OBJECTIVE: The PolyScope(®) (Lumenis, Santa Clara, CA) is a disposable flexible ureteroscope composed of a single-use flexible catheter and a reusable fiber optic bundle. We tested the resilience of the optical properties of the reusable optical fiber after repeat sterilization.
METHODS: Optical properties evaluated were resolution, distortion, and light transmission. Two different optical fibers were tested, one for each sterilization method. Measurements on each fiber were obtained at baseline and were repeated after each set of 10 sterilization cycles for a total of 100 cycles. To measure resolution, the 1951 USAF Contrast Resolution Chart Target was used to subjectively identify the clearest image on the grid by two different operators. To measure distortion, three images of the distortion grid target were captured with a digital camera and analyzed within Image Pro Plus Software. Light transmission was measured using a digital light meter at 50% and 100% transmission.
RESULTS: There was no significant difference in resolution after 100 sterilization cycles between either Steris or Sterrad NX. Distortion changed by 8.7% and 11.2% for Steris and Sterrad NX, respectively. For Steris at 100% light transmission, baseline was 59 foot-candles and this increased to 85 foot-candles after 100 cycles. For Sterrad NX at 100% light transmission, baseline was 50 foot-candles and this increased to 92 foot-candles after 100 cycles. Results were similar at 50%. There was no significant difference between groups at 50%, 100%, or between sterilization modalities.
CONCLUSION: The PolyScope fiber optic bundle appears durable after 100 cycles of sterilization with either Steris or Sterrad NX.

Mesh:

Year:  2012        PMID: 22894548     DOI: 10.1089/end.2012.0323

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  7 in total

1.  Modular flexible ureteroscopy and holmium laser lithotripsy for the treatment of renal and proximal ureteral calculi: A single-surgeon experience of 382 cases.

Authors:  Zejun Yan; Guohai Xie; Hesheng Yuan; Yue Cheng
Journal:  Exp Ther Med       Date:  2015-08-24       Impact factor: 2.447

Review 2.  Retrograde intrarenal surgery for renal stones - Part 1.

Authors:  Ben Van Cleynenbreugel; Özcan Kılıç; Murat Akand
Journal:  Turk J Urol       Date:  2017-06-01

Review 3.  Disposable devices for RIRS: where do we stand in 2013? What do we need in the future?

Authors:  Richard H Shin; Michael E Lipkin; Glenn M Preminger
Journal:  World J Urol       Date:  2014-07-30       Impact factor: 4.226

Review 4.  Single-use flexible ureteropyeloscopy: a systematic review.

Authors:  N F Davis; M R Quinlan; C Browne; N R Bhatt; R P Manecksha; F T D'Arcy; N Lawrentschuk; D M Bolton
Journal:  World J Urol       Date:  2017-11-24       Impact factor: 4.226

5.  Prospective comparative study between the effect of CIDEX® OPA and STERRAD NX on the durability of digital flexible ureteroscope.

Authors:  Saeed H Al Qahtani; Mohamed H Abdelhamied; Abdulrahman H AlMuhrij; Mizyad Y Al Rawashada; Ahmed M Al Askar; Amr M Abdelhamid; Tarek K Fath El-Bab; Ehab M Galal; Mahmoud S Eladawy
Journal:  World J Urol       Date:  2019-05-13       Impact factor: 4.226

Review 6.  Pushing the boundaries of ureteroscopy: current status and future perspectives.

Authors:  Petrisor Geavlete; Razvan Multescu; Bogdan Geavlete
Journal:  Nat Rev Urol       Date:  2014-06-03       Impact factor: 14.432

Review 7.  A comprehensive literature-based equation to compare cost-effectiveness of a flexible ureteroscopy program with single-use versus reusable devices.

Authors:  Giovanni S Marchini; Fábio C Torricelli; Carlos A Batagello; Manoj Monga; Fábio C Vicentini; Alexandre Danilovic; Miguel Srougi; William C Nahas; Eduardo Mazzucchi
Journal:  Int Braz J Urol       Date:  2019 Jul-Aug       Impact factor: 1.541

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.