Icare ONE rebound versus Goldmann applanation tonometry in children with known or suspected glaucoma.

Nandini G Gandhi1, Sasapin G Prakalapakorn, Mays A El-Dairi, Sarah K Jones, Sharon F Freedman.   

Abstract

PURPOSE: To compare Icare ONE tonometry by clinic examiner and parent/guardian to Goldmann applanation in children with known/suspected glaucoma; to evaluate the trend in intraocular pressure (IOP) with 4 repeated measurements using Icare ONE; and to evaluate the feasibility of instructing parents on the use of the Icare ONE device in the clinic setting.
DESIGN: Nonrandomized, prospective clinical study.
METHODS: Patients with known or suspected glaucoma were recruited from the Duke pediatric glaucoma clinic. Parent(s) of all subjects gave informed consent (and children gave assent) for participation in this research study. IOP was measured using Icare ONE by clinic examiner and parent/guardian, then using Goldmann applanation (masked physician). Each parent/guardian completed an ease-of-use survey.
RESULTS: Sixty eyes (60 children) were included. Absolute value of mean IOP difference (ICare ONE clinic examiner vs Goldmann applanation) was 3.3 ± 4.0 mm Hg (P = .001). Icare ONE IOP by clinic examiner was within 3 mm Hg of Goldmann applanation in 68% (n = 41 eyes). In eyes with >3 mm Hg difference, Icare ONE was higher than Goldmann applanation in 84%. IOP demonstrated a statistically significant downward trend with repeated sequential measurements with Icare ONE (P = .0053, r(2) = 0.9894). All parents accomplished Icare ONE tonometry on at least 1 eye; 98% reported it was "easy to learn to use."
CONCLUSION: Icare ONE tonometry appears accurate and well-tolerated compared to Goldmann applanation, and holds promise for clinic and home tonometry in children. IOP trends downward with successive measurements using Icare ONE, demonstrating a possible effect from presumed patient relaxation.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22840485     DOI: 10.1016/j.ajo.2012.05.003

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  12 in total

1.  Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses.

Authors:  Alexandra Anton; Matthias Neuburger; Daniel Böhringer; Jens F Jordan
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-03-28       Impact factor: 3.117

2.  Pediatric ptosis as a sign of treatable autonomic dysfunction.

Authors:  Lara Phillips; David Robertson; Mark R Melson; Emily M Garland; Karen M Joos
Journal:  Am J Ophthalmol       Date:  2013-04-24       Impact factor: 5.258

3.  Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.

Authors:  Alicja Strzalkowska; Nina Pirlich; Julia V Stingl; Alexander K Schuster; Jasmin Rezapour; Felix M Wagner; Justus Buse; Esther M Hoffmann
Journal:  J Clin Med       Date:  2022-05-18       Impact factor: 4.964

4.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

5.  COMPARISON OF IOP READINGS USING REBOUND I CARE TONOMETER AND PERKINS APPLANATION TONOMETER IN AN AFRICAN POPULATION.

Authors:  D S Ademola-Popoola; A F Odi; T M Akande
Journal:  J West Afr Coll Surg       Date:  2014 Jan-Mar

6.  A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma.

Authors:  Dora H AlHarkan; Fatemah T Al-Shamlan; Deepak P Edward; Arif O Khan
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Apr-Jun

7.  Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients.

Authors:  Shaoying Tan; Nafees Baig; Linda Hansapinyo; Vishal Jhanji; Shihui Wei; Clement C Tham
Journal:  PLoS One       Date:  2017-03-23       Impact factor: 3.240

8.  Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

Authors:  Shaoying Tan; Marco Yu; Nafees Baig; Linda Hansapinyo; Clement C Tham
Journal:  Sci Rep       Date:  2017-02-06       Impact factor: 4.379

9.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

Review 10.  Teleglaucoma: ready to go?

Authors:  N G Strouthidis; G Chandrasekharan; J P Diamond; I E Murdoch
Journal:  Br J Ophthalmol       Date:  2014-04-10       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.