Bryan total disc arthroplasty: a replacement disc for cervical disc disease.

Markus Wenger1, Thomas-Marc Markwalder.   

Abstract

Total disc arthroplasty is a new option in the treatment of cervical degenerative disc disease. Several types of cervical disc prostheses currently challenge the gold-standard discectomy and fusion procedures. This review describes the Bryan Cervical Disc System and presents the Bryan prosthesis, its indications, surgical technique, complications, and outcomes, as given in the literature.

Entities:  

Keywords:  bryan prosthesis; cervical spine; complication; degenerative disc disease; disc herniation; myelopathy; outcome; spine surgery

Year:  2010        PMID: 22915917      PMCID: PMC3417861          DOI: 10.2147/mder.s7605

Source DB:  PubMed          Journal:  Med Devices (Auckl)        ISSN: 1179-1470


Introduction

Degeneration of an intervertebral disc involves progressive dehydration and fibrosis of the nucleus pulposus. These modifications induce loss of elasticity, loss of intervertebral height, formation of osseous spurs, cracking and bulging of the annulus fibrosus, and eventually, extrusion of nucleus tissue.1 Advanced degenerative disc disease destabilizes the anterior spinal column. Each of these events may generate pain. Nuchalgia, restriction of head movements, muscular spasms, torticollis, headaches, and pseudoradicular complaints are frequently reported. In the case of nucleus pulposus herniation into the spinal canal or the foraminal area, neural elements are usually compressed. Radicular pain in the supply area of the respective nerve is characteristic of nerve root compression. Nerve root compression may also be associated with neurological deficits such as sensible dysfunction, abolished tendon reflex, and paresis. Compression of the spinal cord results in myelopathy. Common clinical findings of myelopathy are hyperreflexia, clonus, disabling disturbances of gait, equilibrium, coordination, and difficulty handling small objects.2–5 Recent disc herniations often heal spontaneously or favorably respond to conservative treatments. In case of severe symptoms, microdiscectomy combined with segment fusion is required. However, this otherwise successful procedure is associated with a significant incidence of accelerated adjacent-segment degeneration.2,6,7 Ultimately, adjacent-segment morbidity requires further interventions. Maintaining motion at the index segment is believed to be an important approach to reducing problems at transient segments.4,6,8–12 Based on this knowledge, Dr Vincent Bryan developed his total cervical disc arthroplasty device in the 1990s. Dr Goffin of Belgium implanted this prosthesis for the first time in January 2000.5,8,13,14 The aim of this review is to present the Bryan® Cervical Disc System (Medtronic Sofamor Danek; Memphis, Tennesse) and its use in the treatment of degenerative disc disease. Several types of cervical disc prostheses are marketed at present.1,13,15

Surgical gold standard

The standard surgical procedure for cervical degenerative disc disease or disc herniation has been in use for more than 50 years. It involves an anterior approach to the cervical spine and removal of the disc with the offending tissue (anterior cervical discectomy [ACD]).16,17 This procedure is safe in view of nerve decompression.15,18 However, ACD alone shows a tendency to create subsequent problems. The most notable are pain because of chronic instability or malalignment of the operated segment, as well as adjacent-segment disease. In order to improve reliable bony fusion in correct alignment, surgeons began inserting a corticocancellous bone graft into the disc space (ACD with fusion [ACDF]).1,2,16,19,20 Later, these bone grafts were replaced by custom-made interbody fusion cages,21 which reduce the problem of matching the shape of the graft to the disc space and annoying transformations of the bone graft during the healing process. Because cages require only a small amount of autologous cancellous bone or a bone substitute, the severity of graft harvest site problems is significantly reduced. A plate-screw osteosynthesis is added if the stability of the cage construct is considered to be insufficient. Introduction of the operating microscope (microdiscectomy) allowed for magnification and direct illumination of the surgical field, rendering operations more precise and safe. These technical adaptations significantly improved ACD outcome; however, the problem of transitional-segment disease with subsequent interventions remained unsolved. Nevertheless, at present, microdiscectomy with cage insertion and eventual plating represents the gold standard in the surgical treatment of cervical degenerative disc disease and disc herniation.5,18,22 ACDF is associated with favorable long-term results.1,4,8,15,18,22–24 Table 1 compares the ACDF and the Bryan procedures.
Table 1

Comparison between the gold-standard ACDF procedure and Bryan total disc arthroplasty in the treatment of cervical degenerative disc disease and eventual disc herniation

ACDFBryan total disc arthroplasty
IndicationDegenerative disc disease, ±disc herniation, ±arthrosis formation, instability, too low or high disc spaceDegenerative disc disease, ±disc herniation, ±only mild myelopathy, otherwise healthy individual and cervical spine (see text); up to 43%15 of patients
Segments to be treatedC2/3 to C7/T1Approved for C3/4 to C6/7;25 also used rarely at the cervicothoracal junction10,15,28
No. of segments1 to several1–2 (or 3).
FluoroscopyLocalization of the target space, and final control of the cage’s position and alignmentPermanent use for patient adjustment, assessment of axis, burring, milling, and final control; increased exposure for patient and personnel
InstrumentsMicrodiscectomy instrumentsMicrodiscectomy instruments + Bryan frame + Bryan instruments
AccessStandard anterior with oblique, curvilinear, or horizontal incision16,17,19,20Standard anterior with oblique, curvilinear, or horizontal incision; exposure slightly larger to allow access for milling fixture
MicroscopeYesYes
Duration of surgeryShort; graft harvest prolongs operating time; in case of osteosynthesis, probably identical with BryanSlightly longer5,24 (some exceptions3); time for mounting of the scaffolding and adjusting slightly exceeds time for ACDF with graft harvest; learning curve
ImplantCage (usually carbon, PEEK or metallic), eventual platescrew osteosynthesis (titanium, steel)One-piece, bi-articulating, axially and functionally symmetric prosthesis (titanium alloy, polyurethane, saline)
Autologous bone graftYes; possible graft site complications; alternatively bone substitutesNone
Plate-screw osteosynthesisSometimesNone
Blood lossNegligible24Negligible; slightly higher due to burring and milling of bony surfaces5,9,11,24
Possible complicationsBleeding, N. laryngeus, dysphagia, infection; graft harvest siteBleeding, N. laryngeus, dysphagia, infection; device related
Post-operative measuresWe use rigid collars that restrict head movements,2 even though others avoid stiff collarsNo collar; no restriction of head movements14 (exception24)
Costs of surgeryRelatively low (surgery, cage, eventual bone substitute, rigid collar)Relatively high (surgery, prosthesis)33
Back to workLaterEarlier5,18,23
Overall costsProbably moreProbably less
Adjacent segment after surgeryPotential of mechanical overload over the long term (overload adds to inherent degeneration)Assumed protection over the long term
Published follow-upLong term (first operation >50 years ago)16,19,20Short to intermediate term, <5 years (first Bryan implantation in 20005,13)

Abrreviation: ACDF, anterior cervical discectomy with fusion; N, nervuus; PEEK, polyether ether ketone.

Rationale for cervical total disc arthroplasty

There are three main reasons why researchers look for improvements to the successful gold-standard ACDF procedure. First, most surgeons believe that creation of an osseous block out of a functional motion segment has long-term adverse effects on adjacent segments below6,10 and, more frequently, above14,25 the block. Adjacent levels are subject to increased strain, motion, and intradiscal pressure (mechanical overcharge), as they have to take over the work of the fused segment.2,5–7,9,13,18,24,26 Consequently, they tend to degenerate early.2 Elongation of the spondylodesis, including this pain-generating segment, is required later in many patients.14,15 Hilibrand et al2 reported an annual mean rate of 2.9% adjacent-segment disease during the first 10 years after operation; however, this value did not consider the inherent course of degeneration. Their Kaplan–Meier survivorship analysis of 374 fused patients revealed an incidence of 25.6% at 10 years. The risk of adjacent-segment disease was lowest for fusions at C2/3 and C6/T1, intermediate at C3/4 and C4/5 (factor 3.2), and highest at C5/6 and C6/7 (factor 4.9).2 In the long run, more than two-thirds of their patients who developed new transitional-segment disease needed further operations. They considered the inherent process of degeneration to be a more probable cause of adjacent-segment disease than increased stress after spondylodesis.2 Robertson et al10 found new symptomatic adjacent-disc disease after 158 fusions in 13.9% (6.9% annually) of patients at 2 years and considered transitional-level disease to be a sequela of spondylodesis. Goffin et al27 also reported a 92% incidence of radiologic adjacent-segment degeneration after ACDF (mean follow-up of 100.6 months, >60 months for all; mostly degenerative and trauma cases). However, only 6.11% (11 of 180) of patients eventually required surgery. Garrido et al24 analyzed three ACDF patients (12%) who underwent revision surgery for adjacent-segment disease within 48 months: one for remote-level degeneration and two for pseudoarthrosis formation, whereas only one (5%) had revision surgery in the Bryan cohort. There was a clear trend toward fewer secondary interventions in the Bryan group. Wigfield et al6 reported that fusion increased motion at adjacent levels. Second, reoperations of the original segment, even though rare,27 are required to treat pseudoarthrosis formation, graft migration, or graft collapse (eg, graft nonunion rate of 6% at 2 years in 221 single-level ACDF patients18).1,9,15,18,24,27 In the case of plate-screw osteosynthesis, implants fail, may become loose, and even migrate,1,28 and the implants have to be removed. Third, problems occurring at the graft harvest site (usually anterior iliac crest) such as hematoma formation, infections, chronic pain, meralgia paresthetica, lesions of the sacroiliac joint, and even pelvic fractures blur the otherwise favorable results.1,15,29,30 Sandhu et al29 reported a graft harvest site complication rate of up to 25%.

Assessment of patients with cervical degenerative disc disease

In many instances, the history and clinical presentation of a patient with degenerative disc disease and cervical disc herniation are specific.22 At this point, the radiological assessment is initiated, and we prefer magnetic resonance imaging (MRI) of the cervical spine to be the first investigation.15 MRI depicts huge osseous spurs, neural structures, the condition of the disc, disc herniations, and the relation between a sequester and the nerves. In T2-weighted MRI, degenerated discs have a dark, dehydrated appearance (black disc), and sometimes they appear collapsed. Abnormalities in the area of the endplate and the adjacent bone marrow reflect painful mechanical overload (eg, vertebral edema formation, Modic sign).31 In severe myelopathy, MRI reveals signal alterations within the spinal cord. The central canal is studied. However, function of the motion segment cannot be assessed by conventional MRI. This is why X-rays in anteroposterior and lateral directions, as well as a functional series of conventional radiographs with full flexion and extension of the head, are mandatory (caveat: severe myelopathy). These images reliably show hypermobility or hypomobility of the segment, height of the disc space, and significant changes due to arthrosis formation. In a few ambiguous cases, usually a more detailed examination of the relation between discrete osseous changes and neural elements is needed. In these instances and in cases of MRI contraindications, functional myelography and myelography combined with computed tomography (CT) are also used.15 We use nerve conduction or electromyography studies only to exclude nervus medianus or ulnaris compression syndromes.4

Indication and contraindication for Bryan disc implantation

In cervical degenerative disc disease, a conservative treatment is initiated.1 Therapy involves analgesics, muscle relaxants, anti-inflammatory medications, immobilization in a rigid orthosis, and physical therapy. Nonresponders are evaluated for total disc arthroplasty within a reasonable period of time (eg, 3 months in the Swiss protocol; others after 6 weeks9,18,24,32 or 6 months15). Arthroplasty, usually an elective surgery, requires good general health and is possible only in adults (mature skeleton; 21 years25 to <60 years of age,1,33 rarely >60 years10,15). There should be no local or systemic infection (including HIV and hepatitis B and C),15 allergy to any component of the implant, medical therapy interfering with bone healing (eg, steroids), diabetes mellitus, chronic kidney disease, severe obesity (body mass index > 40 kg/m2 15), active malignancy, pregnancy, or a general disorder of the skeleton, such as rheumatoid arthritis, osteopenia, or osteoporosis.1,4,5,11,12,15, 18,25,33 Short necks and superimposed, large shoulder girdles obviate fluoroscopy of caudal segments.8,18,28 Axial neck pain alone should not be treated by arthroplasty.3,15 Minor spondylosis with anterior, posterior, and uncovertebral spurs is frequently encountered. These osteophytes can easily be removed during surgery.15 However, individuals presenting with facet arthrosis, target disc space height less than 50% of the normal,5,15,18 kyphotic target segments,12,34,35 olisthesis of ≥2,5,25,32 3,4,15 or 3.5 mm,9,24 instability with angular motion of ≥11° more than either adjacent segment, 5,15,24,28,32 and spontaneous fusion (ie, <2° motion)5,15 should not undergo arthroplasty.4,5,15 We consider myelopathy to be an absolute contraindication. 25 Without analysis, we believe that the chance of healing will be greater if the diseased part of the spinal cord is immobilized as much as possible by creating an osseous block. The importance of healing myelopathy, or at least stopping its progression, outweighs all the advantages of arthroplasty. Nevertheless, many surgeons treat patients with myelopathy. 3,5,11,15,18,23–25,28,32,33,36,37 The probably treat myelopathy due to strict retrodiscal offending tissue9 and exclude patients with myelopathy due to posterior or retrovertebral compression (eg, ossification of the posterior longitudinal ligament [OPLL]8). In our view, previous cervical spine surgery, at least at the target segment,15 as well as previous cervical trauma (eg, whiplash injury), is a criterion for exclusion.1 Nevertheless, some surgeons implant prostheses after neck surgery,28,37 and others even combine arthroplasty and fusion.11 Because the rate of complications increases per treated level, we never implant more than two prostheses.25,28,38 Some surgeons restrict arthroplasty to a single-level disease,1,12,18 some treat up to three levels,28,37 and some surgeons only advise against the treatment of more than three levels.15 Auerbach et al15 showed that 43% of patients with cervical degenerative disc disease are possible arthroplasty candidates (conversely, 57% have contraindications); however, only 8.4% (14 of 167) had nonfusion surgery. Costs, national healthcare politics, and other nonmedical factors obviously have an impact on patient selection. Individuals with new paresis or myelopathy need either arthroplasty or ACDF on an emergency basis.

Bryan prosthesis

The Bryan prosthesis is composed of two convex titanium alloy spheres, an intervening polyurethane core that articulates with each sphere, a polyurethane circular wall, and a cavity (Figures 1–4).1,3,8,13,14,25,39 It is axially and functionally symmetric (ie, no lordotic shape).1,12,36 The prosthesis comes in a one-piece, preassembled device. Once saline is injected into the cavity, the device is ready for implantation. Currently, the prosthesis is available with diameters of 14–18 mm and a height of 8.5 mm.3,25,36 The device is approved for implantation at C3/4 to C6/7.25 It is rarely used at the cervicothoracic junction as well.10,15,28,37
Figure 1

Bryan cervical disc prosthesis. There are two porous convex shells with the central access port to the void and two punched anterior stops, as well as the polyurethane sheath retained by two wires. Reproduced with permission from Medtronic Sofamor Danek USA, Inc.

Figure 4

Cross section of Bryan prosthesis. The void is located around the center post, squeezed between endplates and the nucleus, and visible again in the area of the sheath (membrane) and the shell rim. Note that the nucleus is convex at the center and concave at the outer border. Reproduced with permission from Medtronic Sofamor Danek USA, Inc.

The shells consist of a titaniumaluminiumvanadium alloy, and their outer surfaces are convex. High-friction porosity is obtained by the fusion of 250-μm titanium pellets onto their surfaces.5 The concave inner surfaces, articulating with the core, are polished to prevent wear debris formation. Each endplate has a metallic center post. The posts fit into the center dents of the nucleus. There is a low circular rim all along the periphery of the spheres, which is directed toward the vertebral endplates. At identical locations of the outer edge, each shell has a perpendicular, punched flange (anterior stop) bent cephalad on the upper shell and caudal on the lower endplate. Most of the device is filled with a low-friction core of medical-grade polyurethane. The surface geometry of the nucleus is convex, and it matches with the titanium alloy endplates. However, in the area of the shell rims, the nucleus changes its shape and becomes biconcave. A circular polyurethane band connects the two spheres. It is retained by a wire in the groove at the outermost margin of each shell. The sheath represents the vertical wall of the device. There is a single void in the center of the device. It is confined by the spheres, the core, and the sheath. In the center of each sphere, there is a port with access to this cavity. Each part of the prosthesis has a specific function. The convex endplates, with their rims, exactly fit to the precision-milled vertebral recesses. This feature provides for immediate or primary stability after the release of disc space distraction. Prior to bone ingrowth, the shell’s rough texture protects against excessive torsion. Maximal capping of the vertebral endplates shares the loads applied to the device and impedes osseous bridging of the device.28,38 Porosity allows for bony ingrowth,14 which represents late or secondary stability. Even though they are not reliable in all instances,28 anterior stops anticipate posterior dislocation during surgery and later in life. Intraoperatively, the introducer instrument is attached to the punches in the stops. The flanges may be useful during revision surgery.33 The high-resistance polyurethane core absorbs axial, torsional, and shear loads. It allows for torsion, a maximum of 11° flexion, extension, lateral bending, 2 mm of translation, and any combination of these.8,14,36,39 Motion is stopped if the shell’s center post comes in contact with the wall of the dent. Contact between the outermost border of the shell and the peripheral concavity of the nucleus limits motion as well. However, motion is stopped by the prosthesis only if surrounding anatomical structures such as facet articulations, ligaments, and remnants of the annulus fail to stop motion before. In this way, the prosthesis is unconstrained for physiological motion.12,39 This semipermeable polyurethane sheath retains polymeric wear debris1 and prevents soft tissue ingrowth.39 Later, a pseudocapsule of noninflammatory fibrovascular tissue may arise around the sheath.9,8 Permanently submerged into sterile physiological saline (initial lubricant) the device is prepared intraoperatively. One seal plug is screwed into the sphere, and the device is turned onto this sphere, repetitively squeezed in order to eliminate as many air bubbles as possible, and slightly compressed (see kyphosis formation discussed later). In this state, the second seal plug is closed. Usually, the height of the prepared prosthesis is less than 8.5 mm, and the shells are no longer parallel. After implantation and release of distraction, the spheres adjust themselves to the vertebral endplates. Titanium, titanium alloy, and polyurethane implants are successfully used in other fields of surgery (eg, orthopedics, vascular surgery), where they show favorable long-term biocompatibility.39–41 To our knowledge, no reports on Bryan material intolerance have been published. However, patients with allergies should not undergo implantation. Polyurethane is preferred to other polymers used in orthopedics (eg, ultrahigh molecular weight polyethylene [UHMWPE]) because it best emulates the elasticity and shock-absorbing capability of the healthy disc.36 Polyurethane has a lower immunologic potential than polyethylene.36 Bryan prostheses were exposed to in vitro cyclic fatigue testing.14,39 After 10 million cycles, the nucleus mass loss was 1.76%, which is assumed to be 50–100 times less than that in total hip arthroplasties.39 Nucleus height loss was on average 0.02 mm per million cycles. After 40 million moves, wear averaged 18% by weight.42 Therefore, contact between the shells was imminent. However, in vitro decay is estimated to be significantly higher than in vivo decay (probably 5–10 times).42 Under the assumption that 1 million moves occur during 1 year with normal daily activities, the Bryan prosthesis is (optimistically) expected to be functional for more than 40 years.42 In animal studies, polymer wear particles had a diameter of 3.9 μm on average (most of them in the range of 1–5 μm).8,39 There was no debris dissemination into distant organs (lymph node, liver, spleen),14 penetration of the dura mater, accumulation of metal ion in body fluids, or significant inflammatory response by activation of various cytokines, macrophages, osteoclasts, or other cells.14,42,43 In contrast to large joint arthroplasties and metal-on-metal prostheses, Bryan discs obviously do not induce metallic debris formation.8,39,42 In a recent study on explanted devices, no case of device fracture, polymer oxidation, or metal corrosion was noted.42 Bryan prostheses allow for later MRI and CT examinations. 1,4 The polyurethane core is not radiopaque. The Bryan prosthesis can be removed if it fails. Normally, the remaining bony endplates allow for conversion to fusion.4,10,18,28,33

Preoperative measures

Preoperative measures and surgical procedures have been previously described in detail.3,8,25,44 On axial MRI (or CT), the diameters of the caudal and cephalad vertebral endplates adjoining the target disc space are assessed. Attention is paid to laying the disc templates on the images that are parallel to the endplates. Bony spurs are excluded from the measurement. The smaller size represents the appropriate prosthesis diameter, and this size will be validated intraoperatively. General anesthesia is induced, and the patient is positioned supine and exactly adjusted to lie in the middle of the horizontal operating table. The patient’s head is slightly elevated and supported (with a donut of 2.5 cm height).8,36 A rolled towel fills the void between the cervical lordosis and the flat table. The head is fixed with tapes, and hence, the cervical spine remains in an accurate neutral position.36 The body is firmly strapped to the operating table. In order to minimize superimposition of the shoulder girdle into lateral fluoroscopy images, wide tapes are used to pull the shoulders and arms in the caudal direction (caveat: plexus brachialis). At this point, processus spinosi should be centered between facets in anteroposterior fluoroscopy. The inclinometer, with a free-swinging, radiopaque pendulum, is taped to the cranial edge of the fluoroscopy apparatus. Using an offprint of the lateral view, the goniometer measures the angle between the pendulum and the line that runs through the superior posterior corner of the cephalad vertebral body and the inferior posterior corner of the caudal vertebral body adjacent to the target disc. This represents the disc space angle, which is eventually used for parallel burring of the endplates and circular precision milling of the recesses (Figure 5).
Figure 5

Preoperative lateral fluoroscopy with the patient in definitive position for surgery. The angle between the line that runs through the superior posterior corner (asterix) of the cephalad vertebral body and the inferior posterior corner of the caudal vertebral body (dot) adjacent to the target disc and the pendulum is depicted. In this case, an interspace deviation of 15° was measured.

Bryan surgical procedure

We prefer an oblique incision along the medial border of the right sternocleidomastoid muscle.22,38,45 Alternatively, a curvilinear or transverse incision may be used.23,36,37 The platysma muscle is divided, and the main vessels are retracted ipsilaterally using blunt dissection. The trachea together with the oesophagus is displaced contralaterally. Trauma to the longus colli muscles by excessive coagulation or retraction is avoided. The target disc is identified using fluoroscopy. The disc and the hernia are removed, and the lateral parts of the annulus fibrosus and the ligaments remain intact on either side. In contrast to others,11 we resect the posterior longitudinal ligament to assure complete decompression.11,12 Any anterior, posterior, and foraminal osseous spurs are removed. Then, the 8.5-mm sagittal cam distractor is introduced to evaluate the disc space height and hypermobility or hypomobility of the segment. This examination defines whether arthroplasty or conversion to fusion should be used. The gravitational referencing system is assembled.3,8,25 Sterile extensions are mounted on either side rail of the table. A vertical rod is clamped to each extension, and these rods carry the retractor frame. The middle of the frame is centered over the incision (Figure 6). Depending on the patient’s trunk anatomy, it is placed a few centimeters above the wound. In this position, the frame is meticulously aligned parallel to the spine and horizontal using a water level. All clamps are firmly closed. Appropriate wound retractor blades are affixed onto either side of the frame. Cephalad and caudal retractions are optional. Exposure is adequate if there is enough space to put the dual-track milling guide on the anterior aspect of the spine. The transverse centering tool marks the sagittal midline. The sagittal wedge is introduced into the space and exactly adjusted to these midline marks. The milling guide is imposed on the wedge. Correct position on the vertebrae is controlled using contact pins and fluoroscopy. The preoperatively assessed sagittal slope of the milling fixture is obtained by using the preset protractor and the water level. In this position, the milling fixture is affixed onto the cranial retractor crossbar and anchored with screws into each adjacent vertebra, and the wedge is removed. Through each guide slot, the milling depth gauge measures the distance between the fixture and the more anterior vertebral edge (Figure 7). The burring block-ring assembly of the assumed prosthesis size is mounted on the fixture. Under fluoroscopic control, the burring depth gauge is lowered through the assembly down to the level of the posterior longitudinal ligament (caveat: injury to the dura) (Figure 7). This returns the burring depth and the definite size of the prosthesis. The calibrated parallel burr is mounted on the assembly. The burr is lowered into the interspace by reducing the assembly height (Figure 8). This feature prevents inadvertent injury of the thecal sac. Bilateral movements of the burr ensure flattening of the whole endplate. After removal of the burr and the assembly, the calibrated milling disc is introduced (Figures 8 and 9). Pivoting movements along the y-axis of the spine are applied to create the exact shell geometry in each endplate (Figure 9). Saline is used for irrigation during the motor burring and milling procedures. All instruments but the anchor screws are removed. The anchor post distractor temporarily enlarges the disc space. The introducer instrument, which grasps the device at the holes in the anterior stops, is used to gently hammer the prepared device into the space. Lateral and anteroposterior fluoroscopy controls the correct device position. After the removal of the anchors, the wound is thoroughly rinsed32 in order to eliminate bone dust and sutured over a closed suction drain. Although blood loss is negligible in general (64 mL,9 80 mL [42 mL for ACDF],24 90 mL [50–130 mL],4 97.5 ± 77.6 mL [maximum 300],5 208 mL11), blood transfusions37 may be indicated in exceptional cases.
Figure 6

Table extension rails, bilateral vertical rods, Bryan frame, and wound retractors in their definitive positions centered on the exposed intervertebral target disc. Note the rolled towel supporting physiological cervical lordosis. Anesthesia devices are not depicted. Reproduced with permission from Medtronic Sofamor Danek USA, Inc.

Figure 7

Intraoperative lateral fluoroscopies. Left: milling depth gauge at the anterior edge of the vertebra (white arrow), anchor screws (black arrows), retractor blades (white asterix), and orotracheal anesthesia tube (black asterix). Right: burring depth gauge at the posterior vertebral edge and the remainders of the posterior longitudinal ligament (white arrow).

Figure 8

Intraoperative lateral fluoroscopies. Left: close to the thecal sac, the burr abrades the posterior edge (arrow) of the caudal vertebral endplate. Right: the milling disc (arrow) prepares the recess. Note the screw-anchored fixture, lateral retractors, vertical bar (asterix), and the orotracheal anesthesia tube.

Figure 9

Pivoting moves (double arrow) are applied to the milling disc during preparation of the vertebral recess. Note the thecal sac containing the myelon next to the posterior rim of the milling disc. The hook in the slit of the fixture avoids inadvertent neural element lesion. The milling fixture is affixed to the anterior aspect of the spine. Two anchor screws and extensions to the cranial crossbar secure its position in the previously determined angle. Optional cranial and caudal wound retractors are mounted. A longitudinal bar of the frame is depicted. Reproduced with permission from Medtronic Sofamor Danek USA, Inc.

The above procedure describes the single-level method.3,8,25 For bi-level procedures, several instruments are exchanged and repositioned. The surgical exposure is more extensive; otherwise, there are no major procedural changes. The complexity and learning curve44,46 of the actual Bryan system have the potential to repel some surgeons. A simplified implantation procedure using the identical prosthesis is currently under development. We do not perform this procedure on an outpatient basis. Our patients remain in the recovery room overnight in order to cover the first of the critical 36 hours after surgery.47 Hematomas may rarely compress the trachea or the thecal sac in a dramatic manner early after surgery.3,11,47 In contrast to ACDF, no collars are used in this procedure. 1,3,14 The patients ambulate within a few hours. They are allowed to resume nonstrenuous activities immediately18 and usually return to work earlier than ACDF patients.11,18,23 Thus, the overall costs are reduced to some extent.

Complications o f the Bryan disc prosthesis and their prevention

Complications occur at various stages.28 They are related to the surgical access, the Bryan system procedure, and the long-term behavior of the device. Complications depend on the number of implants.25,28,47 Goffin et al25 reported a complication rate of 6.8% for single-level procedures, 11.6% for bi-levels, and an overall complication rate of 6.3% per treated level. Pickett et al28 reported 8.1% perioperative complications, equivalent to 6.2% per treated level. They also reported 5.2% of late-onset complications at the mean follow-up of 12 months. Most surgeons switch from ACDFs involving anterior approaches to Bryan procedures. Therefore, there is no major learning curve for the identical Bryan approach. Vital structures such as the internal carotid artery, the superior thyroid artery, the internal jugular vein, the trachea, the esophagus,25 and the thecal sac containing neural elements are seldom injured. Aggravation of preexisting neurological dysfunction is rare.28 However, undetected retraction and compression trauma to adjoining structures is a concern.25 Thus, anterior approaches are associated with temporary dysphagia ( incidence of 12.3%48), dysphonia (7%4), hoarseness (4.9%48), and unilateral, true vocal cord paresis (1.4%48). Hidden bleeding points inadvertently created by transection of small vessels may lead to the formation of life-threatening prevertebral or epidural hematoma shortly after the surgery.10,11,28,33 Horner syndrome, venous air embolism, and chylothorax formation have not been reported. Postoperative kyphosis through the Bryan device and/or the treated functional spinal unit is increasingly reported.11,12,33,36,44 The complexity of the Bryan procedure, which is associated with a substantial learning curve, and the inclusion of patients with preexisting malalignment may contribute to its incidence.44 Kyphosis formation can sometimes be prevented by a slight modification of the milling guide angle, avoidance of asymmetric endplate preparation, endplate overmilling, and compression of the prosthesis during closure of the second seal plug.11,12,34–36,44,49 Robertson et al10 studied preoperative kyphosis in some patients: 12 with slight, 11 with mild, and 12 with severe preoperative kyphosis. Slight kyphosis turned into normal lordosis in six patients, remained unchanged in four, and worsened to severe in two. Mild kyphosis diminished to slight in four patients, remained unchanged in four, and became severe in three. Severe kyphosis remained severe in all instances. Postoperative severe kyphosis decreased prosthesis motion. Fong et al36 reported that 9 of 10 patients had kyphosis through the prosthesis at a mean follow-up of 4 months (3–12 months). Pickett et al49 observed kyphosis formation in all of their 14 patients. Their prostheses moved, and the overall cervical alignment was preserved (follow-up, 6–24 months; mean, 12 months). Kyphosis most frequently remains asymptomatic.11 Both myositis ossificans and heterotopic ossification contribute to fusion of the prosthesis. Myositis ossificans is induced by the excessive coagulation and retraction damage of the longus colli muscles.12,32,50 Mesenchymal cells released from muscle components may transform into ossification-inducing cells. Heterotopic ossification occurs in approximately 6%–12% of cases4,25,28,32,33,38,51,52 and progresses from hardly detectable bony spurs to solid fusion. A grading system (Class 0 [no ossification] to 4 [ossification causing inadvertent arthrodesis]) was introduced for lumbar arthroplasty.53 Endplate burring and milling release bone dust that contains osteogenic precursor cells, bone morphogenic protein, and various cytokines.12,32,43,53,54 Copious rinsing of the wound and use of nonsteroidal anti-inflammatory drugs (NSAIDs) in the early postoperative period (2–4 weeks) is supposed to act against ossification.12,14,18,28,32,52 However, there remains some concern that NSAIDs compromise bony shell ingrowth32 as well. Solid fusion turns the device into a nonmoving disc space filler, which behaves very much like a primary ACDF.4,28,32,51 However, the protective effect on adjacent segments is lost.32 Unless radiological controls are performed, spontaneous prosthesis fusion may remain undiscovered. No treatment is required as long as fusion is asymptomatic. It is mandatory to implant the largest prosthesis possible.38 Huge shells cover most of the vertebral surface from which bridging syndesmophytes might originate. Large spheres also distribute strain over large areas of the device. Longevity of the device is optimized, the probability of subsidence is minimized, and the area of bony ingrowth is maximized. Prior to system adaptation, prosthesis dislocation in the z-axis occurred due to milling mistakes,25 but now, anteroposterior prosthesis dislocation is rare. Inferior plate dislocation of the device requiring removal of the prosthesis with conversion to fusion has been described.4,28 Internal subluxation of the device occurred in a case of segmental hypermobility.37 There is no report of cervical prosthesis infection, prosthesis subsidence, or aseptic osteolysis due to wear debris formation. Little is known about same-segment disease involving facet arthrosis and the recurrence of uncoforaminal bony spurs.55 In case of bi-level treatment, prosthesis spheres may even contact each other.28 Wrong-level surgery should be an exception.3,10,25

Outcome

Late outcome of neural element decompression is expected to be the same for both ACDF and Bryan procedures.14 However, published outcomes at more than 5 years are not available for Bryan arthroplasty.33 The following is not a complete review of the literature, and the patients and the results in some studies may overlap. Bryan14 presented the first 97 patients who received single-level prosthesis implantation in several European centers. Using modified Odom’s56 criteria, clinical success was reported to be 87% (n = 40) after 1 year and 89% (n = 8) after 2 years. At 1 year, in 86% (38 of 44) of patients, the flexion-to-extension angle was 8° ± 5° on average, and in 4 patients, it was 1°. At 2 years, 100% (10 of 10) of prostheses moved (11° ± 5°). High patient satisfaction was noted, and there were no restrictions on the activities of daily living. One patient had evacuation of a wound hematoma, and another had foraminotomy to remove a remaining osteophyte. In one patient, the device migrated over 2 mm in the anteroposterior direction; otherwise, there were no prosthesis dislocations, subsidences, failures, bridging, or inflammatory response to the prosthesis. Goffin et al3 conducted a prospective, concurrently enrolled, European multicenter trial on 60 patients with single-level prosthesis. Using modified Odom’s criteria, clinical success was obtained in 86% (52 of 60) of patients at 6 months and in 90% (27 of 30) of patients at 1 year. At 12 months, patients met or exceeded the US population mean for SF-36 physical and mental component scores.57 At 6 months, flexion-to-extension angles were just less than 9° ± 4° in 93% (53 of 57) of patients. At 1 year, they were just more than 9° ± 6° in 88% (21 of 24) of patients, and two had fused segments (1°). Anteroposterior migration (≤3 mm) was observed in one patient and suspected in another. There was no device failure, subsidence, bridging, or explantation. Two patients had anterior or posterior revision surgery for additional neural element decompression. One had shortness of breath, requiring hematoma evacuation, and one wrong level was operated. Goffin et al25 again reported on the European multicenter study. Single-level Bryan arthroplasty (n = 103) was a clinical success (excellent, good, fair) in 90% (83 of 92), 86% (76 of 89), and 90% (44 of 49) of patients after 6, 12, and 24 months, respectively. In bi-level Bryan arthroplasty (n = 43), clinical success was 82% after 6 months and 96% after 12 months. After 1 year, 88% of single-level and 86% of bi-level Bryan prostheses moved. After 2 years, 93% of single-level Bryan prostheses moved >2°. Flexion-to-extension angles were 7.9° ± 5.3° for single-level prostheses, and 7.4° ± 5.1° for bi-level prostheses at 1-year follow-up. Leung et al32 reviewed 90 patients at 12-month followups. The incidence of heterotopic ossification was 17.8% (16 of 90 patients). Among them, 6.7% (n = 6) of patients had Class 3 or 4 ossification53 and another 11% (n = 10) had a range of motion of <2°. Nevertheless, 89.9% (80 of 89) of patients had a favorable Odom’s outcome at 1 year so that there was no association between ossification and clinical outcome. Lafuente et al4 conducted an observational clinical study involving 46 patients with single-level Bryan prostheses. Visual analog scale (VAS),58 SF-36 physical component, SF-36 mental component, and neck disability index (NDI)59 improved in a statistically significant manner in all patients. Odom’s criteria were satisfactory in 84% and 89% patients at 1-year and 2-year follow-ups, respectively. The range of motion was 7.72° on average (range, 0°–17°). Two patients (4.3%) had ankylosis and a good clinical outcome. In one patient, the prosthesis was removed, and the segment was fused because of interior disc plate dislodgment. There was no subsidence. Interestingly, they operated on a patient who had quadraparesis due to disc herniation, patients with previous spine surgery, and those with other-level fusions. Sekhon et al37 reported on a prospective study involving 15 patients receiving 24 devices (1–3 levels). All had previously undergone anterior or posterior cervical spine surgery. Follow-up was 24.2 ± 10.5 months. VAS was 7.8 ± 3.3 preoperatively and 1.4 ± 2.6 postoperatively. The Oswestry60 disability index was 27.1 ± 15.6 before and 17.6 ± 8.5 after surgery. Range of motion was 5.5° ± 4.7° before and 4.3° ± 2.6° after surgery. No revision surgery was required; however, one patient experienced hypermobility with internal subluxation of the device, and two required blood transfusions. Interestingly, one patient had clicking of the prosthesis. Pickett et al28 assessed 74 patients with 96 Bryan prostheses, who were followed for a mean of 12 months (maximum 36 months). NDI changed from 20.1 ± 9.2 preoperatively to 7.1 ± 9.6 postoperatively; the Oswestry disability index from 34.4 ± 8.7 to 17.0 ± 9.7; VAS from 6.9 ± 3.0 to 1.7 ± 2.2; SF-36 physical component from 34.3 ± 10.5 to 46.5 ± 9.1 and mental component from 44.1 ± 10.3 to 50.3 ± 7.9. Two prostheses fused (heterotopic ossification; undersized prosthesis) and 92 moved 8.13° on average. Neck pain was present in 19 patients (25%). There was one Bryan failure, probably due to internal subluxation of the nucleus in extension, and one late-onset migration of the device inducing regional kyphosis. Three patients required further decompressions, and one required exchange of the prosthesis because of kyphosis formation. One retropharyngeal hematoma was evacuated. Yoon et al11 retrospectively studied 46 single-level Bryan patients at 11.8 (2.9–19.5) months after surgery. Twelve of them received a combination of prosthesis and single-level fusion. NDI changed from 74% preoperatively to 24% at 1 year, radiating pain VAS changed from 8.15 to 1.35, and neck pain VAS changed from 6.5 to 3.8. However, in four patients, neck pain VAS increased from 5.32 to 6.9 and NDI increased from 67% to 75% at 13.8 months on average. The range of motion of the treated segment was 13.3° ± 5.7° before surgery, 9.3° ± 3.7° at 1 month after surgery, and recovered to 14.4° ± 4.5° at 1 year. Using conservative therapy, a patient fully recovered from quadriplegia due to an acute subdural hematoma. There was no fusion or device migration. Sasso et al9 presented a prospective, randomized trial. In the Bryan arm, there were 56 patients treated at one level. With statistical significance, all outcome scores improved from the preoperative period to the postoperative period. NDI changed from 47 preoperatively to 11 at 2 years (n = 49), arm pain VAS from 70 to 14, neck pain VAS from 72 to 16, SF-36 physical component score from 34 to 51, and SF-36 mental component from 46 to 54. Flexion-to-extension angle was 7.9° on average at 2 years. There were no fusions, device failures, explantations, or intraoperative complications. Later, two patients required further surgery for adjacent-segment disease. Heidecke et al33 published a prospective study of 54 patients receiving 59 Bryan prostheses. All patients had either good or excellent outcomes (Odom) at 2 years, and moving protheses had a 6° ± 3° range of motion. No useful motion (<3°) was observed in seven single-level patients (12%). Among them, five had heterotopic ossification Class 3 or 4, and two had none. In the remaining cohort, ossification Class 1 or 2 was present in 12 segments (overall 17 segments or 29%). One retropharyngeal hematoma was evacuated, and one dysfunctional prosthesis had to be removed and was converted to fusion. Migrations or dislocations did not occur. Riew et al5 performed a cross-sectional analysis of two prospective, randomized multicenter trials to evaluate arthroplasty in myelopathy patients. In the Bryan trial, there were 47 arthroplasty patients. NDI, neck pain VAS, arm pain VAS, SF-36 physical component, and SF-36 mental component improved in a statistically significant manner. At 2 years, 89.7% of patients had improvement or maintenance of neurological status, 46.2% had improvement in gait function, nobody had deterioration of gait function, and 94.7% were satisfied with the outcome. There was no revision surgery and no implant-related adverse events. Cheng et al23 carried out a prospective, controlled, randomized clinical trial. In the two-level Bryan arm (n = 31), neck pain VAS changed from 7.3 preoperatively to 1.5 at 2 years, arm pain VAS from 7.1 to 1.4, NDI from 50 to 11, and SF-36 physical component score from 35 to 50. Odom’s scale was excellent, good, or fair for all. Range of motion was 7.9° on average at 2 years. Apart from a deep-vein thrombosis, there were no complications in the Bryan cohort. Yang et al12 studied 15 single-level Bryan arthroplasty patients for 29.5 months on average (24–35 months). For all patients Odom’s criteria were excellent, good, or fair at 2 years. With statistic significance, the VAS changed from 7.0 before surgery to 2.4 at 2 years, NDI from 23.0 to 8.9, and the Bryan range of motion was 10.4 ± 2.7° before surgery and 10.9 ± 2° at 2 years. Interestingly, the range of motion decreased 1 month postoperatively (as seen in other studies), but recovered later. Heller et al18 reported on a randomized, controlled, multicenter clinical trial. In the Bryan arm, 242 patients underwent single-level treatment. Mean NDI improved from 51.4 preoperatively to 16.2 at 2 years, arm pain VAS score from 71.2 to 19.1, neck pain VAS score from 75.4 to 23, SF-36 physical component from 32.6 to 47.9, SF-36 mental component from 42.3 to 51.7, and the range of motion from 6.5° ± 3.4° to 8.1° ± 4.8°. Improvements were statistically significant. Overall success rate was 82.6% (17.4% failure); 1.7% of patients experienced implant-related adverse events, and 2.5% had secondary surgical procedures at the treated level (one revision, three removals, and two reoperations). Range of motion slightly increased during follow-up. There was no fusion, and the Bryan patients returned to work earlier. Garrido et al24 undertook a prospective, randomized, controlled trial. In the Bryan arm, there were 18 patients available for a 48-month follow-up. All had single-level treatment. Their NDI improved from 51 preoperatively to 10 at 4 years, neck pain VAS from 76.2 to 13.6, arm pain VAS from 78.8 to 10.8, SF-36 physical component from 33.1 to 49.4, and SF-36 mental component from 43.2 to 53.5. Overall results were more favorable for the Bryan than the ACDF cohort. Revision surgeries were more frequent in the ACDF group (n = 6; 23%). One patient (5%) of the Bryan group had a follow-up surgery for adjacent-segment disease. They conclude that arthroplasty may progressively replace ACDF. We38 retrospectively followed 25 patients with 29 prostheses for 22.3 ± 9.4 months (range, 11.5–45.6). One prosthesis fused, and the remaining 28 moved 9.5° ± 4.7° on average (range, 3°–20°) (Figure 10). This range of motion was very similar to that of adjacent untreated segments. There was no reoperation, device-related complications, prosthesis dislocation, or subsidence. For unknown reasons, the fused prosthesis had been implanted eccentrically. Although it was not stated in our article, several patients temporarily complained about postoperative dysphagia or neck pain. No patient was treated for myelopathy.
Figure 10

Functional lateral radiographs, 3 months after successful Bryan arthroplasty (diameter, 17 mm) for C5/6 disc herniation in a 49-year-old woman. The prosthesis moves physiologically. There is a minimal area of failed (or delayed) bone ingrowth at the posterior border of the upper shell.

Conclusions

Total disc arthroplasty in the cervical spine is associated with important advantages. Bryan prostheses challenge the gold-standard ACDF in the treatment of cervical degenerative discopathy associated with eventual neural impingement. Neural structures are reliably decompressed. Prostheses maintain physiological alignment, kinematics, and biomechanics of the segment. In this way, accelerated transient-segment degeneration with subsequent radiculomyelopathy is averted with high probability. Patient’s satisfaction, safety, tolerability, and longevity of the Bryan prosthesis system seem to be very acceptable. Bone graft harvest site morbidity is eliminated. Frequently, patients return to work earlier. However, Bryan prostheses have the disadvantage of high implant costs, technical complexity of the implantation procedure, implant-associated complications, and spontaneous fusion. At present, the lack of long-term results of more than 5 years obviates a more definite appreciation. As a consequence, the definite role of the Bryan device in the therapy of degenerative discopathy and disc herniation is not yet settled.
  55 in total

1.  Stabilization of the cervical spine by anterior fusion.

Authors:  R W BAILEY; C E BADGLEY
Journal:  J Bone Joint Surg Am       Date:  1960-06       Impact factor: 5.284

2.  Cervical arthroplasty with the Bryan disc.

Authors:  Michael Y Wang; Clarence H S Leung; Adrian T H Casey
Journal:  Neurosurgery       Date:  2005-01       Impact factor: 4.654

3.  Measurement of pain.

Authors:  E C Huskisson
Journal:  Lancet       Date:  1974-11-09       Impact factor: 79.321

4.  Cervical arthroplasty complicated by delayed spontaneous fusion. Case report.

Authors:  Jonathon F Parkinson; Lali H S Sekhon
Journal:  J Neurosurg Spine       Date:  2005-03

5.  A prospective randomized multicenter clinical evaluation of an anterior cervical fusion cage.

Authors:  R J Hacker; J C Cauthen; T J Gilbert; S L Griffith
Journal:  Spine (Phila Pa 1976)       Date:  2000-10-15       Impact factor: 3.468

6.  The prevalence of indications and contraindications to cervical total disc replacement.

Authors:  Joshua D Auerbach; Kristofer J Jones; Christian I Fras; Jessica R Balderston; Scott A Rushton; Kingsley R Chin
Journal:  Spine J       Date:  2007-11-05       Impact factor: 4.166

7.  Clinical outcomes of Bryan cervical disc arthroplasty a prospective, randomized, controlled, single site trial with 48-month follow-up.

Authors:  Ben J Garrido; Tarek A Taha; Rick C Sasso
Journal:  J Spinal Disord Tech       Date:  2010-08

8.  The SF36 health survey questionnaire: an outcome measure suitable for routine use within the NHS?

Authors:  A M Garratt; D A Ruta; M I Abdalla; J K Buckingham; I T Russell
Journal:  BMJ       Date:  1993-05-29

9.  The Neck Disability Index: a study of reliability and validity.

Authors:  H Vernon; S Mior
Journal:  J Manipulative Physiol Ther       Date:  1991-09       Impact factor: 1.437

10.  Classification of heterotopic ossification (HO) in artificial disk replacement.

Authors:  Paul C McAfee; Bryan W Cunningham; John Devine; Eric Williams; Janet Yu-Yahiro
Journal:  J Spinal Disord Tech       Date:  2003-08
View more
  4 in total

Review 1.  Spinal motion preservation surgery: indications and applications.

Authors:  Ioannis D Gelalis; Dimitrios V Papadopoulos; Dionysios K Giannoulis; Andreas G Tsantes; Anastasios V Korompilias
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-06

2.  The MOVE-C Cervical Artificial Disc - Design, Materials, Mechanical Safety.

Authors:  Annette Kienle; Nicolas Graf; Carina Krais; Hans-Joachim Wilke
Journal:  Med Devices (Auckl)       Date:  2020-09-25

Review 3.  Adverse Events Following Cervical Disc Arthroplasty: A Systematic Review.

Authors:  Jordan C Xu; Chandni Goel; Michael F Shriver; Joseph E Tanenbaum; Michael P Steinmetz; Edward C Benzel; Thomas E Mroz
Journal:  Global Spine J       Date:  2017-08-15

4.  Retrospective evaluation of cervical fusion with DTRAX (R) cervical cage.

Authors:  Hamed Yazdanshenas; Ethan Osias; Richard Hwang; Don Y Park; Elizabeth Lord; Arya N Shamie
Journal:  J Craniovertebr Junction Spine       Date:  2022-03-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.