Are dental color measuring devices CIE compliant?

Karl Martin Lehmann1, Alessandro Devigus, Christopher Igiel, Michael Weyhrauch, Irene Schmidtmann, Stefan Wentaschek, Herbert Scheller.   

Abstract

OBJECTIVES: This in vitro study evaluated the accordance of dental color measurement devices with a Commission Internationale de l'Eclairage (CIE)-compliant reference system, by comparing the CIE lightness, chroma, and hue (L*C*h°) color coordinates of ceramic samples.
METHODS: Four color measurement devices: Vita Easyshade® Advance (A); DeguDent Shadepilot™ (B); X-Rite® Shadevision (C); and Crystaleye Olympus (D), were compared with a CIE-compliant reference system by recording the L*C*h° color coordinates of ceramic samples matching the tooth colors of the Vita Linearguide 3D-Master®, under standardized test conditions. Differences between regression lines for the dental color measurement device data and regression lines for the CIE-compliant reference system data were evaluated.
RESULTS: All devices offered high intraclass correlation coefficients (0.9771- 0.9999) for the L*C*h° color coordinates. The regression lines of the L* and C* coordinates for device A were steeper than those of the CIE-compliant reference system; the regression lines for devices B, C, and D were nearly parallel to those of the reference system, but with an offset. The regression lines of the h° coordinates for all devices were almost parallel to those of the reference system, with slopes near 1. Excluding the L* and h° coordinate measurements of device A, the measurements with the devices exhibited deviations from the reference system that were greater than those expected by chance (P < 0.0002).
CONCLUSIONS: The dental color measurement devices assessed here offered excellent reproducibility, but showed significant deviations from the CIE-compliant reference system regarding the L*C*h° color coordinates.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22908079

Source DB:  PubMed          Journal:  Eur J Esthet Dent        ISSN: 1862-0612


  5 in total

1.  Comparison of the Easyshade Compact and Advance in vitro and in vivo.

Authors:  Andreas Zenthöfer; Tomas Cabrera; Nicoleta Corcodel; Peter Rammelsberg; Alexander J Hassel
Journal:  Clin Oral Investig       Date:  2013-10-08       Impact factor: 3.573

2.  Interdevice agreement of eight equivalent dental color measurement devices.

Authors:  M Weyhrauch; C Igiel; A M Pabst; S Wentaschek; H Scheller; K M Lehmann
Journal:  Clin Oral Investig       Date:  2015-03-24       Impact factor: 3.573

3.  In vivo and in vitro spectrophotometric evaluation of upper central incisors before and after extraction.

Authors:  Enric Soler; Fernando Duran-Sindreu; Juan Basilio; Miguel Roig; Stefano Ardu; Juan R Mayoral
Journal:  Clin Oral Investig       Date:  2017-01-05       Impact factor: 3.573

4.  The reproducibility of electronic color measurements of the marginal gingiva.

Authors:  Henning Staedt; Eva Mally; Herbert Scheller; Stefan Wentaschek; Peer Wolfgang Kämmerer; Adrian Kasaj; Alessandro Devigus; Karl Martin Lehmann
Journal:  Clin Oral Investig       Date:  2020-06-22       Impact factor: 3.573

5.  Mouthrinses affect color stability of composite.

Authors:  Arshia Rashid Baig; Deepa Deepak Shori; Pratima Ramakrishna Shenoi; Syed Navid Ali; Sanjay Shetti; Alkesh Godhane
Journal:  J Conserv Dent       Date:  2016 Jul-Aug
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.