Literature DB >> 22902810

Evaluation of an algorithm based on single-condition decision rules for binary classification of 12-lead ambulatory ECG recording quality.

Luigi Yuri Di Marco1, Wenfeng Duan, Marjan Bojarnejad, Dingchang Zheng, Susan King, Alan Murray, Philip Langley.   

Abstract

A new algorithm for classifying ECG recording quality based on the detection of commonly observed ECG contaminants which often render the ECG unusable for diagnostic purposes was evaluated. Contaminants (baseline drift, flat line, QRS-artefact, spurious spikes, amplitude stepwise changes, noise) were detected on individual leads from joint time-frequency analysis and QRS amplitude. Classification was based on cascaded single-condition decision rules (SCDR) that tested levels of contaminants against classification thresholds. A supervised learning classifier (SLC) was implemented for comparison. The SCDR and SLC algorithms were trained on an annotated database (Set A, PhysioNet Challenge 2011) of 'acceptable' versus 'unacceptable' quality recordings using the 'leave M out' approach with repeated random partitioning and cross-validation. Two training approaches were considered: (i) balanced, in which training records had equal numbers of 'acceptable' and 'unacceptable' recordings, (ii) unbalanced, in which the ratio of 'acceptable' to 'unacceptable' recordings from Set A was preserved. For each training approach, thresholds were calculated, and classification accuracy of the algorithm compared to other rule based algorithms and the SLC using a database for which classifications were unknown (Set B PhysioNet Challenge 2011). The SCDR algorithm achieved the highest accuracy (91.40%) compared to the SLC (90.40%) in spite of its simple logic. It also offers the advantage that it facilitates reporting of meaningful causes of poor signal quality to users.

Mesh:

Year:  2012        PMID: 22902810     DOI: 10.1088/0967-3334/33/9/1435

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  5 in total

Review 1.  A review of methods for the signal quality assessment to improve reliability of heart rate and blood pressures derived parameters.

Authors:  Nicolò Gambarotta; Federico Aletti; Giuseppe Baselli; Manuela Ferrario
Journal:  Med Biol Eng Comput       Date:  2016-02-23       Impact factor: 2.602

2.  Wearable Electrocardiogram Quality Assessment Using Wavelet Scattering and LSTM.

Authors:  Feifei Liu; Shengxiang Xia; Shoushui Wei; Lei Chen; Yonglian Ren; Xiaofei Ren; Zheng Xu; Sen Ai; Chengyu Liu
Journal:  Front Physiol       Date:  2022-06-30       Impact factor: 4.755

3.  Assessing ECG signal quality indices to discriminate ECGs with artefacts from pathologically different arrhythmic ECGs.

Authors:  C Daluwatte; L Johannesen; L Galeotti; J Vicente; D G Strauss; C G Scully
Journal:  Physiol Meas       Date:  2016-07-25       Impact factor: 2.833

4.  A method to extract realistic artifacts from electrocardiogram recordings for robust algorithm testing.

Authors:  Loriano Galeotti; Christopher G Scully
Journal:  J Electrocardiol       Date:  2018-08-18       Impact factor: 1.438

5.  Quality estimation of the electrocardiogram using cross-correlation among leads.

Authors:  Eduardo Morgado; Felipe Alonso-Atienza; Ricardo Santiago-Mozos; Óscar Barquero-Pérez; Ikaro Silva; Javier Ramos; Roger Mark