Literature DB >> 23015562

Differences in long-term fixation between mobile-bearing and fixed-bearing knee prostheses at ten to 12 years' follow-up: a single-blinded randomised controlled radiostereometric trial.

B G Pijls1, E R Valstar, B L Kaptein, R G H H Nelissen.   

Abstract

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (SD 0.76) in the MB group and 0.66 mm (SD 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores. For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (SD 0.014). This was significantly smaller than the wear rate of 0.11 mm/year (SD 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant.

Entities:  

Mesh:

Year:  2012        PMID: 23015562     DOI: 10.1302/0301-620X.94B10.28858

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  13 in total

Review 1.  Mobile bearing and fixed bearing total knee arthroplasty.

Authors:  Marcello Capella; Marco Dolfin; Francesco Saccia
Journal:  Ann Transl Med       Date:  2016-04

2.  Comparison of fixed-bearing and mobile-bearing total knee arthroplasty after high tibial osteotomy.

Authors:  Philippe Hernigou; Maxime Huys; Jacques Pariat; François Roubineau; Charles Henri Flouzat Lachaniette; Arnaud Dubory
Journal:  Int Orthop       Date:  2017-06-30       Impact factor: 3.075

3.  CORR Insights®: Radiostereometric Analysis Permits In Vivo Measurement of Very Small Levels of Wear in TKA.

Authors:  Robert H Hopper
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

4.  No clinical difference between fixed- and mobile-bearing cruciate-retaining total knee arthroplasty: a prospective randomized study.

Authors:  O Bailey; K Ferguson; E Crawfurd; P James; P A May; S Brown; M Blyth; W J Leach
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-02-09       Impact factor: 4.342

5.  CORR Insights®: Rotating-platform TKA no different from fixed-bearing TKA regarding survivorship or performance: a meta-analysis.

Authors:  Tae Kyun Kim
Journal:  Clin Orthop Relat Res       Date:  2014-03-26       Impact factor: 4.176

Review 6.  No differences between fixed- and mobile-bearing total knee arthroplasty.

Authors:  B L Fransen; D C van Duijvenbode; M J M Hoozemans; B J Burger
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-06-20       Impact factor: 4.342

7.  Low implant migration of the SIGMA® medial unicompartmental knee arthroplasty.

Authors:  Daan Koppens; Maiken Stilling; Stig Munk; Jesper Dalsgaard; Søren Rytter; Ole Gade Sørensen; Torben Bæk Hansen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-11-16       Impact factor: 4.342

8.  Minimum 5-year follow-up results and functional outcome of rotating-platform high-flexion total knee arthroplasty: A prospective study of 701 knees.

Authors:  Sanjeev Jain; Aditya C Pathak; K Kalaivanan
Journal:  Arthroplast Today       Date:  2016-06-06

9.  Migration and clinical outcome of mobile-bearing versus fixed-bearing single-radius total knee arthroplasty.

Authors:  Koen T Van Hamersveld; Perla J Marang-Van De Mheen; Huub J L Van Der Heide; Henrica M J Van Der Linden-Van Der Zwaag; Edward R Valstar; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2018-02-16       Impact factor: 3.717

10.  RSA migration of total knee replacements.

Authors:  Bart G Pijls; José W M Plevier; Rob G H H Nelissen
Journal:  Acta Orthop       Date:  2018-03-06       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.