Correlation between CT attenuation value and iodine concentration in vitro: discrepancy between gemstone spectral imaging on single-source dual-energy CT and traditional polychromatic X-ray imaging.

Le Wang1, Bin Liu, Xing-Wang Wu, Jie Wang, Yong Zhou, Wan-Qin Wang, Xiao-Hong Zhu, Yong-Qiang Yu, Xiao-Hu Li, Shuai Zhang, Yun Shen.   

Abstract

INTRODUCTION: To assess the relation between CT attenuation value and iodine concentration in vitro, using gemstone spectral imaging (GSI) with single-source dual-energy CT and traditional polychromatic X-ray imaging (TPXI), respectively.
METHODS: A polypropylene phantom with eight test tubes in which iodine concentrations of solution were 0.4, 0.7, 2, 5, 10, 20, 30 and 50 mg/mL underwent GSI and traditional polychromatic X-ray scans (80, 100, 120 and 140 kV(p)), using single-source dual-energy spectral CT (Discovery CT750HD; GE Healthcare Technologies, Milwaukee, WI, USA) at the same tube speed of 0.8 s/rotation. All spectral imaging data were analysed with GSI viewer to obtain monochromatic images (50-140 keV, interval of 10 keV). Computed tomography attenuation values of iodine solution were measured with the same size of regions of interest and at the exact same level for both monochromatic and polychromatic images. The relation between CT attenuation value and iodine concentration was examined.
RESULTS: A linear correlation was found between CT attenuation value and iodine concentration for both monochromatic and polychromatic images. Moreover, the fitting coefficients for CT attenuation values and iodine concentrations were closer to one with GSI (r(2) = 0.99824-0.99996) than that with TPXI (r(2) = 0.99640-0.99736).
CONCLUSIONS: Owing to the better correlation coefficients between CT attenuation value and iodine concentration, GSI may be a preferred method for quantitative measurement compared with TPXI.
© 2012 The Authors. Journal of Medical Imaging and Radiation Oncology © 2012 The Royal Australian and New Zealand College of Radiologists.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22883644     DOI: 10.1111/j.1754-9485.2012.02379.x

Source DB:  PubMed          Journal:  J Med Imaging Radiat Oncol        ISSN: 1754-9477            Impact factor:   1.735


  13 in total

1.  Monochromatic energy computed tomography image for active intestinal hemorrhage: a model investigation.

Authors:  Wen-Dong Liu; Xing-Wang Wu; Jun-Mei Hu; Bin Wang; Bin Liu
Journal:  World J Gastroenterol       Date:  2015-01-07       Impact factor: 5.742

2.  Determination of glomerular filtration rate (GFR) from fractional renal accumulation of iodinated contrast material: a convenient and rapid single-kidney CT-GFR technique.

Authors:  XiaoDong Yuan; Wei Tang; WenWei Shi; Libao Yu; Jing Zhang; Qing Yuan; Shan You; Ning Wu; Guokun Ao; Tingting Ma
Journal:  Eur Radiol       Date:  2018-02-09       Impact factor: 5.315

3.  Iodine concentration: a new, important characteristic of the spot sign that predicts haematoma expansion.

Authors:  Fan Fu; Shengjun Sun; Liping Liu; Jianying Li; Yaping Su; Yingying Li
Journal:  Eur Radiol       Date:  2018-04-19       Impact factor: 5.315

4.  Comparing the diagnostic utility of sacroiliac spectral CT and MRI in axial spondyloarthritis.

Authors:  Ping Zhang; Kai Hu Yu; Rui Min Guo; Jun Ran; Yao Liu; John Morelli; Val A Runge; Xiao Ming Li
Journal:  Br J Radiol       Date:  2015-12-16       Impact factor: 3.039

5.  Identification of epidermal growth factor receptor mutations in pulmonary adenocarcinoma using dual-energy spectral computed tomography.

Authors:  Meng Li; Li Zhang; Wei Tang; Yu-Jing Jin; Lin-Lin Qi; Ning Wu
Journal:  Eur Radiol       Date:  2018-10-26       Impact factor: 5.315

6.  Quantitative spectral CT evaluation of kidney tumors with the stretched-exponential nonlinear regression analysis model.

Authors:  Xiaotong Yang; Wenge Sun; Desheng Huang; Hui Li; Yu Zhao; Peiling Li; Yi Liu
Journal:  Quant Imaging Med Surg       Date:  2021-02

7.  Imaging assessment of a portable hemodialysis device: detection of possible failure modes and monitoring of functional performance.

Authors:  Olufoladare G Olorunsola; Steven H Kim; Ryan Chang; Yuo-Chen Kuo; Steven W Hetts; Alex Heller; Rishi Kant; Maythem Saeed; William H Fissell; Shuvo Roy; Mark W Wilson
Journal:  Med Instrum (Luton)       Date:  2014-03-27

8.  Computed Tomography Number Measurement Consistency Under Different Beam Hardening Conditions: Comparison Between Dual-Energy Spectral Computed Tomography and Conventional Computed Tomography Imaging in Phantom Experiment.

Authors:  Tian He; Xiaojun Qian; Renyou Zhai; Zongtao Yang
Journal:  J Comput Assist Tomogr       Date:  2015 Nov-Dec       Impact factor: 1.826

9.  Diagnostic value of single-source dual-energy spectral computed tomography in differentiating parotid gland tumors: initial results.

Authors:  Lin Li; Yanfeng Zhao; Dehong Luo; Liang Yang; Lei Hu; Xinming Zhao; Yong Wang; Wensheng Liu
Journal:  Quant Imaging Med Surg       Date:  2018-07

10.  Evaluation of energy spectrum CT for the measurement of thyroid iodine content.

Authors:  Weiguang Shao; Jingang Liu; Dianmei Liu
Journal:  BMC Med Imaging       Date:  2016-08-12       Impact factor: 1.930

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.