Engineering and performance (NEMA and animal) of a lower-cost higher-resolution animal PET/CT scanner using photomultiplier-quadrant-sharing detectors.

Wai-Hoi Wong1, Hongdi Li, Hossain Baghaei, Yuxuan Zhang, Rocio A Ramirez, Shitao Liu, Chao Wang, Shaohui An.   

Abstract

UNLABELLED: The dedicated murine PET (MuPET) scanner is a high-resolution, high-sensitivity, and low-cost preclinical PET camera designed and manufactured at our laboratory. In this article, we report its performance according to the NU 4-2008 standards of the National Electrical Manufacturers Association (NEMA). We also report the results of additional phantom and mouse studies.
METHODS: The MuPET scanner, which is integrated with a CT camera, is based on the photomultiplier-quadrant-sharing concept and comprises 180 blocks of 13 × 13 lutetium yttrium oxyorthosilicate crystals (1.24 × 1.4 × 9.5 mm(3)) and 210 low-cost 19-mm photomultipliers. The camera has 78 detector rings, with an 11.6-cm axial field of view and a ring diameter of 16.6 cm. We measured the energy resolution, scatter fraction, sensitivity, spatial resolution, and counting rate performance of the scanner. In addition, we scanned the NEMA image-quality phantom, Micro Deluxe and Ultra-Micro Hot Spot phantoms, and 2 healthy mice.
RESULTS: The system average energy resolution was 14% at 511 keV. The average spatial resolution at the center of the field of view was about 1.2 mm, improving to 0.8 mm and remaining below 1.2 mm in the central 6-cm field of view when a resolution-recovery method was used. The absolute sensitivity of the camera was 6.38% for an energy window of 350-650 keV and a coincidence timing window of 3.4 ns. The system scatter fraction was 11.9% for the NEMA mouselike phantom and 28% for the ratlike phantom. The maximum noise-equivalent counting rate was 1,100 at 57 MBq for the mouselike phantom and 352 kcps at 65 MBq for the ratlike phantom. The 1-mm fillable rod was clearly observable using the NEMA image-quality phantom. The images of the Ultra-Micro Hot Spot phantom also showed the 1-mm hot rods. In the mouse studies, both the left and right ventricle walls were clearly observable, as were the Harderian glands.
CONCLUSION: The MuPET camera has excellent resolution, sensitivity, counting rate, and imaging performance. The data show it is a powerful scanner for preclinical animal study and pharmaceutical development.

Entities:  

Mesh:

Year:  2012        PMID: 22997277     DOI: 10.2967/jnumed.112.103507

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  5 in total

1.  A Prototype High-Resolution Small-Animal PET Scanner Dedicated to Mouse Brain Imaging.

Authors:  Yongfeng Yang; Julien Bec; Jian Zhou; Mengxi Zhang; Martin S Judenhofer; Xiaowei Bai; Kun Di; Yibao Wu; Mercedes Rodriguez; Purushottam Dokhale; Kanai S Shah; Richard Farrell; Jinyi Qi; Simon R Cherry
Journal:  J Nucl Med       Date:  2016-03-24       Impact factor: 10.057

2.  NEMA characterization of the SAFIR prototype PET insert.

Authors:  Parisa Khateri; Werner Lustermann; Christian Ritzer; Charalampos Tsoumpas; Günther Dissertori
Journal:  EJNMMI Phys       Date:  2022-06-13

3.  An Accurate Timing Alignment Method with Time-to-Digital Converter Linearity Calibration for High-Resolution TOF PET.

Authors:  Hongdi Li; Chao Wang; Shaohui An; Xingyu Lu; Yun Dong; Shitao Liu; Hossain Baghaei; Yuxuan Zhang; Rocio Ramirez; Wai-Hoi Wong
Journal:  IEEE Trans Nucl Sci       Date:  2015-06-03       Impact factor: 1.679

4.  Single-Chain Variable Fragment Antibody of Vascular Cell Adhesion Molecule 1 as a Molecular Imaging Probe for Colitis Model Rabbit Investigation.

Authors:  Chunbao Liu; Jun Zhou; Xiaojie Cheng; Liang Xia; Junfen Zhou; Shufang Xu; Yichun Wang; Yongxue Zhang; Diyu Lu
Journal:  Contrast Media Mol Imaging       Date:  2019-01-20       Impact factor: 3.161

5.  Comments on the NEMA NU 4-2008 Standard on Performance Measurement of Small Animal Positron Emission Tomographs.

Authors:  Patrick Hallen; David Schug; Volkmar Schulz
Journal:  EJNMMI Phys       Date:  2020-02-24