Lara Edbrooke1, Noel Lythgo, Unna Goldsworthy, Linda Denehy.
Abstract
PURPOSE: To investigate the validity and reliability of the Activity Monitoring Pod (AMP331) to record gait parameters in healthy young adults (YA) and intensive care unit inpatients (ICU).Entities:
Mesh:
Year: 2012 PMID: 22980236 PMCID: PMC4776921 DOI: 10.5539/gjhs.v4n3p98
Source DB: PubMed Journal: Glob J Health Sci ISSN: 1916-9736
AMP 331 and Vicon gait measures recorded during over-ground walking in healthy young adults
| Parameter | 2 Laps | 4 Laps | 6 Laps | ||||||
|---|---|---|---|---|---|---|---|---|---|
| AMP | Vicon | Mean difference | AMP | Vicon | Mean difference | AMP | Vicon | Mean difference | |
| 27.93 ± 2.28 | 28.96 | -1.02 (-2.29 to 0.24) | 56.27 ± 3.56 | 57.91 | -1.65 (-3.61 to 0.32) | 84.20 ± 9.53 | 86.87 | -1.34 (-1.30 to 3.98) | |
| 22.3 ± 1.80 | 22.60 ± 1.70 | -0.33 (-0.68 to 0.01) | 44.50 ± 4.10 | 44.70 ± 4.40 | -0.20 (-0.57 to 0.17) | 64.90 ± 5.40 | 65.10 ± 5.20 | -0.20 (-0.51 to 0.11) | |
| 116.2 ± 16.60 | 116.0 ± 19.20 | 0.20(-5.50 to 5.80) | 119.10 ± 19.40 | 118.90 ± 21.80 | 0.20 (-5.00 to 5.30) | 122.50 ± 20.90 | 123.40 ± 21.30 | -1.00 (-9.50 to 7.80) | |
| 63.20 ± 8.20 | 64.40 ± 5.20 | -0.01 (-0.04 to 0.02) | 63.50 ± 6.50 | 65.30 ± 6.50 | -0.02 (-0.04 to 0.01) | 65.20 ± 8.80 | 67.10 ± 5.40 | -0.02 (-0.06 to 0.02) | |
| 111.03 ± 10.96 | 107.69 ± 11.90 | 3.34 (0.24 to 6.43) | 112.21 ± 11.98 | 108.68 ± 12.35 | 3.52 (0.62 to 6.43) | 112.79 ± 11.41 | 109.82 ± 11.90 | 2.97 (0.00 to 5.94) | |
| 24.40 ± 3.25 | 25.55 ± 3.93 | -1.15 (-1.96 to -0.341) | 48.53 ± 9.05 | 50.30 ± 9.77 | -1.77 (-3.15 to -0.39) | 69.33 ± 11.39 | 72.26 ± 11.75 | -2.93 (-5.43 to -0.43) | |
*Data are expressed as mean ± SD and mean differences (95% confidence intervals) AMP331 – Vicon
significant differences p < 0.05
Figure 1Modified Bland Altman plot for the gait measure of distance (m) during over-ground walking in healthy young adults. Difference in distance measured (AMP331 – Vicon)
AMP 331 and Vicon gait measures recorded during treadmill walking in healthy young adults
| Parameter | 2.4 km·h-1 | 3.2 km·h-1 | 4.2 km·h-1 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| AMP | Vicon | Mean difference | AMP | Vicon | Mean difference | AMP | Vicon | Mean difference | |
| 88.73 ±39.79 | 117.81 ± 0.68 | -29.09 | 154.27 ± 12.70 | 162.77 ± 1.23 | -8.50 (-15.66 to -1.58) | 188.93 ± 18.85 | 212.97 ± 1.51 | -24.03 | |
| 87.00 ± 40.22 | 114.13 ± 9.77 | -27.13 | 139.3 ± 11.7 | 140.7 ± 11.4 | -1.33 | 159.33 ± 9.96 | 160.40 ± 11.13 | -1.07 (-2.68 to 0.55) | |
| 65.60 ± 7.76 | 65.59 ± 0.1 | -0.01 (-4.32 to 4.30) | 85.93 ± 6.8 | 90.16 ± 0.2 | -4.07 (-7.85 to -0.02) | 104.87 ± 10.06 | 118.02 ± 0.14 | -13.13 | |
| 52.00 ± 7.97 | 52.04 ± 4.31 | -0.05 (-2.97 to 2.88) | 55.93± 7.36 | 57.94 ± 4.79 | -2.00 (-4.46 to 0.46) | 59.60± 7.50 | 66.54 ± 4.58 | -6.93 | |
| 76.57 ± 6.15 | 76.09 ± 6.51 | 0.48 (-0.58 to 1.53) | 93.12 ± 7.44 | 93.78 ± 7.59 | -0.66 | 106.11 ± 6.74 | 106.93 ± 7.42 | -0.83 (-1.96 to 0.31) | |
| 134.73 ± 59.27 | 179.59 ± 1.04 | -44.86 | 179.60 ± 1.92 | 180.58 ± 1.36 | -0.98 (-2.21 to 0.25) | 180.20 ± 1.66 | 180.48 ± 1.29 | -0.28 (-1.58 to 1.02) | |
| 251.93 ±22.65 | 289.23 ± 2.14 | -37.49† (-49.84 to -25.14) | 178.20 ± 46.53 | 212.66 ± 1.12 | -34.46 | ||||
| 183.73 ± 9.32 | 183.73 ± 9.22 | 0.00 (-0.47 to 0.47) | 155.20 ± 10.41 | 155.40 ± 10.68 | -0.20 (-0.63 to 0.23) | ||||
| 139.13 ±11.99 | 160.29 ± 0.11 | -20.87 | 105.47 ± 8.44 | 118.36 ± 0.14 | -12.67 | ||||
| 68.87 ± 7.82 | 78.70 ± 4.00 | -9.73 | 61.53 ± 6.83 | 64.14 ± 4.33 | -2.47 (-5.34 to 0.40) | ||||
| 121.70 ± 6.22 | 122.49 ± 6.14 | -0.79 (-1.46 to -0.11) | 103.27 ± 7.23 | 103.60 ± 7.12 | -0.33 (-0.65 to -0.01) | ||||
| 181.20 ± 1.74 | 180.56 ± 1.34 | 0.64 (-0.52 to 1.81) | 180.33 ± 1.59 | 180.22 ± 0.95 | 0.16 (-1.15 to 1.38) | ||||
*Data are expressed as mean ± SD and mean difference (95% confidence intervals) AMP331 – Vicon
significant differences p < 0.05
Difference in gait measures recorded in ICU survivors between the AMP 331 and manually measured parameters
| Parameter | Walk 1 Mean (95%CI) | Walk 2 Mean (95%CI) |
|---|---|---|
| -2.79 (-3.66 to -1.92) | -3.11 (-3.99 to -2.22) | |
| 0.93 (0.39 to 1.48) | 0.92 (0.44 to 1.40) | |
| -21.94 (-29.32 to -14.55) | -28.87 (-34.94 to -22.80) |
There was a consistent underestimation of distance by the AMP331 at all distances during walks one and two (walk 2 mean difference = -3.11m, 95% agreement of – 10.84 to 4.62m).
Figure 2Modified Bland Altman plot for the gait measure of distance (m) during over-ground walking in ICU survivors (walk 2). Difference in distance measured (AMP331 – measured distance)