Literature DB >> 22848098

Validation of the efficacy of a solar-thermal powered autoclave system for off-grid medical instrument wet sterilization.

Tremayne Kaseman1, Jean Boubour, Douglas A Schuler.   

Abstract

This work describes the efficacy of a solar-thermal powered autoclave used for the wet sterilization of medical instruments in off-grid settings where electrical power is not readily available. Twenty-seven trials of the solar-thermal powered system were run using an unmodified non-electric autoclave loaded with a simulated bundle of medical instruments and biological test agents. Results showed that in 100% of the trials the autoclave achieved temperatures in excess of 121°C for 30 minutes, indicator tape displayed visible reactions to steam sterilization, and biological tests showed that microbial agents had been eliminated, in compliance with the Centers for Disease Control and Prevention requirements for efficacious wet sterilization.

Mesh:

Substances:

Year:  2012        PMID: 22848098      PMCID: PMC3516307          DOI: 10.4269/ajtmh.2012.12-0061

Source DB:  PubMed          Journal:  Am J Trop Med Hyg        ISSN: 0002-9637            Impact factor:   2.345


  5 in total

Review 1.  Lessons from nosocomial viral haemorrhagic fever outbreaks.

Authors:  Susan P Fisher-Hoch
Journal:  Br Med Bull       Date:  2005-12-22       Impact factor: 4.291

2.  Device-associated nosocomial infections in 55 intensive care units of 8 developing countries.

Authors:  Victor D Rosenthal; Dennis G Maki; Reinaldo Salomao; Carlos Alvarez Moreno; Yatin Mehta; Francisco Higuera; Luis E Cuellar; Ozay Akan Arikan; Rédouane Abouqal; Hakan Leblebicioglu
Journal:  Ann Intern Med       Date:  2006-10-17       Impact factor: 25.391

3.  The relationship between therapeutic injections and high prevalence of hepatitis C infection in Hafizabad, Pakistan.

Authors:  S P Luby; K Qamruddin; A A Shah; A Omair; O Pahsa; A J Khan; J B McCormick; F Hoodbhouy; S Fisher-Hoch
Journal:  Epidemiol Infect       Date:  1997-12       Impact factor: 2.451

4.  The role of parenteral antischistosomal therapy in the spread of hepatitis C virus in Egypt.

Authors:  C Frank; M K Mohamed; G T Strickland; D Lavanchy; R R Arthur; L S Magder; T El Khoby; Y Abdel-Wahab; E S Aly Ohn; W Anwar; I Sallam
Journal:  Lancet       Date:  2000-03-11       Impact factor: 79.321

5.  A hand-powered, portable, low-cost centrifuge for diagnosing anemia in low-resource settings.

Authors:  Jocelyn Brown; Lauren Theis; Lila Kerr; Nazima Zakhidova; Kelly O'Connor; Margaret Uthman; Z Maria Oden; Rebecca Richards-Kortum
Journal:  Am J Trop Med Hyg       Date:  2011-08       Impact factor: 2.345

  5 in total
  1 in total

1.  A Shipping Container-Based Sterile Processing Unit for Low Resources Settings.

Authors:  Jean Boubour; Katherine Jenson; Hannah Richter; Josiah Yarbrough; Z Maria Oden; Douglas A Schuler
Journal:  PLoS One       Date:  2016-03-23       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.