Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid.

Christian Bergek1, Joachim H Zdolsek, Robert G Hahn.   

Abstract

CONTEXT: Measurement of blood haemoglobin concentration by pulse oximetry could be of value in determining when erythrocytes should be transfused during surgery, but the effect of infusion fluids on the results is unclear.
OBJECTIVE: To study the effect of crystalloid and colloid fluid on the accuracy (bias) and precision of pulse oximetry haemoglobin estimation to indicate the venous haemoglobin concentration in volunteers.
DESIGN: Open interventional crossover study.
SETTING: Single university hospital. PARTICIPANTS: Ten male volunteers aged 18-28 (mean 22) years.
INTERVENTIONS: Each volunteer underwent three infusion experiments on separate days and in random order. The infusions were Ringer's acetate (20 ml kg), hydroxyethyl starch 130/0.4 (10 ml kg) and a combination of both.
RESULTS: At the end of the infusions of Ringer's acetate, pulse oximetry haemoglobin concentration had decreased more than the true haemoglobin concentration (15 vs. 8%; P < 0.005; n = 10) whereas starch solution decreased pulse oximetry haemoglobin concentration less than true haemoglobin concentration (7 vs. 11%; P < 0.02; n = 20). The same differences were seen when the fluids were infused separately and when they were combined. The overall difference between all 956 pairs of pulse oximetry haemoglobin concentration and true haemoglobin concentrations (the bias) averaged only -0.7 g l whereas the 95% prediction interval was wide, ranging from -24.9 to 23.7 g l. In addition to the choice of infusion fluid, the bias was strongly dependent on the volunteer (each factor, P < 0.001).
CONCLUSION: The bias of measuring haemoglobin concentration by pulse oximetry is dependent on whether a crystalloid or a colloid fluid is infused. Trial registration ClinicalTrials identifier: NCT01195025.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23047301     DOI: 10.1097/EJA.0b013e3283592733

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  6 in total

Review 1.  Accuracy of pleth variability index to predict fluid responsiveness in mechanically ventilated patients: a systematic review and meta-analysis.

Authors:  Haitao Chu; Yong Wang; Yanfei Sun; Gang Wang
Journal:  J Clin Monit Comput       Date:  2015-08-05       Impact factor: 2.502

2.  Population Kinetics of 0.9% Saline Distribution in Hemorrhaged Awake and Isoflurane-anesthetized Volunteers.

Authors:  Joakim Nyberg; Husong Li; Pehr Wessmark; Viktor Winther; Donald S Prough; Michael P Kinsky; Christer H Svensén
Journal:  Anesthesiology       Date:  2019-09       Impact factor: 7.892

3.  The use of a noninvasive hemoglobin monitor for determining fluid distribution and elimination in pediatric patients undergoing minor surgery.

Authors:  Qingquan Lian; Husong Li; Ruifeng Zeng; Junhui Lang; Wangning Shangguan; Huacheng Liu; Benfu Wang; Peter M Rodhe; Christer H Svensen
Journal:  J Clin Monit Comput       Date:  2014-02-19       Impact factor: 2.502

4.  Can perioperative hemodilution be monitored with non-invasive measurement of blood hemoglobin?

Authors:  Robert G Hahn; Patrick Y Wuethrich; Joachim H Zdolsek
Journal:  BMC Anesthesiol       Date:  2021-05-06       Impact factor: 2.217

5.  Interactions between the volume effects of hydroxyethyl starch 130/0.4 and Ringer´s acetate.

Authors:  Robert G Hahn; Christian Bergek; Tobias Gebäck; Joachim Zdolsek
Journal:  Crit Care       Date:  2013-05-29       Impact factor: 9.097

6.  [Noninvasive hemoglobin monitoring in clinical trials: a systematic review and meta-analysis].

Authors:  Tailyne Zortéa; Daniele Paola da Silva Wizbicki; Kristian Madeira; Pedro Gabriel Ambrosio; Rafaela Okchstein Borges de Souza; Edson Souza Machado Durães
Journal:  Braz J Anesthesiol       Date:  2020-06-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.