Abstract
INTRODUCTION: ISO 15189 was a new standard published in 2003 for accrediting medical laboratories. We believe that some requirements of the ISO 15189 standard are especially difficult to meet for majority of laboratories. The aim of this article was to present the frequency of nonconformities to requirements of the ISO 15189 accreditation standard, encountered during the assessments of medical laboratories in Hong Kong, during 2004 to 2009.Entities:
Mesh:
Year: 2012 PMID: 22838190 PMCID: PMC4062334 DOI: 10.11613/bm.2012.027
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Figure 1Distribution of nonconformities. S - significant nonconformities; M - minor nonconformities reported against management and technical requirements of ISO 15189:2003 in assessment of 40 laboratories conducted from 2004 to 2006.
Figure 2Distribution of nonconformities. S - significant nonconformities; M - minor nonconformities reported against management and technical requirements of ISO 15189:2007 in reassessment and surveillance visits of 61 laboratories conducted in 2009.
Number of nonconformities (NCs) reported in assessments conducted from 2004 to 2006 and in 2009.
| Total number of NCs | 616 | 580 | 233 | 813 |
| Number of NCs per laboratory | 15.4 | 9.5 | 19.4 | 11.1 |
Nonconformities (NCs) are either significant or minor, excluding recommendations
N = Number of laboratories assessed
Most common nonconformities reported against management and technical requirements of ISO 15189 in two study periods (from 2004 to 2006 and in 2009).
| Management requirements | 4.2 Quality Management System | 1 (61, 10%) | 1 (73, 13%) | 1 (35, 15%) | 1.5 | 1.2 | 2.9 |
| 4.13 Quality and technical records | 2 (19, 3%) | 4 (23, 4%) | 2 (22, 9%) | 0.5 | 0.4 | 1.8 | |
| 4.3 Document control | 3 (18, 3%) | 3 (29, 5%) | 3 (9, 4%) | 0.5 | 0.5 | 0.8 | |
| 4.6 External services and supplies | 4 (5, 1%) | 2 (30, 5%) | 3 (9, 4%) | 0.1 | 0.5 | 0.8 | |
|
| |||||||
| Technical requirements | 5.5 Examination Procedures | 1 (136, 22%) | 1 (96, 17%) | 2 (29, 12%) | 3.4 | 1.6 | 2.4 |
| 5.3 Equipment | 2 (85, 14%) | 2 (92, 16%) | 1 (32, 14%) | 2.1 | 1.5 | 2.7 | |
| 5.6 Assuring quality of examination procedures | 3 (72, 12%) | 3 (88, 15%) | 3 (25, 11%) | 1.8 | 1.4 | 2.1 | |
(A, B%) = Figures in brackets are total number of NCs, including significant and minor nonconformities; and percentage of NCs reported against the specified clause number.
N - number of laboratories assessed
61 laboratories underwent reassessment or surveillance visit; and among these 61 laboratories, 27 had their second reassessments in 2009.
12 laboratories underwent initial assessments in 2009
Comparison of performance of 27 laboratories in initial assessments and in their second reassessments.
| Significant (S) | 76 | 3.2 | 18 | 0.7 | 0.023 |
| Minor (M) | 295 | 10.4 | 151 | 5.9 | 0.080 |
| Total | 371 | 13.5 | 169 | 6.6 | 0.032 |
N - No. of laboratories assessed; NC - nonconformity