Immunometric assay by flow cytometry using mixtures of two particle types of different affinity.

T Lindmo1, O Børmer, J Ugelstad, K Nustad.   

Abstract

An improved dynamic range in a particle based flow cytometric immunoassay for carcinoembryonic antigen (CEA) was obtained using a binary mixture of two distinguishable particle types, namely particles of 7 and 10 microns diameter that were distinguishable by their light scattering characteristics in the flow cytometer. The two particle types were coated with antibody of the same specificity but different affinity. The association constants were 3.2 x 10(10) and 3.3 x 10(9) for the antibodies on the 7 and 10 micron particles, respectively. A dilution series of CEA samples was incubated with aliquots of the particle mixture and secondary biotin-streptavidin-phycoerythrin-conjugated antibody directed against a different epitope on the CEA molecule. The fluorescence intensity of the two particle types was measured flow cytometrically, and a double standard curve plotted from the mean logarithmic fluorescence values. The precision profile derived from the standard curve demonstrated that an increase in the dynamic range of about 50% (from 2 to 3 log) was obtained by using a mixture of high and low affinity particles, compared to using the high affinity particles alone.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2303729     DOI: 10.1016/0022-1759(90)90149-p

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303


  1 in total

1.  Flow cytometric microsphere-based immunoassay: analysis of secreted cytokines in whole-blood samples from asthmatics.

Authors:  C Camilla; L Mély; A Magnan; B Casano; S Prato; S Debono; F Montero; J P Defoort; M Martin; V Fert
Journal:  Clin Diagn Lab Immunol       Date:  2001-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.