Echocardiographic measures of myocardial deformation by speckle-tracking technologies: the need for standardization?

Matthew R Nelson1, R Todd Hurst, Serageldin F Raslan, Stephen Cha, Susan Wilansky, Steven J Lester.   

Abstract

BACKGROUND: Multiple vendor-specific two-dimensional speckle-tracking echocardiographic algorithms with which to characterize myocardial mechanics are commercially available. The purpose of this study was to compare global longitudinal strain (GLS) results between two independent software vendors using a neutral image platform.
METHODS: A convenience sample of 100 prospectively collected patients was evaluated. Subjects with more than two left ventricular endocardial segments poorly delineated were excluded. GLS was obtained from the apical four-chamber, three-chamber, and two-chamber views using two independent speckle-tracking echocardiographic software packages (EchoInsight version 1.5.0 and Image-Arena version 4.5). Linear regression analysis and paired t tests were used to compare GLS results. Intraclass correlation coefficients and Bland-Altman plots were used for assessments of reliability.
RESULTS: The "out-of-the-box" mean GLS was -12.99 ± 2.38% using EchoInsight and -16.87 ± 2.84% using Image-Arena (mean difference, 3.87 ± 2.42%; P = .0001). Agreement between the software packages was moderate (intraclass correlation coefficient, 0.43; 95% confidence interval, 0.32-0.55). Using uniform variables to derive GLS (Lagrangian strain measured in systole and diastole at the endocardium and averaging the peak segmental strain curves), EchoInsight GLS was -16.17 ± 2.90% and Image-Arena GLS was -16.87 ± 2.84% (mean difference, 0.70 ± 2.75%; P = .02), with an intraclass correlation coefficient of 0.70 (95% confidence interval, 0.52-0.79).
CONCLUSIONS: Image-Arena GLS results were consistently different (more negative) than EchoInsight measures out of the box but became similar when information used to derive GLS was uniform. The evolution of measures of myocardial mechanics into routine clinical practice will require vigilance and standardization of the various techniques, necessitating independent validation of commercially available speckle-tracking echocardiographic products.
Copyright © 2012 American Society of Echocardiography. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22981227     DOI: 10.1016/j.echo.2012.08.006

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  17 in total

1.  Feasibility and reproducibility of systolic right ventricular strain measurement by speckle-tracking echocardiography in premature infants.

Authors:  Philip Thaler Levy; Mark R Holland; Timothy J Sekarski; Aaron Hamvas; Gautam K Singh
Journal:  J Am Soc Echocardiogr       Date:  2013-07-20       Impact factor: 5.251

Review 2.  Strain, strain rate, torsion, and twist: echocardiographic evaluation.

Authors:  Anders Opdahl; Thomas Helle-Valle; Helge Skulstad; Otto A Smiseth
Journal:  Curr Cardiol Rep       Date:  2015-03       Impact factor: 2.931

3.  Global longitudinal strain in septic cardiomyopathy: the hidden part of the iceberg?

Authors:  Philippe Vignon; Stephen J Huang
Journal:  Intensive Care Med       Date:  2015-07-17       Impact factor: 17.440

Review 4.  Left ventricular function assessment in cirrhosis: Current methods and future directions.

Authors:  Francisco Sampaio; Joana Pimenta
Journal:  World J Gastroenterol       Date:  2016-01-07       Impact factor: 5.742

5.  Reproducibility of speckle-tracking-based strain measures of left ventricular function in a community-based study.

Authors:  Susan Cheng; Martin G Larson; Elizabeth L McCabe; Ewa Osypiuk; Birgitta T Lehman; Plamen Stanchev; Jayashri Aragam; Emelia J Benjamin; Scott D Solomon; Ramachandran S Vasan
Journal:  J Am Soc Echocardiogr       Date:  2013-08-14       Impact factor: 5.251

6.  Reproducibility and Agreement of Tissue Tracking versus Feature Tracking for Strain Measurement on Cardiac MR Images in Patients with Repaired Tetralogy of Fallot.

Authors:  Jimmy C Lu; Sowmya Balasubramanian; Sunkyung Yu; Maryam Ghadimi Mahani; Prachi P Agarwal; Adam L Dorfman
Journal:  Radiol Cardiothorac Imaging       Date:  2019-04-25

7.  Advantage and validation of vendor-independent software for myocardial strain analysis compared to vendor-specific software.

Authors:  Robert Chamberlain; Kenji Shiino; Gregory M Scalia; Surendran Sabapathy; Jonathan Chan
Journal:  Australas J Ultrasound Med       Date:  2020-10-14

8.  4 Dimensional XStrain speckle tracking echocardiography: comprehensive evaluation of left ventricular strain and twist parameters in healthy Indian adults during COVID-19 pandemic.

Authors:  Akhil Mehrotra; Shubham Kacker; Mohammad Shadab; Naveen Chandra; Alok Kumar Singh
Journal:  Am J Cardiovasc Dis       Date:  2022-08-15

Review 9.  Perioperative assessment of myocardial deformation.

Authors:  Andra E Duncan; Andrej Alfirevic; Daniel I Sessler; Zoran B Popovic; James D Thomas
Journal:  Anesth Analg       Date:  2014-03       Impact factor: 5.108

10.  Unique Abnormalities in Right Ventricular Longitudinal Strain in Systemic Sclerosis Patients.

Authors:  Monica Mukherjee; Shang-En Chung; Von Khue Ton; Ryan J Tedford; Laura K Hummers; Fredrick M Wigley; Theodore P Abraham; Ami A Shah
Journal:  Circ Cardiovasc Imaging       Date:  2016-06-07       Impact factor: 7.792

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.