Literature DB >> 22865254

The articular surface replacement implant recall: a United Kingdom district hospital experience.

George S Whitwell1, Ashokan Shine, Steve K Young.   

Abstract

We present our experience of the articular surface replacement (ASR) hip and the implant recall process. One hundred and twenty-one ASR components were implanted (21 resurfacing hip arthroplasty (RHA) and 100 ASR/XL modular total hip replacements). At the time of the implant recall in August 2010 there were 111 surviving hips (92%) with a mean follow-up of 44 months. Nine hips had been revised and one had been listed for revision surgery. Ninety-two percent of surviving implants were reviewed in the recall clinics, and blood metal ion levels or ultrasound scans were indicated in 38 hips (34%). Immediately after the recall process seven hips (6 ASR/XL and 1 RHA) were listed for revision and a further 9 were kept under close surveillance. One year after completion of the recall process 23 hips (19 ASR/XL and 4 RHA's) had been revised. A diagnosis of adverse reaction to metal debris (ARMD) was made at surgery in all but two hips. Our current revision rate for ASR RHA is 19% (mean follow-up 62 months, range 29-80) and for the ASR/XL is 19% (mean follow-up 53 months, range 10-80). The 5-year cumulative survival rates with revision for any reason for the ASR/XL, was 80.8% (95% confidence interval 72.0 - 89.5). Given experience elsewhere we expect this rate may increase significantly with time.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22865254     DOI: 10.5301/HIP.2012.9351

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  9 in total

1.  CORR Insights(®): Early Lessons From a Worldwide, Multicenter, Followup Study of the Recalled Articular Surface Replacement Hip System.

Authors:  Jean-Noel A Argenson
Journal:  Clin Orthop Relat Res       Date:  2015-09-08       Impact factor: 4.176

2.  Early Lessons From a Worldwide, Multicenter, Followup Study of the Recalled Articular Surface Replacement Hip System.

Authors:  Rami Madanat; Daniel K Hussey; Gabrielle S Donahue; Hollis G Potter; Robert Wallace; Charles Bragdon; Orhun Muratoglu; Henrik Malchau
Journal:  Clin Orthop Relat Res       Date:  2015-08-27       Impact factor: 4.176

3.  Editorial Comment: Symposium: Advances in Polyethylene Biomaterials.

Authors:  Steven M Kurtz; Pierangiola Bracco
Journal:  Clin Orthop Relat Res       Date:  2016-11-29       Impact factor: 4.176

4.  Five Hundred Fifty-five Retrieved Metal-on-metal Hip Replacements of a Single Design Show a Wide Range of Wear, Surface Features, and Histopathologic Reactions.

Authors:  Sang-Hyun Park; Zhen Lu; Robert S Hastings; Patricia A Campbell; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

5.  The Irish National Joint Registry: where are we now?

Authors:  B J O'Neill; M Nugent; J P Cashman; S J O'Flanagan; P Keogh; P J Kenny
Journal:  Ir J Med Sci       Date:  2013-06-19       Impact factor: 1.568

6.  Are Females at Greater Risk for Revision Surgery After Hip Resurfacing Arthroplasty With the Articular Surface Replacement Prosthesis?

Authors:  Gabrielle S Donahue; Viktor Lindgren; Vincent P Galea; Rami Madanat; Orhun Muratoglu; Henrik Malchau
Journal:  Clin Orthop Relat Res       Date:  2016-04-27       Impact factor: 4.176

7.  Hip arthroplasty with the articular surface replacement (ASR) system: survivorship analysis and functional outcomes.

Authors:  S Shemesh; Y Kosashvili; S Heller; E Sidon; L Yaari; N Cohen; S Velkes
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-07-11

8.  Clinical Results and Serum Metal Ion Concentrations following Ceramic-on-Metal Total Hip Arthroplasty at a Mean Follow-Up of 60 Months.

Authors:  W Maurer-Ertl; D Pranckh-Matzke; J Friesenbichler; G Bratschitsch; L A Holzer; M Maier; A Leithner
Journal:  Biomed Res Int       Date:  2017-03-08       Impact factor: 3.411

Review 9.  Prevalence of Failure due to Adverse Reaction to Metal Debris in Modern, Medium and Large Diameter Metal-on-Metal Hip Replacements--The Effect of Novel Screening Methods: Systematic Review and Metaregression Analysis.

Authors:  Aleksi Reito; Olli Lainiala; Petra Elo; Antti Eskelinen
Journal:  PLoS One       Date:  2016-03-01       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.