Accuracy and reliability of fetal heart rate monitoring using maternal abdominal surface electrodes.

Wayne R Cohen1, Sophia Ommani, Sarmina Hassan, Fadi G Mirza, Molham Solomon, Raymond Brown, Barry S Schifrin, John M Himsworth, Barrie R Hayes-Gill.   

Abstract

OBJECTIVE: Compare the accuracy and reliability of fetal heart rate identification from maternal abdominal fetal electrocardiogram signals (ECG) and Doppler ultrasound with a fetal scalp electrode.
DESIGN: Prospective open method equivalence study.
SETTING: Three urban teaching hospitals in the Northeast United States. SAMPLE: 75 women with normal pregnancies in labor at >37 weeks of gestation.
METHODS: Three fetal heart rate detection methods were used simultaneously in 75 parturients. The fetal scalp electrode was the standard against which abdominal fetal ECG and ultrasound were judged. MAIN OUTCOME MEASURES: The positive percent agreement with the fetal scalp electrode indicated reliability. Bland-Altman analysis determined accuracy. The confusion rate indicated how frequently the devices tracked the maternal heart rate.
RESULTS: Positive percent agreement was 81.7 and 73% for the abdominal fetal ECG and ultrasound, respectively (p = 0.002). The abdominal fetal ECG had a lower root mean square error than ultrasound (5.2 vs. 10.6 bpm, p < 0.001). The confusion rate for ultrasound was 20-fold higher than for abdominal ECG (8.9 vs. 0.4%, respectively, p < 0.001).
CONCLUSION: Compared with the fetal scalp electrode, fetal heart rate detection using abdominal fetal ECG was more reliable and accurate than ultrasound, and abdominal fetal ECG was less likely than ultrasound to display the maternal heart rate in place of the fetal heart rate.
© 2012 The Authors Acta Obstetricia et Gynecologica Scandinavica© 2012 Nordic Federation of Societies of Obstetrics and Gynecology.

Entities:  

Mesh:

Year:  2012        PMID: 22924738     DOI: 10.1111/j.1600-0412.2012.01533.x

Source DB:  PubMed          Journal:  Acta Obstet Gynecol Scand        ISSN: 0001-6349            Impact factor:   3.636


  21 in total

1.  STUDIES IN FETAL BEHAVIOR: REVISITED, RENEWED, AND REIMAGINED.

Authors:  Janet A DiPietro; Kathleen A Costigan; Kristin M Voegtline
Journal:  Monogr Soc Res Child Dev       Date:  2015-09

2.  An open source autocorrelation-based method for fetal heart rate estimation from one-dimensional Doppler ultrasound.

Authors:  Camilo E Valderrama; Lisa Stroux; Nasim Katebi; Elianna Paljug; Rachel Hall-Clifford; Peter Rohloff; Faezeh Marzbanrad; Gari D Clifford
Journal:  Physiol Meas       Date:  2019-02-26       Impact factor: 2.833

3.  Computer-based intrapartum fetal monitoring and beyond: A review of the 2nd Workshop on Signal Processing and Monitoring in Labor (October 2017, Oxford, UK).

Authors:  Antoniya Georgieva; Patrice Abry; Václav Chudáček; Petar M Djurić; Martin G Frasch; René Kok; Christopher A Lear; Sebastiaan N Lemmens; Inês Nunes; Aris T Papageorghiou; Gerald J Quirk; Christopher W G Redman; Barry Schifrin; Jiri Spilka; Austin Ugwumadu; Rik Vullings
Journal:  Acta Obstet Gynecol Scand       Date:  2019-06-18       Impact factor: 3.636

4.  Influence of electrode placement on signal quality for ambulatory pregnancy monitoring.

Authors:  M J Rooijakkers; S Song; C Rabotti; S G Oei; J W M Bergmans; E Cantatore; M Mischi
Journal:  Comput Math Methods Med       Date:  2014-02-03       Impact factor: 2.238

5.  Circadian changes and sex-related differences in fetal heart rate parameters.

Authors:  Habiba Kapaya; Fiona Broughton Pipkin; Barrie Hayes-Gill; Pamela V Loughna
Journal:  Matern Health Neonatol Perinatol       Date:  2016-09-02

6.  Linear Phase Sharp Transition BPF to Detect Noninvasive Maternal and Fetal Heart Rate.

Authors:  Niyan Marchon; Gourish Naik; K R Pai
Journal:  J Healthc Eng       Date:  2018-03-29       Impact factor: 2.682

7.  Non-Invasive Fetal Monitoring: A Maternal Surface ECG Electrode Placement-Based Novel Approach for Optimization of Adaptive Filter Control Parameters Using the LMS and RLS Algorithms.

Authors:  Radek Martinek; Radana Kahankova; Homer Nazeran; Jaromir Konecny; Janusz Jezewski; Petr Janku; Petr Bilik; Jan Zidek; Jan Nedoma; Marcel Fajkus
Journal:  Sensors (Basel)       Date:  2017-05-19       Impact factor: 3.576

Review 8.  Fetal ECG extraction from abdominal signals: a review on suppression of fundamental power line interference component and its harmonics.

Authors:  Dragoş-Daniel Ţarălungă; Georgeta-Mihaela Ungureanu; Ilinca Gussi; Rodica Strungaru; Werner Wolf
Journal:  Comput Math Methods Med       Date:  2014-02-09       Impact factor: 2.238

9.  Standard External Doppler Fetal Heart Tracings versus External Fetal Electrocardiogram in Very Preterm Gestation: A Pilot Study.

Authors:  Gary Fruhman; Jeffrey A Gavard; Kristina McCormick; Judy Wilson-Griffin; Erol Amon; Gilad A Gross
Journal:  AJP Rep       Date:  2016-10

10.  Evaluation of an external fetal electrocardiogram monitoring system: a randomized controlled trial.

Authors:  Martha Monson; Cara Heuser; Brett D Einerson; Isaac Esplin; Greg Snow; Michael Varner; M Sean Esplin
Journal:  Am J Obstet Gynecol       Date:  2020-02-20       Impact factor: 8.661

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.