Simulation of 20-year deterioration of acrylic IOLs using severe accelerated deterioration tests.

Kenji Kawai1, Kenji Hayakawa, Takahiro Suzuki.   

Abstract

PURPOSE: To investigate IOL deterioration by conducting severe accelerated deterioration testing of acrylic IOLs.
SETTING: Department of Ophthalmology, Tokai University School of Medicine
Methods: Severe accelerated deterioration tests performed on 7 types of acrylic IOLs simulated 20 years of deterioration. IOLs were placed in a screw tube bottle containing ultra-pure water and kept in an oven (100°C) for 115 days. Deterioration was determined based the outer appearance of the IOL in water and under air-dried conditions using an optical microscope. For accelerated deterioration of polymeric material, the elapse of 115 days was considered to be equivalent to 20 years based on the Arrhenius equation.
RESULTS: All of the IOLs in the hydrophobic acrylic group except for AU6 showed glistening-like opacity. The entire optical sections of MA60BM and SA60AT became yellowish white in color. Hydrophilic acrylic IOL HP60M showed no opacity at any of the time points examined.
CONCLUSION: Our data based on accelerated testing showed differences in water content to play a major role in transparency. There were differences in opacity among manufacturers. The method we have used for determining the relative time of IOL deterioration might not represent the exact clinical setting, but the appearance of the materials would presumably be very similar to that seen in patients.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23032245

Source DB:  PubMed          Journal:  Tokai J Exp Clin Med        ISSN: 0385-0005


  5 in total

1.  Polarized glasses may help in symptomatic cases of intraocular lens glistenings.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein
Journal:  Clin Optom (Auckl)       Date:  2019-05-06

2.  Surgical experience with a redesigned, fully preloaded, hydrophobic acrylic intraocular lens in challenging cases of pseudoexfoliation syndrome, phacodonesis, and small pupils.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein
Journal:  Clin Ophthalmol       Date:  2019-01-22

3.  In-vitro glistening formation in six different foldable hydrophobic intraocular lenses.

Authors:  Tamer Tandogan; Gerd U Auffarth; Hyeck-Soo Son; Patrick Merz; Chul Young Choi; Ramin Khoramnia
Journal:  BMC Ophthalmol       Date:  2021-03-08       Impact factor: 2.209

4.  Scanning electron microscopic features of explanted degraded hydrophobic acrylic intraocular lenses which were in vivo for a prolonged period.

Authors:  Harsha Bhattacharjee; Suklengmung Buragohain; Henal J Javeri; Dipankar Das
Journal:  Indian J Ophthalmol       Date:  2020-06       Impact factor: 1.848

5.  Comparative analysis of in vitro accelerated glistening formation in foldable hydrophobic intraocular lenses.

Authors:  Tamer Tandogan; Gerd U Auffarth; Chul Young Choi; Hyeck-Soo Son; Ramin Khoramnia
Journal:  Int Ophthalmol       Date:  2021-05-19       Impact factor: 2.031

  5 in total