Joseph M Zabramski1, Mark C Preul, Josef Debbins, Daniel J McCusker.
Abstract
BACKGROUND: Exposure of externally programmable shunt-valves (EPS-valves) to magnetic resonance imaging (MRI) may lead to unexpected changes in shunt settings, or affect the ability to reprogram the valve. We undertook this study to examine the effect of exposure to a 3T MRI on a group of widely used EPS-valves.Entities:
Keywords: 3-Tesla; Magnetic resonance imaging; Testing; programmable; reliability; shunt-valve
Year: 2012 PMID: 22937481 PMCID: PMC3424682 DOI: 10.4103/2152-7806.99171
Source DB: PubMed Journal: Surg Neurol Int ISSN: 2152-7806
List of externally programmable shunt-valves
Figure 1Photograph of fluid filled phantom demonstrating the locations of the left-side of the phantom used for placement of the shunt valves. These same positions were used on the right-side of the phantom [Used with permission from Barrow Neurological Institute]
Figure 2Photograph demonstrating the imaging position used for the phantom whether the head coil was active or not [Used with permission from Barrow Neurological Institute]
Details of the imaging parameters used in Part II of the protocol
Figure 3Photograph demonstrating the artifact testing phantom with a shunt valve suspended in the center of the chamber for evaluation. During testing, the phantom is filled with gadolinium-doped saline [Used with permission from Barrow Neurological Institute]
Figure 4Sample image captured during evaluation for artifact testing; coronal T1-weighted, spin echo pulse sequence. Image locations for measurement were selected to represent the largest, or worse-case, artifacts for each valve [Used with permission from Barrow Neurological Institute]
Results of exposure of Codman Medos® valves to 3T magnetic field simulating four consecutive magnetic resonance imaging scans
Exposure of Sophysa Polaris® Valves to 3-Tesla field and RF coil energy simulating two consecutive magnetic resonance imaging scans (without and with contrast)
Results of exposure of Medtronic Strata® valves to 3-Tesla magnetic field simulating four consecutive magnetic resonance imaging scans
Exposure of Codman Medos® Valves to 3-Tesla magnetic field and RF coil energy simulating two consecutive magnetic resonance imaging scans (without and with contrast)
Exposure of Medtronic Strata® Valves to 3-Tesla magnetic field and RF coil energy simulating two consecutive magnetic resonance imaging scans (without and with contrast)
Results of exposure of Codman Certas® valves to 3-Tesla magnetic field simulating four consecutive magnetic resonance imaging scans
Results of exposure of Sophysa Polaris® valves to 3-Tesla magnetic field simulating 4 consecutive magnetic resonance imaging scans
Exposure of Codman Certas® valves to 3-Tesla magnetic field and RF coil energy simulating two consecutive magnetic resonance imaging scans (without and with contrast)
Maximal area of metallic artifact (cm2)
Results of exposure of Meithke proGAV® (PG) shunt valve to 3-Tesla magnetic field of simulating four consecutive magnetic resonance imaging scans
Exposure of Meithke proGAV® valves to 3-Tesla magnetic field and RF coil energy simulating two consecutive magnetic resonance imaging scans (without and with contrast)