Effects of thread shape on the pullout strength of miniscrews.

Antonio Gracco1, Costantino Giagnorio, Serena Incerti Parenti, Giulio Alessandri Bonetti, Giuseppe Siciliani.   

Abstract

INTRODUCTION: The aim of this study was to determine the effects of variations in thread shape on the axial pullout strength of orthodontic miniscrews.
METHODS: A total of 35 miniscrews, 7 of each design being considered, were tested by performing pullout tests on a synthetic bone support. We used self-tapping and self-drilling miniscrews having a diameter of 2 mm and a thread shaft length of 12 mm (the longest and the largest supplied by the manufacturer). A buttress reverse thread shape served as the control design and was tested against 4 experimental designs, each manufactured with a modification in thread shape while maintaining all other characteristics. The experimental groups had the following thread designs: buttress, 75° joint profile, rounded, and trapezoidal.
RESULTS: The control group with a buttress reverse thread shape had consistently higher pullout strength values than did the other designs. A statistically significant reduction in pullout force was found between the buttress reverse and the buttress thread miniscrews.
CONCLUSIONS: Thread design influenced the resistance to pullout of the orthodontic miniscrews. The buttress reverse thread shape provided the greatest pullout strength.
Copyright © 2012 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

Mesh:

Substances:

Year:  2012        PMID: 22858327     DOI: 10.1016/j.ajodo.2012.03.023

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  9 in total

1.  Does Change in Thread Shape Influence the Pull Out Strength of Mini Implants? An In vitro Study.

Authors:  Aniruddh V Yashwant; S Dilip; R Krishnaraj; K Ravi
Journal:  J Clin Diagn Res       Date:  2017-05-01

2.  Analysis of biological and structural factors implicated in the clinical success of orthodontic miniscrews at posterior maxillary interradicular sites.

Authors:  M Palone; A Darsiè; G B Maino; G Siciliani; G A Spedicato; L Lombardo
Journal:  Clin Oral Investig       Date:  2021-11-27       Impact factor: 3.573

3.  Mechanical stability of orthodontic miniscrew depends on a thread shape.

Authors:  Keiichiro Watanabe; Bobby Mitchell; Takuma Sakamaki; Yuzo Hirai; Do-Gyoon Kim; Toru Deguchi; Masakazu Suzuki; Kanji Ueda; Eiji Tanaka
Journal:  J Dent Sci       Date:  2021-12-11       Impact factor: 3.719

4.  The Effects of Different Miniscrew Thread Designs and Force Directions on Stress Distribution by 3-dimensional Finite Element Analysis.

Authors:  Hamidreza Fattahi; Shabnam Ajami; Ali Nabavizadeh Rafsanjani
Journal:  J Dent (Shiraz)       Date:  2015-12

5.  The effect of screw thread length on initial stability of Schatzker type 1 tibial plateau fracture fixation: a biomechanical study.

Authors:  Ahmet Salduz; Fevzi Birisik; Gokhan Polat; Bugra Bekler; Ergun Bozdag; Onder Kilicoglu
Journal:  J Orthop Surg Res       Date:  2016-11-22       Impact factor: 2.359

6.  Comparative evaluation of insertion torque and mechanical stability for self-tapping and self-drilling orthodontic miniscrews - an in vitro study.

Authors:  Michele Tepedino; Francesco Masedu; Claudio Chimenti
Journal:  Head Face Med       Date:  2017-05-30       Impact factor: 2.151

7.  In vitro study of the orthodontic mini-implants influence on the growth of human osteoblasts.

Authors:  Andreea Moldoveanu; Mihnea Ioan Nicolescu; Mirela Veronica Bucur; George Gabriel Moldoveanu; Cristian Funieru; Ionela Victoria Neagoe; Gina Manda; Tamara Rahela Ioana; Lucian Toma Ciocan
Journal:  Rom J Morphol Embryol       Date:  2021 Jul-Sep       Impact factor: 0.833

8.  Primary Stability of Dental Implants in Low-Density (10 and 20 pcf) Polyurethane Foam Blocks: Conical vs Cylindrical Implants.

Authors:  Luca Comuzzi; Margherita Tumedei; Ana Emilia Pontes; Adriano Piattelli; Giovanna Iezzi
Journal:  Int J Environ Res Public Health       Date:  2020-04-11       Impact factor: 3.390

9.  Risk Factor Analysis of Facet Fusion Following Cervical Lateral Mass Screw Fixation with a Minimum 1-Year Follow-up: Assessment of Maximal Insertional Screw Torque and Incidence of Loosening.

Authors:  Ryo Kanematsu; Junya Hanakita; Toshiyuki Takahashi; Manabu Minami; Tomoo Inoue; Fumiaki Honda
Journal:  Neurol Med Chir (Tokyo)       Date:  2020-11-18       Impact factor: 1.742

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.