Literature DB >> 22884619

Auxiliary locking plate improves fracture stability and healing in intertrochanteric fractures fixated by intramedullary nail.

Sebastian Eberle1, Johannes Gabel, Sven Hungerer, Stefanie Hoffmann, Robert Pätzold, Peter Augat, Volker Bühren.   

Abstract

BACKGROUND: Intertrochanteric fractures present a significant management challenge due to their low inherent stability. The objective of this study was to determine whether an auxiliary locking plate decreases interfragmentary motions and improves fracture healing in intertrochanteric fractures treated by intramedullary nail.
METHODS: Biomechanical tests and a clinical retrospective study in intertrochanteric to subtrochanteric nonunions were performed. Six synthetic femurs were osteotomized intertrochanterically and fixated with a long gamma nail and an additional locking compression plate. Mechanical tests were conducted that simulated the hip joint force during gait cycle. Following the initial test, the locking compression plate (LCP) was removed from each specimen and the test was repeated. Interfragmentary motions, strains on implants and osteosynthesis stiffness were determined. For the clinical part of the study, 13 intertrochanteric to subtrochanteric nonunions were treated with revisional long gamma nail and additional locking compression plate. Complications and time to union were determined.
FINDINGS: Biomechanically, interfragmentary rotation was 48% smaller (P=0.047) and interfragmentary shear movement was 42% smaller (P=0.007) with locking compression plate. Strains on the nail decreased by 20-27% (P<0.027) and the osteosynthesis stiffness increased by 23% (P=0.005) with locking compression plate. Clinically, fracture healing was achieved in eleven out of 13 patients after 9.0months (range 4 to 22months).
INTERPRETATION: The findings of our study indicate that auxiliary locked plating considerably improves biomechanical performance and results in successful healing of unstable intertrochanteric to subtrochanteric femur fractures.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22884619     DOI: 10.1016/j.clinbiomech.2012.07.008

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  6 in total

1.  [Evolution and principles of intramedullary locked nailing].

Authors:  C Hierholzer; J Friederichs; P Augat; A Woltmann; O Trapp; V Bühren; C von Rüden
Journal:  Unfallchirurg       Date:  2018-03       Impact factor: 1.000

2.  The effect on outcomes of the application of circumferential cerclage cable following intramedullary nailing in reverse intertrochanteric femoral fractures.

Authors:  Ahmet Imerci; Nevres Hurriyet Aydogan; Kursad Tosun
Journal:  Eur J Orthop Surg Traumatol       Date:  2018-12-06

3.  Successful management of lower limb nonunion using locking plates and bone graft with retention of intramedullary nail.

Authors:  Hong-Jun Zhang; Shuai Wang; Yong-Hui Dong; Wen-Di Zheng; Zhe Sun; Jia Zheng
Journal:  Medicine (Baltimore)       Date:  2019-03       Impact factor: 1.889

4.  Supplemental cerclage wiring in angle stable plate fixation of distal tibial spiral fractures enables immediate post-operative full weight-bearing: a biomechanical analysis.

Authors:  Sabrina Sandriesser; Stefan Förch; Edgar Mayr; Falk Schrödl; Christian von Rüden; Peter Augat
Journal:  Eur J Trauma Emerg Surg       Date:  2020-09-28       Impact factor: 3.693

5.  Results of revision intramedullary nailing with and without auxillary plate in aseptic trochanteric and subtrochanteric nonunion.

Authors:  Dorien Schneidmueller; Christian von Rüden; Christina Dietze; Andreas Brand; Jan Friederichs; Fabian Stuby
Journal:  Eur J Trauma Emerg Surg       Date:  2021-04-16       Impact factor: 2.374

Review 6.  Biomechanics of Osteoporotic Fracture Fixation.

Authors:  Marianne Hollensteiner; Sabrina Sandriesser; Emily Bliven; Christian von Rüden; Peter Augat
Journal:  Curr Osteoporos Rep       Date:  2019-12       Impact factor: 5.096

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.