Implementation of dual-energy technique for virtual monochromatic and linearly mixed CBCTs.

Hao Li1, William Giles, Lei Ren, James Bowsher, Fang-Fang Yin.   

Abstract

PURPOSE: To implement dual-energy imaging technique for virtual monochromatic (VM) and linearly mixed (LM) cone beam CTs (CBCTs) and to demonstrate their potential applications in metal artifact reduction and contrast enhancement in image-guided radiation therapy (IGRT).
METHODS: A bench-top CBCT system was used to acquire 80 kVp and 150 kVp projections, with an additional 0.8 mm tin filtration. To implement the VM technique, these projections were first decomposed into acrylic and aluminum basis material projections to synthesize VM projections, which were then used to reconstruct VM CBCTs. The effect of VM CBCT on the metal artifact reduction was evaluated with an in-house titanium-BB phantom. The optimal VM energy to maximize contrast-to-noise ratio (CNR) for iodine contrast and minimize beam hardening in VM CBCT was determined using a water phantom containing two iodine concentrations. The LM technique was implemented by linearly combining the low-energy (80 kVp) and high-energy (150 kVp) CBCTs. The dose partitioning between low-energy and high-energy CBCTs was varied (20%, 40%, 60%, and 80% for low-energy) while keeping total dose approximately equal to single-energy CBCTs, measured using an ion chamber. Noise levels and CNRs for four tissue types were investigated for dual-energy LM CBCTs in comparison with single-energy CBCTs at 80, 100, 125, and 150 kVp.
RESULTS: The VM technique showed substantial reduction of metal artifacts at 100 keV with a 40% reduction in the background standard deviation compared to a 125 kVp single-energy scan of equal dose. The VM energy to maximize CNR for both iodine concentrations and minimize beam hardening in the metal-free object was 50 keV and 60 keV, respectively. The difference of average noise levels measured in the phantom background was 1.2% between dual-energy LM CBCTs and equivalent-dose single-energy CBCTs. CNR values in the LM CBCTs of any dose partitioning are better than those of 150 kVp single-energy CBCTs. The average CNR for four tissue types with 80% dose fraction at low-energy showed 9.0% and 4.1% improvement relative to 100 kVp and 125 kVp single-energy CBCTs, respectively. CNRs for low-contrast objects improved as dose partitioning was more heavily weighted toward low-energy (80 kVp) for LM CBCTs.
CONCLUSIONS: Dual-energy CBCT imaging techniques were implemented to synthesize VM CBCT and LM CBCTs. VM CBCT was effective at achieving metal artifact reduction. Depending on the dose-partitioning scheme, LM CBCT demonstrated the potential to improve CNR for low contrast objects compared to single-energy CBCT acquired with equivalent dose.

Entities:  

Mesh:

Year:  2012        PMID: 23039644     DOI: 10.1118/1.4752212

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Characterization and potential applications of a dual-layer flat-panel detector.

Authors:  Linxi Shi; Minghui Lu; N Robert Bennett; Edward Shapiro; Jin Zhang; Richard Colbeth; Josh Star-Lack; Adam S Wang
Journal:  Med Phys       Date:  2020-05-18       Impact factor: 4.071

2.  Evaluation of the effect of respiratory and anatomical variables on a Fourier technique for markerless, self-sorted 4D-CBCT.

Authors:  I Vergalasova; J Cai; W Giles; W P Segars; F F Yin
Journal:  Phys Med Biol       Date:  2013-09-24       Impact factor: 3.609

Review 3.  Technical Principles of Dual-Energy Cone Beam Computed Tomography and Clinical Applications for Radiation Therapy.

Authors:  Shailaja Sajja; Young Lee; Markus Eriksson; Håkan Nordström; Arjun Sahgal; Masoud Hashemi; James G Mainprize; Mark Ruschin
Journal:  Adv Radiat Oncol       Date:  2019-07-30

4.  Investigating the feasibility of generating dual-energy CT from one 120-kVp CT scan: a phantom study.

Authors:  Wen-Hui Huang; Kai-Jie Jhan; Ching-Ching Yang
Journal:  J Appl Clin Med Phys       Date:  2020-10-14       Impact factor: 2.102

5.  Improving Image Quality of On-Board Cone-Beam CT in Radiation Therapy Using Image Information Provided by Planning Multi-Detector CT: A Phantom Study.

Authors:  Ching-Ching Yang; Fong-Lin Chen; Yeh-Chi Lo
Journal:  PLoS One       Date:  2016-06-09       Impact factor: 3.240

6.  Metal artifact reduction by filter-based dual-energy cone-beam computed tomography on a bench-top micro-CBCT system: concept and demonstration.

Authors:  Hiraku Iramina; Takumi Hamaguchi; Mitsuhiro Nakamura; Takashi Mizowaki; Ikuo Kanno
Journal:  J Radiat Res       Date:  2018-07-01       Impact factor: 2.724