Literature DB >> 23047863

Study-parameter impact in quantitative 90-Yttrium PET imaging for radioembolization treatment monitoring and dosimetry.

Andreas Goedicke1, Yannick Berker, Frederik Anton Verburg, Florian Friedrich Behrendt, Oliver Winz, Felix Manuel Mottaghy.   

Abstract

A small positron-generating branch in 90-Yttrium ((90)Y) decay enables post-therapy dose assessment in liver cancer radioembolization treatment. The aim of this study was to validate clinical (90)Y positron emission tomography (PET) quantification, focusing on scanner linearity as well as acquisition and reconstruction parameter impact on scanner calibration. Data from three dedicated phantom studies (activity range: 55.2 MBq-2.1 GBq) carried out on a Philips Gemini TF 16 PET/CT scanner were analyzed after reconstruction with up to 361 parameter configurations. For activities above 200 MBq, scanner linearity could be confirmed with relative error margins 4%. An acquisition-time-normalized calibration factor of 1.04 MBq·s/CNTS was determined for the employed scanner. Stable activity convergence was found in hot phantom regions with relative differences in summed image intensities between -3.6% and +2.4%. Absolute differences in background noise artifacts between - 79.9% and + 350% were observed. Quantitative accuracy was dominated by subset size selection in the reconstruction. Using adequate segmentation and optimized acquisition parameters, the average activity recovery error induced by the axial scanner sensitivity profile was reduced to +2.4%±3.4% (mean ± standard deviation). We conclude that post-therapy dose assessment in (90)Y PET can be improved using adapted parameter setups.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23047863     DOI: 10.1109/TMI.2012.2221135

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  10 in total

Review 1.  Theranostic Imaging of Yttrium-90.

Authors:  Chadwick L Wright; Jun Zhang; Michael F Tweedle; Michael V Knopp; Nathan C Hall
Journal:  Biomed Res Int       Date:  2015-05-28       Impact factor: 3.411

2.  A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres : The QUEST Phantom Study.

Authors:  Kathy P Willowson; Michael Tapner; Dale L Bailey
Journal:  Eur J Nucl Med Mol Imaging       Date:  2015-05-13       Impact factor: 9.236

3.  Post-radioembolization yttrium-90 PET/CT - part 2: dose-response and tumor predictive dosimetry for resin microspheres.

Authors:  Yung-Hsiang Kao; Jeffrey D Steinberg; Young-Soon Tay; Gabriel Ky Lim; Jianhua Yan; David W Townsend; Charley A Budgeon; Jan A Boucek; Roslyn J Francis; Timothy St Cheo; Mark C Burgmans; Farah G Irani; Richard Hg Lo; Kiang-Hiong Tay; Bien-Soo Tan; Pierce Kh Chow; Somanesan Satchithanantham; Andrew Eh Tan; David Ce Ng; Anthony Sw Goh
Journal:  EJNMMI Res       Date:  2013-07-25       Impact factor: 3.138

4.  Post-radioembolization yttrium-90 PET/CT - part 1: diagnostic reporting.

Authors:  Yung-Hsiang Kao; Jeffrey D Steinberg; Young-Soon Tay; Gabriel Ky Lim; Jianhua Yan; David W Townsend; Angela Takano; Mark C Burgmans; Farah G Irani; Terence Kb Teo; Tow-Non Yeow; Apoorva Gogna; Richard Hg Lo; Kiang-Hiong Tay; Bien-Soo Tan; Pierce Kh Chow; Somanesan Satchithanantham; Andrew Eh Tan; David Ce Ng; Anthony Sw Goh
Journal:  EJNMMI Res       Date:  2013-07-25       Impact factor: 3.138

5.  Quantitative and qualitative assessment of Yttrium-90 PET/CT imaging.

Authors:  Ali Asgar Attarwala; Flavia Molina-Duran; Karen-Anett Büsing; Stefan O Schönberg; Dale L Bailey; Kathy Willowson; Gerhard Glatting
Journal:  PLoS One       Date:  2014-11-04       Impact factor: 3.240

6.  Phantom validation of quantitative Y-90 PET/CT-based dosimetry in liver radioembolization.

Authors:  Marco D'Arienzo; Maria Pimpinella; Marco Capogni; Vanessa De Coste; Luca Filippi; Emiliano Spezi; Nick Patterson; Francesca Mariotti; Paolo Ferrari; Paola Chiaramida; Michael Tapner; Alexander Fischer; Timo Paulus; Roberto Pani; Giuseppe Iaccarino; Marco D'Andrea; Lidia Strigari; Oreste Bagni
Journal:  EJNMMI Res       Date:  2017-11-28       Impact factor: 3.138

7.  Comparison of PET/CT and PET/MR imaging and dosimetry of yttrium-90 (90Y) in patients with unresectable hepatic tumors who have received intra-arterial radioembolization therapy with 90Y microspheres.

Authors:  Karin Knešaurek; Abbas Tuli; Edward Kim; Sherif Heiba; Lale Kostakoglu
Journal:  EJNMMI Phys       Date:  2018-08-30

8.  Optimising quantitative 90Y PET imaging: an investigation into the effects of scan length and Bayesian penalised likelihood reconstruction.

Authors:  Nathaniel P Scott; Daniel R McGowan
Journal:  EJNMMI Res       Date:  2019-05-10       Impact factor: 3.138

Review 9.  Radioembolization and the Dynamic Role of (90)Y PET/CT.

Authors:  Alexander S Pasciak; Austin C Bourgeois; J Mark McKinney; Ted T Chang; Dustin R Osborne; Shelley N Acuff; Yong C Bradley
Journal:  Front Oncol       Date:  2014-02-27       Impact factor: 6.244

10.  Comparison of posttherapy 90Y positron emission tomography/computed tomography dosimetry methods in liver therapy with 90Y microspheres.

Authors:  Karin Knešaurek
Journal:  World J Nucl Med       Date:  2020-08-22
  10 in total