Literature DB >> 10641775

Fluoroscopically-guided manipulation of malfunctioning peritoneal dialysis catheters.

M E Simons1, G Pron, M Voros, L C Vanderburgh, P S Rao, D G Oreopoulos.   

Abstract

OBJECTIVE: To review our experience with fluoroscopic evaluation and manipulation of malpositioned, malfunctioning, peritoneal dialysis (PD) catheters.
MATERIALS AND METHODS: Thirty-one patients, over a 5-year period (1 May 1992 to 30 April 1997) with malfunctioning PD catheters, who had fluoroscopically-guided manipulation were reviewed. Catheters were manipulated using a malleable aluminum bar and, if necessary, guide wires or other stiffeners. Technical success was assessed on the basis of adequate, fluoroscopically verified, catheter placement at the time of the procedure and improved flows. A functional PD catheter at 30 days post manipulation was considered to be a clinically successful manipulation.
RESULTS: There were 41 manipulations [33 initial (IM) and 8 remanipulations (RM)] for malpositioned or kinked catheters. In 31 (19 male, 12 female) patients ranging in age from 31 to 76 years (mean age 60 years), the initial technical success rate was 85% for IM (n = 28/33) and 63% (n = 5/8) for RM. The overall clinical success rate, or 30-day primary patency, was 55% for IM (n = 18/33) and 63% for RM. Catheter function (combined IM and RM) continued for a median 869 days (95% CI: 118, 1620). No early complications were noted.
CONCLUSIONS: Fluoroscopic manipulation, including attempts at remanipulation, of PD catheters is a safe procedure. The technique is a simple, inexpensive, and effective way of prolonging PD catheter life, thereby reducing the number of surgical interventions.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10641775

Source DB:  PubMed          Journal:  Perit Dial Int        ISSN: 0896-8608            Impact factor:   1.756


  6 in total

1.  Peritoneal dialysis catheter malfunction because of encasement by an extraluminal fibrin sheath.

Authors:  S K S Singh; A Common; J Perl
Journal:  Perit Dial Int       Date:  2012 Mar-Apr       Impact factor: 1.756

2.  Fluoroscopic manipulation of peritoneal dialysis catheters: outcomes and factors associated with successful manipulation.

Authors:  Matthew Miller; Brendan McCormick; Susan Lavoie; Mohan Biyani; Deborah Zimmerman
Journal:  Clin J Am Soc Nephrol       Date:  2012-02-23       Impact factor: 8.237

3.  Minilaparotomy salvage of malfunctioning catheters in peritoneal dialysis.

Authors:  Jian-Ri Li; Chi-Hung Cheng; Kun-Yuan Chiu; Chen-Li Cheng; Chi-Rei Yang; Hao-Chung Ho; Jiunn-Liang Ko; Yen-Chuan Ou
Journal:  Perit Dial Int       Date:  2012-09-01       Impact factor: 1.756

4.  Guidelines for laparoscopic peritoneal dialysis access surgery.

Authors:  Stephen Haggerty; Scott Roth; Danielle Walsh; Dimitrios Stefanidis; Raymond Price; Robert D Fanelli; Todd Penner; William Richardson
Journal:  Surg Endosc       Date:  2014-10-08       Impact factor: 4.584

5.  Efficacy and safety of fluoroscopic manipulation using the alpha-replacer for peritoneal catheter malposition.

Authors:  Yosuke Saka; Yasuhiko Ito; Yoshiyasu Iida; Shoichi Maruyama; Seiichi Matsuo
Journal:  Clin Exp Nephrol       Date:  2014-07-17       Impact factor: 2.801

6.  Minimally invasive fluoroscopic percutaneous peritoneal dialysis catheter salvage.

Authors:  Rajeev Narayan; Terrance Fried; Gerardo Chica; Mathew Schaefer; Daniel Mullins
Journal:  Clin Kidney J       Date:  2014-06
  6 in total