OBJECTIVE: The objective of this study was to investigate the applicability of micro-analytical methods with high spatial resolution to the characterization of the composition and corrosion behavior of two bracket systems. MATERIAL AND METHODS: The surfaces of six nickel-free brackets and six nickel-containing brackets were examined for signs of corrosion and qualitative surface analysis using an electron probe microanalyzer (EPMA), prior to bonding to patient's tooth surfaces and four months after clinical use. The surfaces were characterized qualitatively by secondary electron (SE) images and back scattered electron (BSE) images in both compositional and topographical mode. Qualitative and quantitative wavelength-dispersive analyses were performed for different elements, and by utilizing qualitative analysis the relative concentration of selected elements was mapped two-dimensionally. The absolute concentration of the elements was determined in specially prepared brackets by quantitative analysis using pure element standards for calibration and calculating correction-factors (ZAF). RESULTS: Clear differences were observed between the different bracket types. The nickel-containing stainless steel brackets consist of two separate pieces joined by a brazing alloy. Compositional analysis revealed two different alloy compositions, and reaction zones on both sides of the brazing alloy. The nickel-free bracket was a single piece with only slight variation in element concentration, but had a significantly rougher surface. After clinical use, no corrosive phenomena were detectable with the methods applied. Traces of intraoral wear at the contact areas between the bracket slot and the arch wire were verified. CONCLUSION: Electron probe microanalysis is a valuable tool for the characterization of element distribution and quantitative analysis for corrosion studies.
OBJECTIVE: The objective of this study was to investigate the applicability of micro-analytical methods with high spatial resolution to the characterization of the composition and corrosion behavior of two bracket systems. MATERIAL AND METHODS: The surfaces of six nickel-free brackets and six nickel-containing brackets were examined for signs of corrosion and qualitative surface analysis using an electron probe microanalyzer (EPMA), prior to bonding to patient's tooth surfaces and four months after clinical use. The surfaces were characterized qualitatively by secondary electron (SE) images and back scattered electron (BSE) images in both compositional and topographical mode. Qualitative and quantitative wavelength-dispersive analyses were performed for different elements, and by utilizing qualitative analysis the relative concentration of selected elements was mapped two-dimensionally. The absolute concentration of the elements was determined in specially prepared brackets by quantitative analysis using pure element standards for calibration and calculating correction-factors (ZAF). RESULTS: Clear differences were observed between the different bracket types. The nickel-containing stainless steel brackets consist of two separate pieces joined by a brazing alloy. Compositional analysis revealed two different alloy compositions, and reaction zones on both sides of the brazing alloy. The nickel-free bracket was a single piece with only slight variation in element concentration, but had a significantly rougher surface. After clinical use, no corrosive phenomena were detectable with the methods applied. Traces of intraoral wear at the contact areas between the bracket slot and the arch wire were verified. CONCLUSION: Electron probe microanalysis is a valuable tool for the characterization of element distribution and quantitative analysis for corrosion studies.
Concomitant increases in the prevalence of allergic reactions to alloys, predominantly
nickel (Ni) and chromium (Cr), have added to the general interest in alloy composition
and corrosive phenomena in dentistry. Ni-containing alloys with a Ni content exceeding
50 wt.-% are routinely applied in restorative dentistry due to cost-effectiveness and
ideal material properties such as strength, elasticity, and hardness[18]. However, unlike high-noble alloys,
which exhibit good resistance to corrosion due to their low reactivity as stated by
Geurtsen[8](2002), Ni-containing
alloys are more prone to surface wear[16].Corrosion cannot be entirely prevented in the oral cavity, as dental materials and
orthodontic appliances are exposed to various biological, mechanical, and environmental
stresses depending on their material composition, manufacturing process, and
microstructure[9,21,26]. Galvanic
interaction in orthodontics is most likely to occur between the archwire and the bracket
due to friction, as shown by Eliades and Athanasiou[6] (2002) or can occur within a bracket's own components[29]. While conventional metal brackets
contain different stainless steel alloys in the bracket base and tie wings, which are
then soldered with silver (Ag), Ni, or gold alloys, alternative manufacturing techniques
such asmetal injection molding are applied in an attempt to minimize corrosive
potential. In an in vitro study, Siargos, et al.[27] (2007) stated that single-component
brackets may provide a measurable benefit when compared to conventional brackets, due to
uniform elemental distribution. Also external factors such as the interaction of
fluoride containing solutions and metals such astitanium are well investigated in
dental literature and should not be underestimated when interpreting results[4,25].Controversy exists regarding whether the ion release from orthodontic appliances due to
corrosion has a localized or systemic effect on patient health. An in
vitro study on human cell cultures[28]has described potential carcinogenic, mutagenic, and cytotoxic
effects of metal ions in general, while Tomakidi, et al.[24] (2000) found no such effects. The amount of ions
released from orthodontic appliances in vivo remains unclear.Hypersensitivity is the most common effect to Ni and Cr, and is frequently reported in
dermatology. In a review, Noble, et al.[15] (2008) reported on the influence of gender for the prevalence of
nickelhypersensitivity which is supported by other research results[12]. With the exception of some isolated
case reports on gingival hyperplasia, labial desquamation, angular cheilitis, swelling,
and burning sensations of the oral mucosa, a general correlation between the respective
ions and hypersensitive reaction has not been substantiated in dentistry[23]. In a review of the literature, House,
et al.[26] (2008) stated that most
patients with confirmed Ni sensitivity show no reaction to intra-oral Ni-containing
alloys.There are three ways to investigate metal ion release: in vitro,
retrieval (ex-vivo investigation of in-vivo aged
samples), and in vivo. in vitro investigations are
excellent methods to investigate a specific topic while eliminating interfering side
effects. in vitro results can only provide general guidance due to the
multi-factorial environmental conditions in the oral cavity, but these studies must be
considered because there are few alternatives[11].In vivo research lacks obtainable explanatory power, as saliva, blood,
and urine samples are always a representation of local effects and general external
influences. The greatest challenge of in vivo investigations concerning
ion release of alloys is differentiating between ions released through corrosion and ion
intake from nutrition. On average, 200 to 300 mg Ni and 280 mg Cr are consumed by a
person per day[11].
In vivo investigations using saliva samples to determine the ionic
release of alloys found Cr and Ni concentrations far below that of normal dietary
intake[3].An alternative investigative method is retrieval studies that try to combine in
vivo and in vitro methodologies. As opposed to medical
disciplines such as orthopedics, where ex-vivo investigations on hip
replacements can only be performed if an implant fails, orthodontic brackets and wires
offer the possibility of easy retrieval. Before initiating time- and cost-intensive
research projects, analytical methods must be validated in pilot investigations. While
several ex-vivo studies have been performed on orthodontic archwires,
no such studies are available for different bracket systems. To specify and measure the
amount of ion release (especially the release of Ni and Cr ions) during orthodontic
treatment, highly sensitive analytical methods are needed.The aim of the present pilot investigation was to assess whether surface changes like
ionic reduction occur on alloy surfaces during clinical use before extensive clinical
in vivo investigations are initiated. In a similar study, Eliades,
et al.[7] (2004) investigated Ni release
from two different orthodontic archwires (stainless steel and Nickel-Titanium [Ni-Ti]
wires) retrieved after an intra-oral service period between 1.5 and 12 months and
compared them with as-received wires. Scanning electron microscopy (SEM) and
energy-dispersive X-ray microanalysis were used to assess the elemental composition of
the wires. They found no significant differences in Ni content between as-retrieved and
as-received wires.Electron probe microanalysis is a standard method in the quality control of steel
manufacturing. Special protocols have been developed for wavelength-dispersive
spectrometry (WDS) to simultaneously obtain an element distribution map, the
concentrations of the elements and a resulting phase map, as well as a volumetric
estimate of all phases present[1]. In
the field of orthodontic applications, stainless steels are predominantly analyzed with
a scanning electron microscope equipped with electron detectors and a semiconductor
X-ray detector[29,30]. In this case, the quantitative analysis is based on
energy-dispersive spectrometry (EDS) without calibration (standardization), but with a
correction procedure for atomic number (Z), absorption (A), and fluorescence
(F)[30]. The energy resolution and
the intensity ratio between line maximum (peak) and background
(IPK/IBG) is less in EDS than in WDS. The non-standardized
ZAF-corrected EDS analysis is sufficient for a general compositional characterization of
the materials. However, in more detailed studies, such as phase composition and
corrosion potential, WDS of electron microprobe analysis allows lower concentrations to
be analyzed and, by calibration, more reliable results for elements present in low
concentration (<0.5% by weight).The hypothesis of this study was that there is a difference in corrosive potential in
orthodontic brackets of different alloy composition and that the sensitivity of electron
probe microanalysis allows detection of corrosive potential in the metal surfaces.
MATERIAL AND METHODS
In the ex-vivo/retrieval study, quantitative and qualitative analysis
of two bracket systems (Ni-free and Ni-containing) was performed focusing on corrosion
resistance, alloy composition, and homogeneity. To compare element component alteration
between the states prior to bracket bonding and after retrieval, each bracket was
documented in images of secondary electrons (SE) and backscattered electrons (BSE) at
low magnification. In addition, several pictures at higher magnification were obtained
from selected areas.
Patients
Prior to initiation of the pilot investigation, approval was obtained from the local
ethics committee of the University (IRB). All participants were patients at the
department of orthodontics in need of orthodontic treatment with fixed appliances.
Since all patients were adolescents, written consent to participate was obtained from
both parents. Twelve patients were randomly divided into two groups (Group 1 and
Group 2). Exclusion criteria were any metallic restoration or the need for additional
metallic orthodontic devices such as headgear or a trans-palatal arch (TPA). The six
patients in Group 1 received Ni-free brackets (Sprint-Bracket, Forestadent,
Pforzheim, Germany) and the six patients in Group 2 received Ni-containing brackets
(Victory Series MBT RX, 3M Unitek, Monrovia, USA).Full arch bonding was performed in the upper and lower jaw using established bonding
protocols. To isolate the effect of bracket type, the same type of orthodontic
archwires were used for both groups (.014 Sentalloy, GAC International Inc., Bohemia,
NY, USA). After 4 months of clinical use, the bracket of tooth 14 was removed from
each patient and re-examined.
Electron probe microanalysis
Electron probe microanalysis is a destruction-free physical microanalytical method to
determine the chemical composition of solid matter in high spatial resolution using
detectors for electrons and for X-rays. The principle of electron probe
microanalysis, as well as basic descriptions of instrumentation and technique, are
summarized in Reed[19] (1997) and
Potts[7](1992).The electron probe microanalyzer (JEOL Superprobe JXA-8200, Tokyo, Japan) is equipped
with an energy-dispersive system [EDS; Si(Li)] and five wavelength-dispersive (WD)
spectrometers. Two of the WD spectrometers (CH-1, CH-2) have argon-methane gas flow
counters (Ar:CH4 90%:10% by volume, P10-gas), two spectrometers (CH-3,
CH-4) have sealed xenon (Xe)-filled counters, all with a Rowland circle of 140 mm
radius. One spectrometer (CH-5) is a high-resolution spectrometer with larger
monochromator crystals, a sealed Xe-filled counter, and a Rowland circle with a
radius of 100 mm.
Qualitative analysis
Elements present at higher concentrations were initially detected by qualitative EDS
analysis. Element distribution (intensity) maps of rectangular areas were obtained
using WDS spectrometers (20 kV/20 nA, focused beam, 5 µm step width, 1000 ms dwell
time). For spectrometers CH-3, CH-4, and CH-5, the upper part of the basal plate may
project into the X-ray path due to the take-off angle of 40º. Therefore, the mapped
area had to be carefully selected to avoid absorption of the X-rays by the basal
plate. Prior to the quantitative analysis, WDS scans (20 kV/50 nA, 25 µm step width,
500 ms dwell time) were also performed over the whole accessible wavelength range to
detect elements with minor concentrations. Some scans were also performed to obtain
the correct spectrometer offsets to higher and lower background positions, as well as
information on possible line interferences.
Quantitative analysis
For the quantitative analyses (20 kV/20 nA, focused beam), the spectrometers
(channels), selected monochromator crystals, elements, and lines were: CH-1 PET:
Cr-Ka; CH-2 TAP: Si-Ka; CH-3 PET: Mo-La; CH-4 LIF: Fe-Ka, Mn-Ka, Ni-Ka, and Cu-Ka;
and CH-5 LIFH: Co-Ka. To get reliable counting statistics the measurement times on
the peak and the two background positions were set to 20/10/10 s for the major
elements [Iron (Fe), Cr, Ni, and Manganese (Mn)], to 30/15/15 s for Molybdenum (Mo),
and to 40/20/20 s for Silicon (Si) and Cobalt (Co). No light elements [Carbon (C),
Nitrogen CN)] were analyzed during this preliminary study. Prior to measurement, the
elements were calibrated on pure element standards. Two analysis routines were set up
according to the alloys of the two bracket types. Both the calibration and the effect
of flank or line overlap were checked by analysis of the standards using the previous
analysis routines. The quantitative analysis program utilizes the CITZAF correction
by Armstrong[2] (1995) as implemented
by the manufacturer.
Sample preparation
As electron probe microanalysis needs the analyzed surfaces to be planar and
horizontally oriented, two types of specimen holders for both investigated bracket
types (Figure 1) and a plate needed for
guidance were custom made.
Figure 1
Specimen holder with mounted bracket in holder block
Specimen holder with mounted bracket in holder blockThe brackets were cleaned with petroleum ether (Rotipuran, ACS, Carl Roth, Karlsruhe,
Germany) to remove surface contamination. For the investigation of the original
surface prior to insertion, the cleaned brackets were otherwise mounted as delivered.
After recovery from the patient, the organic substances and contamination from the
oral cavity deposited on the bracket surface were carefully manually removed using
petroleum ether, interdental brushes, toothsticks, and cotton wool.To obtain information on the internal fabric and to perform quantitative analysis,
one additional bracket of each type was selected. The part of the basal plate
projecting into the ray path was removed and a planar surface was then obtained by
careful manual grinding and polishing to avoid fabric distortion.
RESULTS
Figure 2 compares SE, BSE compositional, and BSE
topographical images of one sample of both investigated bracket types at low
magnification. Most brackets showed scratches resulting from clinical use (Figure 3). It is evident from higher-magnification
images (Figure 4) that both bracket types are
characterized, with slight differences between the individual samples of the same type,
by a surface roughness caused by steps and by linear depressions or cavities. Comparison
of the same surfaces before insertion and after retrieval shows no apparent
modifications in the mechanically undisturbed areas (Figures 5 and 6). In several places of
the investigated bracket, and observed to a lesser degree at the other brackets, some
pits appeared to have an increased size after retrieval.
Figure 2
Overview images of two brackets, as delivered and after cleaning with petroleum
ether. (a-c) Group 1; (d-f) Group 2. Images: (a, d) secondary electron (SE), back
scattered electron in (b, e) compositional mode (BSE), and in (c, f) topographic
mode (TOPO)
Figure 3
Overview secondary electron (SE) images of Group 1 brackets, after recovery from
patients (a) with dried plaque and (b) after cleaning
Figure 4
Surface details of brackets at larger magnification displaying the rough surface
of the wing section. (a -b) Group 1; (c-d) Group 2. Images: (a, c) secondary
electron (SE), (b, d) back scattered electron in compositional mode (BSE) In BSE
images the rectangles mark the position of the element distribution maps: (b)
Group 1 (H 350 μm, V 750 μm), (d) Group 2 (H 650 μm, V 250 μm). The size and
orientation of the element distribution maps is confined by the geometry of the
bracket, especially the part of the basal plate projecting into the ray path
between excitation volume and spectrometers
Figure 5
Detailed secondary electron (SE) and back scattered electron in topographic mode
(BSE-TOPO) images of characteristic features before insertion (a-b) and after
recovery and cleaning (c-d) of a Group 1 bracket. Besides the obvious scratches
present in the wing portion after recovery, cavities in mechanically untreated
areas show no detectable change. In the brazing alloy part, a few minor changes
are visible at the right-hand side of the image, very likely due to mechanical
treatment during retrieval
Figure 6
Detailed secondary electron (SE) and back scattered electron in topographic mode
(BSE-TOPO) images of characteristic features before insertion (a-b) and after
recovery and cleaning (c-d) of a Group 2 bracket. After recovery, cavities in
mechanically untreated areas show no detectable change
Overview images of two brackets, as delivered and after cleaning with petroleum
ether. (a-c) Group 1; (d-f) Group 2. Images: (a, d) secondary electron (SE), back
scattered electron in (b, e) compositional mode (BSE), and in (c, f) topographic
mode (TOPO)Overview secondary electron (SE) images of Group 1 brackets, after recovery from
patients (a) with dried plaque and (b) after cleaningSurface details of brackets at larger magnification displaying the rough surface
of the wing section. (a -b) Group 1; (c-d) Group 2. Images: (a, c) secondary
electron (SE), (b, d) back scattered electron in compositional mode (BSE) In BSE
images the rectangles mark the position of the element distribution maps: (b)
Group 1 (H 350 μm, V 750 μm), (d) Group 2 (H 650 μm, V 250 μm). The size and
orientation of the element distribution maps is confined by the geometry of the
bracket, especially the part of the basal plate projecting into the ray path
between excitation volume and spectrometersDetailed secondary electron (SE) and back scattered electron in topographic mode
(BSE-TOPO) images of characteristic features before insertion (a-b) and after
recovery and cleaning (c-d) of a Group 1 bracket. Besides the obvious scratches
present in the wing portion after recovery, cavities in mechanically untreated
areas show no detectable change. In the brazing alloy part, a few minor changes
are visible at the right-hand side of the image, very likely due to mechanical
treatment during retrievalDetailed secondary electron (SE) and back scattered electron in topographic mode
(BSE-TOPO) images of characteristic features before insertion (a-b) and after
recovery and cleaning (c-d) of a Group 2 bracket. After recovery, cavities in
mechanically untreated areas show no detectable changeNarrowly-spaced changes in composition were visible in the element distribution maps. In
Group 2, the compositional contrast between wings and brazing alloy was striking,
especially for Ni and Si. The narrow net-like pattern that was visible in BSE, Cr, and
Si intensity maps (Figure 7A-C) is apparently not a surface effect. In addition, several spots of
brazing alloy composition were detected in SE and BSE images of the wing portion of the
shown bracket (Figure 4) and were already visible
by their lighter grey value in the BSE image (Figure
7A). In the Group 1 brackets, BSE images show polygonal to irregular spots of
different grey values (Figure 2, Figure 7E), corresponding to two different alloy
compositions with anti-correlating Mn and Mo concentrations. Therefore a compositionally
homogenous surface was not present, and, as a consequence, there is potential for
corrosion.
Figure 7
Back scattered electron (BSE) image and enhanced element distribution (intensity)
maps of two investigated brackets: (a-d) Group 1 series, (e-h) Group 2 series. The
grey value in BSE CP (compositional) images corresponds to a difference in mean
atomic number (compare to Figures 4b, d). In the element distribution maps, the
intensities, and thus the relative concentrations, are colour-coded; colour scale
from low intensity (blue) to high intensity (white). In (c), showing the
concentration of Si, two maps were combined due to very different Si
concentrations in steel and brazing alloy. Note in group 1 series the sharp
contrasts in composition between brazing alloy and wing (b-d) and in group 2
series the juxtaposed areas of differing composition (f-h). In (f), the intensity
gradient is the result of a slightly inclined sample surface causing the
excitation volume to move out of the spectrometer focus
Back scattered electron (BSE) image and enhanced element distribution (intensity)
maps of two investigated brackets: (a-d) Group 1 series, (e-h) Group 2 series. The
grey value in BSE CP (compositional) images corresponds to a difference in mean
atomic number (compare to Figures 4b, d). In the element distribution maps, the
intensities, and thus the relative concentrations, are colour-coded; colour scale
from low intensity (blue) to high intensity (white). In (c), showing the
concentration of Si, two maps were combined due to very different Si
concentrations in steel and brazing alloy. Note in group 1 series the sharp
contrasts in composition between brazing alloy and wing (b-d) and in group 2
series the juxtaposed areas of differing composition (f-h). In (f), the intensity
gradient is the result of a slightly inclined sample surface causing the
excitation volume to move out of the spectrometer focusQuantitative analysis showed that two different steels were used for the base and wings
of the group 2 brackets. Nearly all the material of the base consisted of Cr- and
Ni-rich steel, with minor amounts of Mo and copper (Cu) (Table 1, analysis 1). A relatively Cr- and Mo-rich, Ni-poor material
was present in very small quantity and confined to grain boundaries (Table 1, analysis 2). The wings consisted of a
steel lower in Cr, Mn, Mo, and Ni, but considerably higher in Cu (Table 1, analysis 4). Generally, the steel of the basal plate was
similar to 1.4550 (SST-347, UNS S 34700, X6 CrNiNb 18 10), except for the presence of Mo
and the lack of detectable Niobium (Nb). The steel of the wings resembled 1.4542
(17-4PH; X5CrNiCuNb16-4), except that Nb and Tantalum (Ta) were not detected during our
study. One composition of the brazing alloy in the center of a larger area of the filler
(Table 1, analysis 3) shows a Si- and Ni-rich
composition. Some adjacent material was analyzed due to the small size of the grain, but
the composition tends towards the one given by Zinelis, et al.[29](2004) for the braze alloy of Gemini brackets (3M
Unitek).
Table 1
Chemical composition of brackets (wt.-%)
Element
Cr
Si
Mo
Mn
Fe
Co
Ni
Cu
Total
Comment
Group 2 (Ni containing)
1
18,700
0.415
0.496
2,000
68,100
0,000
9,810
0.338
99,920
base, light grey
2
26,700
0.448
0.866
1,650
65,100
0,000
4,610
0.197
99,590
base, dark grey
3
9,860
13,100
0.041
5,070
7,380
0,000
62,000
1,040
98,540
brazing alloy
4
16,100
0.363
0.133
0.671
75,400
0,000
3,950
3,050
99,670
wing
Group 1 (Ni-free)
5
17,500
0.663
3,110
11,100
67,500
0.072
0.069
0,000
100,020
core
6
17,300
0.798
4,590
9,950
67,300
0.065
0.064
0,000
100,080
rim
7
17,100
1,020
6,900
8,750
66,400
0.065
0.055
0,000
100,310
phase, light grey
Chemical composition of brackets (wt.-%)The group 1 bracket consists of a virtually Ni-free high-N austenitic steel (X15 CrMnMoN
17 11 3) named P.A.N.A.C.E.A. (Protection Against Ni Allergy, Corrosion, Erosion and
Abrasion) developed at the Institute of Metallurgy, ETH Zürich[27]. The compositional range is given (in weight
percentages) asCr 16.5-17.5%, Mo 3.0-3.5%, Mn 10-12%, N 0.6-1.2%, Fe balance[13], or asCr 15-18%, Mo 3-6%, Mn 10-12%, N
0.9%, Fe balance[27], the average values
asCr 17.3%. Mo 3.2%, Ni<0.05%, Mn 12%, N 0.9%, balance Fe[22]. When properly processed the steel is fully
austenitic[22].
DISCUSSION
Both working hypotheses, that varying orthodontic brackets have alloys with different
corrosive potentials, and that the sensitivity of the electron probe microanalysis is an
adequate measure to detect small changes in element composition, were accepted.
Investigation of six Ni-free brackets and six Ni-containing brackets revealed that
corrosion potential exists due to surface morphology and the juxtaposition of materials
of different composition. The use of three different materials for the Ni-containing
brackets (i.e. basal plate and wings joined by a brazing alloy) yields sharp edges where
the brazing alloy projects over the other parts.The observation that some pits appeared to have increased in size after retrieval cannot
be attributed to corrosion phenomena, as mechanical treatment during the period of wear
or during retrieval cannot be excluded. Major concerns pertaining to the
biocompatibility, potential hypersensitivity, and corrosion resistance of metallic
orthodontic devices could not be proven or disproven due to the pilot character of the
investigation, but results clearly underscore an obvious corrosion potential in the
examined specimens[5,21]. Scientific data on corrosion potential in orthodontic
devices is contradictory. A previous in vitro investigation[5 ]found that while none of the tested
orthodontic wires and brackets was susceptible to pitting corrosion, galvanic corrosion
was measurable after 28 days of exposure in lactic acid solution. Siargos, et
al.[21] (2007) compared
conventional and metal injection molded brackets with commonly used orthodontic
archwires and found comparable potential differences for both bracket types. The
detected steps and linear depressions or cavities can be considered additional primary
loci for corrosion, especially as such voids provide spaces for
microbial plaque that cannot be easily removed by conventional tooth cleaning
procedures. The same is true for the observed areas of brazing alloy composition in
Group 2 that most likely resemble material sputtered onto the wing during the procedure
of joining the basal plate and the wing, thus producing local areas with greater
potential for corrosion by half cell generation. Additionally, as shown by Siargos, et
al.[21] (2007), the heterogeneous
material composition, manufacturing processes, and microstructure of brackets have an
immediate influence on the corrosion potential, and the galvanic susceptibility might be
triggered by applying different archwires.Organic substances and contamination deposited on the bracket surface had to be removed
prior to the re-assessment of the retrieved brackets. This phenomenon of a protective
passive film, on Fe-Cr-Ni-based stainless steel alloys composed of
Cr2O3 is well known. In an investigation by Lin, et
al.[14] (2006) a variation in the
corrosion resistance of different stainless steel brackets is clearly shown, even though
they all had an identical passive film structure. In the oral cavity, this passive film
is susceptible to mechanical and chemical factors (e.g. carbonate drinks or fluoride
containing products). An increase of corrosion susceptibility caused by toothpastes and
mouthwashes containing fluoride has also been shown in in vitro
investigations[20,25] which demonstrated the destruction of the oxide layer.
In an in vivo split mouth study, Harzer, et al.[10] (2001) investigated the sensitivity of
titanium and stainless steel brackets to fluoride toothpaste and tea. They found no
significant differences, which is contrary to the findings of other studies but can
probably be explained by the short exposure time. This phenomenon has an immediate
effect on orthodontic treatment. Corrosion of the bracket or wire surface leads to
surface roughening and a resultant increase of adhering hard tartar. As a consequence,
control of orthodontic forces may become unpredictable.The increasing availability and application of different materials for orthodontic
treatment, coupled with the complexity of the oral environment, indicate a need for
further investigations on the biocompatibility of metallic materials to ensure patient
health.
CONCLUSION
Electron probe microanalysis is a valuable tool for the characterization of element
distribution and quantitative analysis for corrosion studies. The duration and intensity
of the exposure of the brackets to saliva and food were not sufficient to produce
corrosion observable with SE or BSE. Future studies employing larger sample sizes and
longer periods of clinical use will provide important information on the local behavior
of orthodontic devices.