| Literature DB >> 22997510 |
Gerard D Henry1, Caroline Jennermann, J Francois Eid.
Abstract
The inflatable penile prosthesis (IPP) has high patient satisfaction rates and good mechanical reliability rates in multiple studies. The number one patient compliant at six months is penile length. Recently, new technique for aggressive sizing of the cylinders has been published on in the literature. One IPP company has produced a new product that has longer length cylinders (XL) than those available. However, traditionally long cylinders were felt to lack axial rigidity. Therefore, a prospective, multicenter, central IRB-approved, monitored study was performed on the new product to address these concerns. At 2 centers, a total of 17 patients underwent surgical implantation of these new XL cylinders. These patients were questioned for patient satisfaction and tested for axial rigidity using a Fastsize Erectile Quality Monitor. The results showed excellent patient satisfaction rates and great axial rigidity with the Fastsize Erectile Quality Monitor. The XL cylinders appear to give the IPP surgeon the ability to use longer cylinders with good patient satisfaction and great axial rigidity.Entities:
Year: 2012 PMID: 22997510 PMCID: PMC3446653 DOI: 10.1155/2012/896070
Source DB: PubMed Journal: Adv Urol ISSN: 1687-6369
Demographic data of study subjects.
| Characteristic |
| Mean ± SD | Median (Min, Max) | 95% CI |
|---|---|---|---|---|
| Age at implant | 16 | 66.8 ± 7.0 | 67.0 (53.6, 79.4) | 63.0, 70.5 |
| Height (cm) | 16 | 180.9 ± 11.9 | 184.0 (142, 191) | 174.5, 187.2 |
| Weight (kg) | 16 | 105.6 ± 14.0 | 106.5 (83, 126) | 98.2, 113.1 |
| Body mass index (kg/m2) | 16 | 32.5 ± 5.0 | 32.5 (24.8, 41.7) | 29.8, 35.2 |
Note: SD: standard deviation; CI: confidence interval.
General ED etiology and presence of peyronie's disease.
| Medical history condition |
|
|---|---|
| Diabetes mellitus | 1/16 (6%) |
| Vascular disease | 6/16 (38%) |
| Pelvic surgery | 0/16 (0%) |
| Pelvic trauma | 0/16 (0%) |
| Post-cancer treatment | 5/16 (31%) |
| Psychological causes | 0/16 (0%) |
| Other | 6/16 (38%) |
| Hypertension | 2/16 (13%) |
| Malfunctioning penile prosthesis | 3/16 (19%) |
| Severe impotence | 1/16 (6%) |
| Peyronie's Disease | 3/16 (19%) |
| Modeling conducted | 2/3 |
| Device maintained enough pressure to sufficiently model subject's anatomy | 2/3 |
Characteristics of implants assessed.
|
| |
|---|---|
| Implant approach/technique: | |
| Scrotal | 16/16 (100%) |
| Infrapubic | 0/16 (0%) |
| Right proximal measurement | |
| 10 cm | 2 (12%) |
| NA1 | 14 (88%) |
| Right distal measurement | |
| 16 cm | 2 (12%) |
| NA1 | 14 (88%) |
| Left proximal measurement | |
| 10 cm | 2 (12%) |
| NA1 | 14 (88%) |
| Left distal measurement | |
| 15 cm | 1 (6%) |
| 16 cm | 1 (6%) |
| NA1 | 14 (88%) |
| Cylinder size | |
| 24 cm | 7/16 (44%) |
| 26 cm | 7/16 (44%) |
| 28 cm | 2/16 (12%) |
| Right RTE | |
| None | 9 (56%) |
| 1 cm | 5 (31%) |
| 2 cm | 2 (12%) |
| 3 cm | 0 (0%) |
| Left RTE | |
| None | 9 (56%) |
| 1 cm | 6 (38%) |
| 2 cm | 1 (6%) |
| 3 cm | 0 (0%) |
| Reservoir size | |
| 100 cc | 15/16 (94%) |
| 130 cc | 1/16 (6%) |
Note: 1NA: not available; no result given.
Physician feedback.
|
| |
|---|---|
| Rate final result (5 = excellent, 1 = poor): | |
| 5 | 14/16 (88%) |
| 4 | 2/16 (12%) |
| 3 | 0/16 (0%) |
| 2 | 0/16 (0%) |
| 1 | 0/16 (0%) |
| What device would you have previously used? | |
| Titan | 15/16 (94%) |
| AMS CX | 1/16 (6%) |
| AMS Ultrex/LGX | 0/16 (0%) |
| AMS Ambicor | 0/16 (0%) |
| What cylinder size? | |
| 21 | 1/16 (6%) |
| 22 | 14/16 (88%) |
| 24 | 1/16 (6%) |
| What RTE size? | |
|
| 16 |
| Mean ± SD | 3.7 ± 1.4 |
| Median (Min, Max) | 3.8 (2, 7) |
| 95% CI | 2.9, 4.4 |
| Revision subject (yes) | 3/16 (19%) |
| Previous device | |
| Titan | 3 |
| AMS CX | 0 |
| AMS Ultrex/LGX | 0 |
| AMS Ambicor | 0 |
| Cylinder size | |
| 18 cm | 1 |
| 22 cm | 1 |
| RTE size | |
| 2 cm | 1 |
| 3 cm | 1 |
Rigidity measurements.
| Patient | Physician | |
|---|---|---|
| Number of pumps until perceiving full inflation | ||
|
| 16 | 16 |
| Mean ± SD | 30.4 ± 14.6 | 44.4 ± 8.3 |
| Median (Min, Max) | 28.5 (9, 58) | 44.5 (33, 62) |
| 95% CI | 22.6, 38.2 | 40.0, 48.9 |
| Rigidometer reading (g) | ||
|
| 16 | 16 |
| Mean ± SD | 1500.0 ± 342.5 | 1787.5 ± 212.5 |
| Median (Min, Max) | 1600 (800, 2000) | 1800 (1400, 2000) |
| 95% CI | 1317.5, 1682.5 | 1674.3, 1900.7 |
| Did the penis buckle? | ||
| No ( | 5/16 (31%) | 16/16 (100%) |
| Yes ( | 11/16 (69%) | 0/16 (0%) |
Satisfaction of initial implantation patients.
| How satisfied are you with the following aspects of your implant? |
| ||||
|---|---|---|---|---|---|
| Strongly agree | Agree | Neutral | Disagree | Strongly disagree | |
|
|
|
|
|
| |
| Overall function | 8 (62%) | 3 (23%) | 2 (15%) | 0 (0%) | 0 (0%) |
| Soft enough to conceal when deflated | 4 (31%) | 4 (31%) | 1 (8%) | 4 (31%) | 0 (0%) |
| Ease of deflation | 4 (31%) | 5 (38%) | 3 (23%) | 1 (8%) | 0 (0%) |
| Ease of inflation | 7 (54%) | 6 (46%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Hardness of erection when inflated | 11 (85%) | 2 (15%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Girth when inflated | 6 (46%) | 3 (23%) | 4 (31%) | 0 (0%) | 0 (0%) |
| Length when inflated | 6 (46%) | 2 (15%) | 2 (15%) | 3 (23%) | 0 (0%) |
Satisfaction of revision patients.
| Compared to your previous implant, how satisfied are you with the following aspects of your new large cylinder implant? |
| ||||
|---|---|---|---|---|---|
| Strongly agree | Agree | Neutral | Disagree | Strongly disagree | |
|
|
|
|
|
| |
| Overall function | 1 (33%) | 2 (67%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Soft enough to conceal when deflated | 2 (67%) | 1 (33%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Ease of deflation | 1 (33%) | 2 (67%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Ease of inflation | 1 (33%) | 1 (33%) | 0 (0%) | 1 (33%) | 0 (0%) |
| Hardness of erection when inflated | 1 (33%) | 2 (67%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Girth when inflated | 1 (33%) | 1 (33%) | 1 (33%) | 0 (0%) | 0 (0%) |
| Length when inflated | 1 (33%) | 1 (33%) | 1 (33%) | 0 (0%) | 0 (0%) |