Adjustable shunt valve-induced magnetic resonance imaging artifact: a comparative study.

Ahmed K Toma1, Andrew Tarnaris, Joan P Grieve, Laurence D Watkins, Neil D Kitchen.   

Abstract

OBJECT: In this paper, the authors' goal was to compare the artifact induced by implanted (in vivo) adjustable shunt valves in spin echo, diffusion weighted (DW), and gradient echo MR imaging pulse sequences.
METHODS: The MR images obtained in 8 patients with proGAV and 6 patients with Strata II adjustable shunt valves were assessed for artifact areas in different planes as well as the total volume for different pulse sequences.
RESULTS: Artifacts induced by the Strata II valve were significantly larger than those induced by proGAV valve in spin echo MR imaging pulse sequence (29,761 vs 2450 mm(3) on T2-weighted fast spin echo, p = 0.003) and DW images (100,138 vs 38,955 mm(3), p = 0.025). Artifacts were more marked on DW MR images than on spin echo pulse sequence for both valve types.
CONCLUSIONS: Adjustable valve-induced artifacts can conceal brain pathology on MR images. This should influence the choice of valve implantation site and the type of valve used. The effect of artifacts on DW images should be highlighted pending the development of less MR imaging artifact-inducing adjustable shunt valves.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19817540     DOI: 10.3171/2009.9.JNS09171

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  3 in total

1.  The Contribution of Common Surgically Implanted Hardware to Functional MR Imaging Artifacts.

Authors:  A A Desai; M K Strother; C C Faraco; V L Morgan; T R Ladner; L M Dethrage; L C Jordan; M J Donahue
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

2.  Valve exchange towards an adjustable differential pressure valve with gravitational unit, clinical outcome of a single-center study.

Authors:  S Alavi; M Schulz; A Schaumann; K Schwarz; Ulrich W Thomale
Journal:  Childs Nerv Syst       Date:  2017-03-22       Impact factor: 1.475

3.  3T magnetic resonance imaging testing of externally programmable shunt valves.

Authors:  Joseph M Zabramski; Mark C Preul; Josef Debbins; Daniel J McCusker
Journal:  Surg Neurol Int       Date:  2012-07-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.