Literature DB >> 22942269

Minilaparotomy salvage of malfunctioning catheters in peritoneal dialysis.

Jian-Ri Li1, Chi-Hung Cheng, Kun-Yuan Chiu, Chen-Li Cheng, Chi-Rei Yang, Hao-Chung Ho, Jiunn-Liang Ko, Yen-Chuan Ou.   

Abstract

BACKGROUND: Catheter malfunction is a common and significant complication during peritoneal dialysis (PD). We developed a minilaparotomy procedure to rescue malfunctioning catheters and to prevent recurrence of malfunction.
METHODS: From 2006 to 2011, 11 patients receiving PD had a malfunctioning catheter. In all patients, a 2-cm incision, 5 cm caudally to the previous peritoneal entry site was used to correct the malfunctioning catheter, with concomitant fixation of the catheter to the peritoneum.
RESULTS: Catheter tip migration occurred in 7 patients, omental wrapping in 3, and blood clot obstruction in 1. The mean onset time to catheter malfunction was 197.5 days (range: 4 - 1270 days), and the mean operating time was 41 minutes (range: 35 - 56 minutes). There was no recurrence of catheter malfunction and no surgery-related peritonitis.
CONCLUSIONS: Our minilaparotomy procedure is safe and feasible for the salvage of malfunctioning PD catheters.

Entities:  

Mesh:

Year:  2012        PMID: 22942269      PMCID: PMC3598265          DOI: 10.3747/pdi.2011.00237

Source DB:  PubMed          Journal:  Perit Dial Int        ISSN: 0896-8608            Impact factor:   1.756


  12 in total

1.  Fluoroscopically-guided manipulation of malfunctioning peritoneal dialysis catheters.

Authors:  M E Simons; G Pron; M Voros; L C Vanderburgh; P S Rao; D G Oreopoulos
Journal:  Perit Dial Int       Date:  1999 Nov-Dec       Impact factor: 1.756

2.  Videolaparoscopy as rescue therapy and placement of peritoneal dialysis catheters: a thirty-two case single centre experience.

Authors:  Stefano Santarelli; Matthias Zeiler; Rita Marinelli; Tania Monteburini; Antonio Federico; Emilio Ceraudo
Journal:  Nephrol Dial Transplant       Date:  2006-01-18       Impact factor: 5.992

3.  Correction of displaced peritoneal dialysis catheters with an angular stiff rod.

Authors:  U Ozyer; A Harman; C Aytekin; F Boyvat; N Ozdemir
Journal:  Acta Radiol       Date:  2009-03       Impact factor: 1.990

4.  Minilaparotomy with manual correction for malfunctioning peritoneal dialysis catheters.

Authors:  Sung-Ho Kim; Duk-Hyun Lee; Hee-Jeong Choi; Hye-Jin Seo; Ye-Soo Jang; Dong-Hyun Kim; Jong-Hoon Park; Chan-Duck Kim; Yong-Lim Kim
Journal:  Perit Dial Int       Date:  2008 Sep-Oct       Impact factor: 1.756

5.  Complications after tenckhoff catheter insertion: a single-centre experience using multiple operators over four years.

Authors:  Wen Jiun Liu; Lai Seong Hooi
Journal:  Perit Dial Int       Date:  2010-03-12       Impact factor: 1.756

6.  The use of a channel-cleaning brush for malfunctioning Tenckhoff catheters.

Authors:  M J Kumwenda; F K Wright
Journal:  Nephrol Dial Transplant       Date:  1999-05       Impact factor: 5.992

7.  Role of Fogarty catheter manipulation in management of migrated, nonfunctional peritoneal dialysis catheters.

Authors:  M F Gadallah; N Arora; R Arumugam; K Moles
Journal:  Am J Kidney Dis       Date:  2000-02       Impact factor: 8.860

8.  Mini-laparotomy implantation of peritoneal dialysis catheters: outcome and rescue.

Authors:  Po-Jen Yang; Chih-Yuan Lee; Chi-Chuan Yeh; Hsiao-Ching Nien; Tun-Jun Tsai; Meng-Kung Tsai
Journal:  Perit Dial Int       Date:  2010-02-26       Impact factor: 1.756

9.  Omental folding: a novel laparoscopic technique for salvaging peritoneal dialysis catheters.

Authors:  Y Heng Goh
Journal:  Perit Dial Int       Date:  2008 Nov-Dec       Impact factor: 1.756

Review 10.  Malfunctioning peritoneal dialysis catheter and accompanying surgical pathology repaired by laparoscopic surgery.

Authors:  Güner Ogünç
Journal:  Perit Dial Int       Date:  2002 Jul-Aug       Impact factor: 1.756

View more
  1 in total

1.  Identification of the appropriate fixation site to avoid peritoneal catheter migration based on a mechanical analysis.

Authors:  Yujuan Wang; Yao Zou; Xinghua Chen; Jili Zhu; Cuizhi Xiang; Houjun Jia; Guohua Ding; Huiming Wang
Journal:  Ren Fail       Date:  2017-11       Impact factor: 2.606

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.