Quantitative SPECT/CT reconstruction for ¹⁷⁷Lu and ¹⁷⁷Lu/⁹⁰Y targeted radionuclide therapies.

S Shcherbinin1, H Piwowarska-Bilska, A Celler, B Birkenfeld.   

Abstract

We investigated the quantitative accuracy of SPECT/CT imaging studies as would be performed before and after targeted radionuclide therapy (TRT) using phantom experiments with (i) (99m)Tc, (ii) ¹⁷⁷Lu and (iii) ⁹⁰Y/¹⁷⁷Lu. While the experiment with (99m)Tc imitated a diagnostic scan, the experiments with ¹⁷⁷Lu and ⁹⁰Y/¹⁷⁷Lu modeled post-therapy acquisitions. At the next stage, we reconstructed images from pre- and post-therapy patient studies. The data were first reconstructed using two methods with limited corrections for the physics effects. Then, to generate quantitatively accurate absolute activity distributions, we applied a hybrid (model-based and window-based) reconstruction strategy where some of the physics effects were accurately modeled while corrections for other effects were empirical and based on information obtained from the projection data. The accuracies of absolute activity recovered by the hybrid method from the six phantom experiments were very similar to each other and acceptable for potential use in TRT. When measured in identical regions of interest, the (99m)Tc ⁹⁰activity was reconstructed with errors ranging between -3.3% and 2.9%, while the ¹⁷⁷Lu activity was reconstructed from experiments with ¹⁷⁷Lu and Y/¹⁷⁷Lu with errors ranging between -1.6% and 1.6%. The reconstruction algorithms with limited corrections led to larger and case-specific errors as might have been expected. From a clinical prospective, our results showed that physics-based reconstructions improved resolution of images corresponding to both diagnostic scans with (99m)Tc and post-therapy scans with ¹⁷⁷Lu. Our analysis of patient study demonstrated that lack of corrections led to overestimation of activities in organs and tumor by 29-39% for the diagnostic scan with (99m)Tc and by 105-218% for post-therapy scan with ¹⁷⁷Lu.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22948135     DOI: 10.1088/0031-9155/57/18/5733

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Quantitative SPECT/CT Imaging of (177)Lu with In Vivo Validation in Patients Undergoing Peptide Receptor Radionuclide Therapy.

Authors:  J C Sanders; T Kuwert; J Hornegger; P Ritt
Journal:  Mol Imaging Biol       Date:  2015-08       Impact factor: 3.488

Review 2.  Preclinical Voxel-Based Dosimetry in Theranostics: a Review.

Authors:  Arun Gupta; Min Sun Lee; Joong Hyun Kim; Dong Soo Lee; Jae Sung Lee
Journal:  Nucl Med Mol Imaging       Date:  2020-04-19

3.  Quantitative accuracy of 177Lu SPECT imaging for molecular radiotherapy.

Authors:  Emilio Mezzenga; Vincenzo D'Errico; Marco D'Arienzo; Lidia Strigari; Koutla Panagiota; Federica Matteucci; Stefano Severi; Giovanni Paganelli; Andrew Fenwick; David Bianchini; Francesco Marcocci; Anna Sarnelli
Journal:  PLoS One       Date:  2017-08-14       Impact factor: 3.240

4.  Accuracy of 177Lu activity quantification in SPECT imaging: a phantom study.

Authors:  Carlos F Uribe; Pedro L Esquinas; Jesse Tanguay; Marjorie Gonzalez; Emilie Gaudin; Jean-Mathieu Beauregard; Anna Celler
Journal:  EJNMMI Phys       Date:  2017-01-07

5.  The effect of calibration factors and recovery coefficients on 177Lu SPECT activity quantification accuracy: a Monte Carlo study.

Authors:  Keamogetswe Ramonaheng; Johannes A van Staden; Hanlie du Raan
Journal:  EJNMMI Phys       Date:  2021-03-18

6.  Quantitative accuracy of (177)Lu SPECT reconstruction using different compensation methods: phantom and patient studies.

Authors:  Eero Hippeläinen; Mikko Tenhunen; Hanna Mäenpää; Antti Sohlberg
Journal:  EJNMMI Res       Date:  2016-02-18       Impact factor: 3.138

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.