Literature DB >> 22871840

Monitoring of cornea elastic properties changes during UV-A/riboflavin-induced corneal collagen cross-linking using supersonic shear wave imaging: a pilot study.

Thu-Mai Nguyen1, Jean-François Aubry, David Touboul, Mathias Fink, Jean-Luc Gennisson, Jeremy Bercoff, Mickael Tanter.   

Abstract

PURPOSE: Keratoconus disease or post-LASIK corneal ectasia are increasingly treated using UV-A/riboflavin-induced corneal collagen cross-linking (CXL). However, this treatment suffers from a lack of techniques to provide an assessment in real-time of the CXL effects. Here, we investigated the potential interest of corneal elasticity as a biomarker of the efficacy of this treatment.
METHODS: For this purpose, supersonic shear wave imaging (SSI) was performed both ex vivo and in vivo on porcine eyes before and after CXL. Based on ultrasonic scanners providing ultrafast frame rates (~30 kHz), the SSI technique generates and tracks the propagation of shear waves in tissues. It provides two- and three-dimensional (2-D and 3-D) quantitative maps of the corneal elasticity.
RESULTS: After CXL, quantitative maps of corneal stiffness clearly depicted the cross-linked area with a typical 200-μm lateral resolution. The CXL resulted in a 56% ± 15% increase of the shear wave speed for corneas treated in vivo (n = 4).
CONCLUSIONS: The in vivo CXL experiments performed on pigs demonstrated that the quantitative estimation of local stiffness and the 2-D elastic maps of the corneal surface provide an efficient way to monitor the local efficacy of corneal cross-linking.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22871840     DOI: 10.1167/iovs.11-9142

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  23 in total

1.  Noninvasive measurement of wave speed of porcine cornea in ex vivo porcine eyes for various intraocular pressures.

Authors:  Boran Zhou; Arthur J Sit; Xiaoming Zhang
Journal:  Ultrasonics       Date:  2017-06-07       Impact factor: 2.890

2.  Spatially heterogeneous corneal mechanical responses before and after riboflavin-ultraviolet-A crosslinking.

Authors:  Joel R Palko; Junhua Tang; Benjamin Cruz Perez; Xueliang Pan; Jun Liu
Journal:  J Cataract Refract Surg       Date:  2014-04-18       Impact factor: 3.351

3.  Analysis of the effects of curvature and thickness on elastic wave velocity in cornea-like structures by finite element modeling and optical coherence elastography.

Authors:  Zhaolong Han; Jiasong Li; Manmohan Singh; Salavat R Aglyamov; Chen Wu; Chih-Hao Liu; Kirill V Larin
Journal:  Appl Phys Lett       Date:  2015-06-12       Impact factor: 3.791

4.  Assessing the effects of riboflavin/UV-A crosslinking on porcine corneal mechanical anisotropy with optical coherence elastography.

Authors:  Manmohan Singh; Jiasong Li; Zhaolong Han; Raksha Raghunathan; Achuth Nair; Chen Wu; Chih-Hao Liu; Salavat Aglyamov; Michael D Twa; Kirill V Larin
Journal:  Biomed Opt Express       Date:  2016-12-19       Impact factor: 3.732

5.  Optical coherence elastography for evaluating customized riboflavin/UV-A corneal collagen crosslinking.

Authors:  Manmohan Singh; Jiasong Li; Srilatha Vantipalli; Zhaolong Han; Kirill V Larin; Michael D Twa
Journal:  J Biomed Opt       Date:  2017-09-01       Impact factor: 3.170

6.  In-vivo 3D corneal elasticity using air-coupled ultrasound optical coherence elastography.

Authors:  Zi Jin; Reza Khazaeinezhad; Jiang Zhu; Junxiao Yu; Yueqiao Qu; Youmin He; Yan Li; Tomas E Gomez Alvarez-Arenas; Fan Lu; Zhongping Chen
Journal:  Biomed Opt Express       Date:  2019-11-14       Impact factor: 3.732

7.  Noncontact depth-resolved micro-scale optical coherence elastography of the cornea.

Authors:  Shang Wang; Kirill V Larin
Journal:  Biomed Opt Express       Date:  2014-10-06       Impact factor: 3.732

8.  Effects of Thickness on Corneal Biomechanical Properties Using Optical Coherence Elastography.

Authors:  Srilatha Vantipalli; Jiasong Li; Manmohan Singh; Salavat R Aglyamov; Kirill V Larin; Michael D Twa
Journal:  Optom Vis Sci       Date:  2018-04       Impact factor: 1.973

9.  Shear wave imaging optical coherence tomography (SWI-OCT) for ocular tissue biomechanics.

Authors:  Shang Wang; Kirill V Larin
Journal:  Opt Lett       Date:  2014-01-01       Impact factor: 3.776

Review 10.  Application of Elastography for the Noninvasive Assessment of Biomechanics in Engineered Biomaterials and Tissues.

Authors:  Woong Kim; Virginia L Ferguson; Mark Borden; Corey P Neu
Journal:  Ann Biomed Eng       Date:  2016-01-20       Impact factor: 3.934

View more