Literature DB >> 22841532

Bone properties affect loosening of half-pin external fixators at the pin-bone interface.

Finn E Donaldson1, Pankaj Pankaj, A Hamish R W Simpson.   

Abstract

INTRODUCTION: Local bone yielding at the pin-bone interface of external fixation half-pins has been known to initiate fixator loosening. Deterioration of bone properties due to ageing and disease can lead to an increase in the risk of pin loosening. This study determines the extent, locations and mechanics of bone yielding for unilateral external fixation systems at the tibial midshaft with changes in age-related bone structure and properties. The study also evaluates the effect of the number of pins used in the fixation system and use of titanium pins (in place of steel) on bone yielding.
METHODS: We employ nonlinear finite element (FE) simulations. Strain-based plasticity is used to simulate bone yielding within FE analyses. Our analyses also incorporate contact behaviour at pin-bone interfaces, orthotropic elasticity and periosteal-endosteal variation of bone properties.
RESULTS: The results show that peri-implant yielded bone volume increases by three times from young to old-aged cases. The use of three, rather than two half-pins (on either side of the fracture), reduces the volume of yielded bone by 80% in all age groups. The use of titanium half-pins resulted in approximately 60-65% greater volumes of yielded bone.
CONCLUSIONS: We successfully simulate half-pin loosening at the bone-implant interface which has been found to occur clinically. Yielding across the full cortical thickness may explain the poor performance of these devices for old-aged cases. The models are able to identify patients particularly at risk of half-pin loosening, who may benefit from alternative fixator configurations or techniques such as those using pre-tensioned fine wires.
Copyright © 2012 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22841532     DOI: 10.1016/j.injury.2012.07.001

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  9 in total

1.  Biomechanical features of six design of the delta external fixator for treating Pilon fracture: a finite element study.

Authors:  Muhammad Hanif Ramlee; Mohd Ayub Sulong; Evelyn Garcia-Nieto; Daniel Angure Penaranda; Antonio Ros Felip; Mohammed Rafiq Abdul Kadir
Journal:  Med Biol Eng Comput       Date:  2018-04-21       Impact factor: 2.602

2.  Effect of the starting point of half-pin insertion on the insertional torque of the pin at the tibia.

Authors:  Sung Jae Kim; Sung Hwan Kim; Young Hwan Kim; Yong Min Chun
Journal:  Yonsei Med J       Date:  2015-01       Impact factor: 2.759

3.  Time Dependent Behaviour of Trabecular Bone at Multiple Load Levels.

Authors:  Shuqiao Xie; Krishnagoud Manda; Robert J Wallace; Francesc Levrero-Florencio; A Hamish R W Simpson; Pankaj Pankaj
Journal:  Ann Biomed Eng       Date:  2017-01-27       Impact factor: 3.934

4.  Experimental and numerical investigation into the influence of loading conditions in biomechanical testing of locking plate fracture fixation devices.

Authors:  A MacLeod; A H R W Simpson; P Pankaj
Journal:  Bone Joint Res       Date:  2018-01       Impact factor: 5.853

5.  Metaphyseal cones in revision total knee arthroplasty: The role of stems.

Authors:  Shuqiao Xie; Noel Conlisk; David Hamilton; Chloe Scott; Richard Burnett; Pankaj Pankaj
Journal:  Bone Joint Res       Date:  2020-05-16       Impact factor: 5.853

6.  Time-dependent behaviour of bone accentuates loosening in the fixation of fractures using bone-screw systems.

Authors:  S Xie; K Manda; P Pankaj
Journal:  Bone Joint Res       Date:  2018-11-03       Impact factor: 5.853

Review 7.  An engineering review of external fixators.

Authors:  P L N Fernando; Aravinda Abeygunawardane; Pci Wijesinghe; Parakrama Dharmaratne; Pujitha Silva
Journal:  Med Eng Phys       Date:  2021-11-04       Impact factor: 2.242

8.  Screw Analysis, Modeling and Experiment on the Mechanics of Tibia Orthopedic with the Ilizarov External Fixator.

Authors:  Peng Su; Sikai Wang; Yuliang Lai; Qinran Zhang; Leiyu Zhang
Journal:  Micromachines (Basel)       Date:  2022-06-11       Impact factor: 3.523

9.  The effect of plate design, bridging span, and fracture healing on the performance of high tibial osteotomy plates: An experimental and finite element study.

Authors:  A R MacLeod; G Serrancoli; B J Fregly; A D Toms; H S Gill
Journal:  Bone Joint Res       Date:  2019-01-04       Impact factor: 5.853

  9 in total