Reference standard and statistical model for intersite and temporal comparisons of CT attenuation in a multicenter quantitative lung study.

J P Sieren1, J D Newell, P F Judy, D A Lynch, K S Chan, J Guo, E A Hoffman.   

Abstract

PURPOSE: The purpose of this study was to detect and analyze anomalies between a large number of computed tomography (CT) scanners, tracked over time, utilized to collect human pulmonary CT data for a national multicenter study: chronic obstructive pulmonary disease genetic epidemiology study (COPDGene).
METHODS: A custom designed CT reference standard "Test Object" has been developed to evaluate the relevant differences in CT attenuation between CT scanners in COPDGene. The materials used in the Test Object to assess CT scanner accuracy and precision included lung equivalent foam (-856 HU), internal air (-1000 HU), water (0 HU), and acrylic (120 HU). Nineteen examples of the Test Object were manufactured. Initially, all Test Objects were scanned on the same CT scanner before the Test Objects were sent to the 20 specific sites and 42 individual CT scanners that were used in the study. The Test Objects were scanned over 17 months while the COPDGene study continued to recruit subjects. A mixed linear effect statistical analysis of the CT scans on the 19 Test Objects was performed. The statistical model reflected influence of reconstruction kernels, tube current, individual Test Objects, CT scanner models, and temporal consistency on CT attenuation.
RESULTS: Depending on the Test Object material, there were significant differences between reconstruction kernels, tube current, individual Test Objects, CT scanner models, and temporal consistency. The two Test Object materials of most interest were lung equivalent foam and internal air. With lung equivalent foam, there were significant (p < 0.05) differences between the Siemens B31 (-856.6, ±0.82; mean ± SE) and the GE Standard (-856.6 ± 0.83) reconstruction kernel relative to the Siemens B35 reference standard (-852.5 ± 1.4). Comparing lung equivalent foam attenuation there were also significant differences between CT scanner models (p < 0.01), tube current (p < 0.005), and in temporal consistency (p < 0.005) at individual sites. However, there were no significant effects measurable using different examples of the Test Objects at the various sites compared to the reference scans of the 19 Test Objects. For internal air, significant (p < 0.005) differences were found between all reconstruction kernels (Siemens B31, GE Standard, and Phillips B) compared to the reference standard. There were significant differences between CT models (p < 0.005), and tube current (p < 0.005). There were no significant effects measurable using different examples of the Test Objects at the various sites compared to the reference scans of the 19 Test Objects. Differences, across scanners, between external air and internal air measures in this simple (relative to the in vivo lung) test object varied by as much as 15 HU.
CONCLUSIONS: The authors conclude that the Test Object designed for this study was able to detect significant effects regarding individual CT scanners that altered the CT attenuation measurements relevant to the study that are used to determine lung density. Through an understanding of individual scanners, the Test Object analysis can be used to detect anomalies in an individual CT scanner and to statistically model out scanner differences and individual scanner changes over time in a large multicenter trial.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22957640      PMCID: PMC3448623          DOI: 10.1118/1.4747342

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Comparison of the sensitivities of 5 different computed tomography scanners for the assessment of the progression of pulmonary emphysema: a phantom study.

Authors:  Berend C Stoel; M Els Bakker; Jan Stolk; Asger Dirksen; Robert A Stockley; Eeva Piitulainen; Erich W Russi; Johan H C Reiber
Journal:  Invest Radiol       Date:  2004-01       Impact factor: 6.016

2.  Heterogeneity of pulmonary perfusion as a mechanistic image-based phenotype in emphysema susceptible smokers.

Authors:  Sara K Alford; Edwin J R van Beek; Geoffrey McLennan; Eric A Hoffman
Journal:  Proc Natl Acad Sci U S A       Date:  2010-04-05       Impact factor: 11.205

3.  The phantom portion of the American College of Radiology (ACR) computed tomography (CT) accreditation program: practical tips, artifact examples, and pitfalls to avoid.

Authors:  Cynthia H McCollough; Michael R Bruesewitz; Michael F McNitt-Gray; Krista Bush; Thomas Ruckdeschel; J Thomas Payne; James A Brink; Robert K Zeman
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

4.  An electron density calibration phantom for CT-based treatment planning computers.

Authors:  C Constantinou; J C Harrington; L A DeWerd
Journal:  Med Phys       Date:  1992 Mar-Apr       Impact factor: 4.071

Review 5.  State of the Art. A structural and functional assessment of the lung via multidetector-row computed tomography: phenotyping chronic obstructive pulmonary disease.

Authors:  Eric A Hoffman; Brett A Simon; Geoffrey McLennan
Journal:  Proc Am Thorac Soc       Date:  2006-08

6.  Airway and parenchymal disease in chronic obstructive pulmonary disease are distinct phenotypes.

Authors:  Bip Patel; Barry Make; Harvey O Coxson; Nestor L Muller; Sreekumar Pillai; Wayne Anderson; Edwin Silverman; David Lomas
Journal:  Proc Am Thorac Soc       Date:  2006-08

7.  The CT's sample volume as an approximate, instrumental measure for density resolution in densitometry of the lung.

Authors:  G J Kemerink; H H Kruize; R J Lamers
Journal:  Med Phys       Date:  1997-10       Impact factor: 4.071

8.  Development of radiographic chest phantoms.

Authors:  C Constantinou; J Cameron; L DeWerd; M Liss
Journal:  Med Phys       Date:  1986 Nov-Dec       Impact factor: 4.071

9.  Deaths: final data for 1999.

Authors:  D L Hoyert; E Arias; B L Smith; S L Murphy; K D Kochanek
Journal:  Natl Vital Stat Rep       Date:  2001-09-21

10.  Density resolution in quantitative computed tomography of foam and lung.

Authors:  G J Kemerink; H H Kruize; R J Lamers; J M van Engelshoven
Journal:  Med Phys       Date:  1996-10       Impact factor: 4.071

View more
  29 in total

1.  Relationships between airflow obstruction and quantitative CT measurements of emphysema, air trapping, and airways in subjects with and without chronic obstructive pulmonary disease.

Authors:  Joyce D Schroeder; Alexander S McKenzie; Jordan A Zach; Carla G Wilson; Douglas Curran-Everett; Douglas S Stinson; John D Newell; David A Lynch
Journal:  AJR Am J Roentgenol       Date:  2013-09       Impact factor: 3.959

2.  Effect of Reducing Field of View on Multidetector Quantitative Computed Tomography Parameters of Airway Wall Thickness in Asthma.

Authors:  Ajay Sheshadri; Alfonso Rodriguez; Ryan Chen; James Kozlowski; Dana Burgdorf; Tammy Koch; Jaime Tarsi; Rebecca Schutz; Brad Wilson; Kenneth Schechtman; Joseph K Leader; Eric A Hoffman; Mario Castro; Sean B Fain; David S Gierada
Journal:  J Comput Assist Tomogr       Date:  2015 Jul-Aug       Impact factor: 1.826

3.  Variation in the percent of emphysema-like lung in a healthy, nonsmoking multiethnic sample. The MESA lung study.

Authors:  Eric A Hoffman; Firas S Ahmed; Heather Baumhauer; Mathew Budoff; J Jeffrey Carr; Richard Kronmal; S Reddy; R Graham Barr
Journal:  Ann Am Thorac Soc       Date:  2014-07

4.  Lung density changes with growth and inflation.

Authors:  H Brown Robert; A Wise Robert; Gregory Kirk; M Bradley Drummond; Wayne Mitzner
Journal:  Chest       Date:  2015-10       Impact factor: 9.410

5.  Standardizing CT lung density measure across scanner manufacturers.

Authors:  Huaiyu Heather Chen-Mayer; Matthew K Fuld; Bernice Hoppel; Philip F Judy; Jered P Sieren; Junfeng Guo; David A Lynch; Antonio Possolo; Sean B Fain
Journal:  Med Phys       Date:  2017-02-21       Impact factor: 4.071

6.  Improved CT-based estimate of pulmonary gas trapping accounting for scanner and lung-volume variations in a multicenter asthmatic study.

Authors:  Sanghun Choi; Eric A Hoffman; Sally E Wenzel; Mario Castro; Ching-Long Lin
Journal:  J Appl Physiol (1985)       Date:  2014-08-07

7.  Sinogram Affirmed Iterative Reconstruction (SAFIRE) versus weighted filtered back projection (WFBP) effects on quantitative measure in the COPDGene 2 test object.

Authors:  J P Sieren; E A Hoffman; M K Fuld; K S Chan; J Guo; J D Newell
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

8.  Effect of inspiration on airway dimensions measured in maximal inspiration CT images of subjects without airflow limitation.

Authors:  Jens Petersen; Mathilde M W Wille; Lars Lau Rakêt; Aasa Feragen; Jesper H Pedersen; Mads Nielsen; Asger Dirksen; Marleen de Bruijne
Journal:  Eur Radiol       Date:  2014-06-06       Impact factor: 5.315

Review 9.  Advances in Imaging and Automated Quantification of Pulmonary Diseases in Non-neoplastic Diseases.

Authors:  Fernanda Balbinot; Álvaro da Costa Batista Guedes; Douglas Zaione Nascimento; Juliana Fischman Zampieri; Giordano Rafael Tronco Alves; Edson Marchiori; Adalberto Sperb Rubin; Bruno Hochhegger
Journal:  Lung       Date:  2016-09-23       Impact factor: 2.584

10.  SPIROMICS Protocol for Multicenter Quantitative Computed Tomography to Phenotype the Lungs.

Authors:  Jered P Sieren; John D Newell; R Graham Barr; Eugene R Bleecker; Nathan Burnette; Elizabeth E Carretta; David Couper; Jonathan Goldin; Junfeng Guo; MeiLan K Han; Nadia N Hansel; Richard E Kanner; Ella A Kazerooni; Fernando J Martinez; Stephen Rennard; Prescott G Woodruff; Eric A Hoffman
Journal:  Am J Respir Crit Care Med       Date:  2016-10-01       Impact factor: 21.405

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.