payen.   

Abstract

The purpose of this study is to compare Near Infrared Spectroscopy (NIRS) thenar eminence parameters obtained with 2 different devices from the same manufacturer (InSpectra Models 325 and 650, Hutchinson Tech, Min USA), and 2 different probes (15 vs. 25 mm spacing), in healthy volunteers (HV) and ICU patients. Prospective, observational study in ICU setting. Simultaneous, cross over NIRS inter-device comparison and comparison between different probes (25 vs. 15 mm spacing) were done at baseline and during vascular occlusion tests (VOTs). Forty patients (19 septic shock, 21 trauma), and 29 HV were included. NIRS inter-device comparison showed similar baseline StO(2) values in HV and patients. The VOT result were significantly different for minimal StO(2) value reached during VOT (StO(2min)) (intraclass concordance coefficient (ICC) = 0.18), the occlusion slope (ICC = 0.16) and the reperfusion slope (StO(2reperf)) (ICC = 0.26). The probe comparison was also significantly different for VOT parameters (StO(2min) (ICC = 0.43), occlusion (ICC = 0.50) and StO(2reperf) (ICC = 0.48). The low concordance, poor agreement and large bias (ICC and Bland & Altman) observed, were related both to the device used and the probe spacing. StO(2) data obtained with NIRS model 650 and 15 mm probe differ from values obtained with the previous device (325 and probe spacing 25 or 15 mm). This difference is not related to the population tested, but to the device and probe spacing. As a consequence, despite similar trends for variations between HV and patients during VOT, threshold and predictive values for outcome should be revisited with the new device before the acceptance for routine clinical use.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23054384     DOI: 10.1007/s10877-012-9400-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  34 in total

1.  Noninvasive method for measuring local hemoglobin oxygen saturation in tissue using wide gap second derivative near-infrared spectroscopy.

Authors:  Dean E Myers; LeAnn D Anderson; Roxanne P Seifert; Joseph P Ortner; Chris E Cooper; Greg J Beilman; John D Mowlem
Journal:  J Biomed Opt       Date:  2005 May-Jun       Impact factor: 3.170

2.  Correlates of thenar near-infrared spectroscopy-derived tissue O2 saturation after cardiac surgery.

Authors:  Renske J Uilkema; Arie Bastiaan Johan Groeneveld
Journal:  Interact Cardiovasc Thorac Surg       Date:  2007-02-06

3.  Measurement error and correlation coefficients.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-07-06

Review 4.  2001 SCCM/ESICM/ACCP/ATS/SIS International Sepsis Definitions Conference.

Authors:  Mitchell M Levy; Mitchell P Fink; John C Marshall; Edward Abraham; Derek Angus; Deborah Cook; Jonathan Cohen; Steven M Opal; Jean-Louis Vincent; Graham Ramsay
Journal:  Crit Care Med       Date:  2003-04       Impact factor: 7.598

5.  Statistical evaluation of agreement between two methods for measuring a quantitative variable.

Authors:  J Lee; D Koh; C N Ong
Journal:  Comput Biol Med       Date:  1989       Impact factor: 4.589

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Masseter tissue oxygen saturation predicts normal central venous oxygen saturation during early goal-directed therapy and predicts mortality in patients with severe sepsis.

Authors:  Gwenhael Colin; Olivier Nardi; Andrea Polito; Jérôme Aboab; Virginie Maxime; Bernard Clair; Diane Friedman; David Orlikowski; Tarek Sharshar; Djillali Annane
Journal:  Crit Care Med       Date:  2012-02       Impact factor: 7.598

8.  Changes in muscle tissue oxygenation during stagnant ischemia in septic patients.

Authors:  Roman Pareznik; Rajko Knezevic; Gorazd Voga; Matej Podbregar
Journal:  Intensive Care Med       Date:  2005-11-01       Impact factor: 17.440

9.  Characterization of tissue oxygen saturation and the vascular occlusion test: influence of measurement sites, probe sizes and deflation thresholds.

Authors:  Hernando Gómez; Jaume Mesquida; Peter Simon; Hyung Kook Kim; Juan C Puyana; Can Ince; Michael R Pinsky
Journal:  Crit Care       Date:  2009-11-30       Impact factor: 9.097

10.  Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers.

Authors:  Rick Bezemer; Alexandre Lima; Dean Myers; Eva Klijn; Michal Heger; Peter T Goedhart; Jan Bakker; Can Ince
Journal:  Crit Care       Date:  2009-11-30       Impact factor: 9.097

View more
  2 in total

1.  Multichannel wearable fNIRS-EEG system for long-term clinical monitoring.

Authors:  Ali Kassab; Jérôme Le Lan; Julie Tremblay; Phetsamone Vannasing; Mahya Dehbozorgi; Philippe Pouliot; Anne Gallagher; Frédéric Lesage; Mohamad Sawan; Dang Khoa Nguyen
Journal:  Hum Brain Mapp       Date:  2017-10-23       Impact factor: 5.038

2.  Predicting the Need for Renal Replacement Therapy Using a Vascular Occlusion Test and Tissue Oxygen Saturation in Patients in the Early Phase of Multiorgan Dysfunction Syndrome.

Authors:  Franz Haertel; Diana Reisberg; Martin Peters; Sebastian Nuding; P Christian Schulze; Karl Werdan; Henning Ebelt
Journal:  J Clin Med       Date:  2022-03-04       Impact factor: 4.241

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.